Maintenance And Testing Log; Service; Warranty Service - Smiths Level 1 H-1000 Operator's And Service Manual

Fluid warmer
Table of Contents

Advertisement

EVERY 12 MONTHS:
1. REPLACE O-RINGS
Remove the O-Rings by pulling them out of the bottom and top sockets with a pair of
needle nose pliers or by prying them out with a small screwdriver. Lubricate the new O-
Rings and then press them into place.
2. CLEAN FAN FILTER
Remove the fan filter with a screwdriver from the bottom of the unit and clean with warm
soapy water.

Maintenance and Testing Log

All maintenance and testing should be done by qualified personnel.
Every 30 days:
Change Distilled water
Lubricate O-Rings
Test alarm switch
Every 12 months:
Replace O-Rings
Clean Fan Filter
Test Add Water Alarm
Test Disposable Alarm
Test Over Temp Alarm
Performance Test/
Verify Temperature
Calibration
Earth Leakage
Ground Continuity
Dialectric Strength

Service

ALL SERVICE MUST BE PERFORMED BY SMITHS MEDICAL OR ITS AUTHORIZED
AGENTS. SERVICE BY OTHERS VOIDS THE WARRANTY AND TRANSFERS
LIABILITY FOR MALFUNCTIONS OF THE DEVICE TO THE SERVICING
ORGANIZATION.

Warranty Service

Units received for repair which have not been obviously abused or impact damaged and
are still under warranty will be promptly repaired and returned at no charge. Contact your
22

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

H-1000 100v

Table of Contents