Page 1
INSTALLATION, OPERATION MAINTENANCE MANUAL COSA 9610™ GENERAL PURPOSE & EXPLOSION PROOF Version: 1.3.3 Software version: 2.3.0.0 Revision date: August 17, 2009 Print date: 08/17/09...
Page 2
COSA Instrument Corporation reserves the right to revise or to improve the design or specifications of the COSA 9610™ at any time and without notice.
TORAGE 2.3. P .................. 12 LACEMENT 2.3.1. General ....................12 2.3.2. COSA 9610™ in general purpose execution (Type 01 & 02) .....13 2.3.3. COSA 9610™ in explosion proof execution (Type 01-Ex & 02-Ex) .....13 2.4. M .............. 14 ECHANICAL ONNECTIONS 2.4.1.
Page 4
INSTALLATION, OPERATION AND MAINTENANCE MANUAL – COSA 9610™ 3.2.2. Programming the measurement parameters ...........21 3.2.3. Main screen ..................21 3.3. P ..............27 ROGRAMMING ENUS 3.3.1. Calibration Menu ..................27 3.3.2. Operation Menu ...................28 3.3.3. Measurement Menu................29 3.3.4. Output Menu ..................30 3.3.5.
Purpose of the analyzer The continuous COSA 9610™ analyzer determines online the Wobbe-index of a gas. The COSA 9610™ can be used both, as feed forward and feedback analyzer for gases mixing or as a feed forward analyzer for burning control.
INSTALLATION, OPERATION AND MAINTENANCE MANUAL – COSA 9610™ 1.2.1. Oven with oxygen sensor The gas/air mixture is burnt catalytically in an oven, which is kept at 812ºC with a burning spiral. The temperature is maintained with a temperature controller using a K-type thermocouple.
INSTALLATION, OPERATION AND MAINTENANCE MANUAL – COSA 9610™ 1.2.2. The sample system (SCS) In the sample conditioning system (SCS), gas and air are mixed in a constant proportion, such that a small excess of air is present (± 2.5% oxygen) in the flue gas. The gas and air pressure, are equalized by a dome-loaded pressure reducer (or booster relay), where the gas pressure governs the air pressure.
INSTALLATION, OPERATION AND MAINTENANCE MANUAL – COSA 9610™ 1.3. ALIBRATION ROCEDURE The analyzer can be calibrated in three different ways: • Single point calibration Only one calibration gas is used. The value of the gas is chosen middle of the measuring range.
Page 9
INSTALLATION, OPERATION AND MAINTENANCE MANUAL – COSA 9610™ The two-point calibration/validation procedure will be executed as followed: 1. Analyzer activates calibration/validation contact. 2. The procedure pauses for the specified “Calibration Start Delay” time for the external host to prepare for calibration/validation.
INSTALLATION, OPERATION AND MAINTENANCE MANUAL – COSA 9610™ 1.4. XTENDED ANGE PTION 1.4.1. Operation When the measuring range of the analyzer is larger than 40 MJ/Nm , an extended range option is available which covers a Wobbe index of 0-95 MJ/Nm .
INSTALLATION, OPERATION AND MAINTENANCE MANUAL – COSA 9610™ 1.5. COSA 9610™ W PECIFICATIONS OBBE NDEX NALYZER 1.5.1. Analyzer performance Make Cosa Instrument Corporation Service Natural gas, fuel-gas, biogas, etc. Ranges Wobbe index 0-3000 BTU/scf (0-95 MJ/Nm , span 0-1100 BTU/scf (40 MJ/Nm...
INSTALLATION, OPERATION AND MAINTENANCE MANUAL – COSA 9610™ 2. INSTALLATION 2.1. ENERAL Upon receipt and unpacking of the COSA 9610™ a visual inspection must be carried out to check for any visual damage, caused by transport. Any damage must be reported immediately to: COSA INSTRUMENT CORPORATION...
The COSA 9610™ can optionally be supplied on a 304L Stainless Steel freestanding frame. This frame is to be placed on a flat surface (i.e. concrete slab). Two holes (∅12) in the base of the frame enable the COSA 9610™ to be fixed to the floor 2.3.3.
• Location and amount of connections may vary depending on type and execution of the analyzer. • Tubing connections on the COSA 9610™ are Swagelok double ferrule compression type fittings for imperial sizes. • Only seamless and annealed imperial size instrument tubing according ASTM A-249 at a maximum permissible hardness of Rockwell B-90 may be used.
2.4.3. Calibration gasses Calibration gas composition is depending on range and process gas. COSA INSTRUMENT can advise suitable compositions. For a single range analyzer, 2 calibration gasses (low and high value) are recommended.
SCS. For termination details see specific drawings. Signal cables The COSA 9610™ has multiple input and output signals, which can be split in two groups, analog and digital signals. For both groups, an NEC approved cable gland for use with multi core cables are foreseen.
Page 17
INSTALLATION, OPERATION AND MAINTENANCE MANUAL – COSA 9610™ Signal cables The COSA 9610™ has multiple input and output signals, which can be split in two groups, analog and digital signals. Please make sure that the cable glands used are EExd/NCE certified.
The supply and drainage tube can now be connected. Because the supply line is under pressure and has been closed off on the COSA 9610™ side, the connecting fittings can be squirted with soap in order to detect any possible leaks.
INSTALLATION, OPERATION AND MAINTENANCE MANUAL – COSA 9610™ Relationship of pressure differential (gas/air) versus Wobbe-Index with different air orifices and a fixed residual oxygen concentration of 5% (theoretically determined). It is therefore recommended that a residual oxygen concentration of 2.5% is chosen corresponding to the reference point.
Page 20
INSTALLATION, OPERATION AND MAINTENANCE MANUAL – COSA 9610™ Caution The adjusting screw is located in the gas compartment of the relay. When the insert is released, gas will escape. Be prepared before the relay is going to be adjusted, e.g. have the right equipment readily available, so the gas compartment has to be open for only a minimum of time.
3.2.1. Description 3.2.2. Programming the measurement parameters Programming the COSA 9610™ is easy with the menu-controlled software. Menu can be navigated using the cursor keys. The key operations are: Menu navigation Change setting value or Enter...
Page 22
INSTALLATION, OPERATION AND MAINTENANCE MANUAL – COSA 9610™ Page 22 of 43...
Page 23
INSTALLATION, OPERATION AND MAINTENANCE MANUAL – COSA 9610™ The lower right section displays the status of each digital input and output. The signal assignments are user-programmable. Explanation outputs cal gas 1 Digital output driving low value calibration gas air actuator...
Page 24
INSTALLATION, OPERATION AND MAINTENANCE MANUAL – COSA 9610™ Flow alarm B (optional) As above Flow alarm C (optional) As above Flow alarm D (optional) As above Flow alarm E (optional) As above Page 24 of 43...
Page 25
INSTALLATION, OPERATION AND MAINTENANCE MANUAL – COSA 9610™ Global Settings Menu Tree V2.3.0.0 1. Calibration 1. Start Calibration 2. Start Validation 1. Semi-automatic 3. Calibration gases 1. Semi-automatic 2. 1-point manual 4. Settings 1. Gas 1 Wobbe 2. 1-point manual 3.
Page 26
INSTALLATION, OPERATION AND MAINTENANCE MANUAL – COSA 9610™ 4. Output (Analog & During Cal) 1. Channel A 2. Channel B 3. Channel C 4. Channel D Communication 1. Modbus 2. Serial 3. Ethernet 6. System 1. About 2. Date & Time 3.
INSTALLATION, OPERATION AND MAINTENANCE MANUAL – COSA 9610™ 3.3. ROGRAMMING ENUS The paragraph numbers correspond with the key sequence from the ‘menu tree’. In this way it is easy to see how a specific menu is reached. One exception is the measuring menu, this is reached from the main screen by either the or Enter key.
INSTALLATION, OPERATION AND MAINTENANCE MANUAL – COSA 9610™ 3.3.5. Communications Menu Submenus 1-3 with default communication settings: Page 31 of 43...
INSTALLATION, OPERATION AND MAINTENANCE MANUAL – COSA 9610™ 3.4. EMPERATURE ONTROLLED 3.4.1. Furnace temperature control unit The zirconia cell must operate at a temperature above 600ºC. For optimal performance, a set-point temperature of 812ºC was chosen. The oven that serves as the heating device for the zirconia cell is made of a metal wire-wound heating element.
INSTALLATION, OPERATION AND MAINTENANCE MANUAL – COSA 9610™ : Set oven temp to 812ºC & sample system to 40ºC for standard NOTE units, 100ºC for mid-temp models 3.4.2. Adjustment procedure temperature regulator This procedure describes the setting of the temperature controller.
INSTALLATION, OPERATION AND MAINTENANCE MANUAL – COSA 9610™ 4.2. NNUAL AINTENANCE • Check the calibration gas bottles for pressure. • Check the sample system for correct pressure. • Check the bypass flow meter for correct flow rate. • Replace the instrument air filters as required.
INSTALLATION, OPERATION AND MAINTENANCE MANUAL – COSA 9610™ 4.4. EPLACEMENT OF ESIDUAL XYGEN ENSOR Before going any further make sure that the oven has been disconnected and cooled off so that no physical injury can occur caused by coming into contact with any parts which may still be hot.
INSTALLATION, OPERATION AND MAINTENANCE MANUAL – COSA 9610™ 6. ORDERING OF SPARE PARTS All spare parts may be ordered quoting number and specification from: COSA INSTRUMENT CORPORATION New Jersey Office: Texas Office: 55 Oak Street 7125 North Loop East Norwood, NJ 07648...
Need help?
Do you have a question about the 9610 and is the answer not in the manual?
Questions and answers