Download Print this page

ETS 5503 Operating Manual

Compact environmental chamber

Advertisement

Quick Links

Compact Environmental Chamber
ETS Model 5503
Operating Manual
Electro-Tech Systems – 700 West Park Avenue, Perkasie, Pennsylvania 18944
D01307 Revision B

Advertisement

loading
Need help?

Need help?

Do you have a question about the 5503 and is the answer not in the manual?

Questions and answers

Summary of Contents for ETS 5503

  • Page 1 Compact Environmental Chamber ETS Model 5503 Operating Manual Electro-Tech Systems – 700 West Park Avenue, Perkasie, Pennsylvania 18944 D01307 Revision B...
  • Page 2: Table Of Contents

    Products described in this manual are designed and assembled in the U.S.A. by Electro-Tech Systems, Inc. 700 West Park Avenue Perkasie, PA 19844 www.electrotechsystems.com 833-ENV-GURU (833-368-4878) Model 5503 Compact Chamber – Operating Manual D01307 Revision B - 2021-03-16 - Page 2 of 20...
  • Page 3: Important Safety Information

    DO NOT USE IN ANY MANNER NOT SPECIFIED OR APPROVED BY THE MANUFACTURER: Unapproved use may result in damage to the equipment or present an electrical shock or fire hazard. Model 5503 Compact Chamber – Operating Manual D01307 Revision B - 2021-03-16 - Page 3 of 20...
  • Page 4: Description Of Contents

    II. Description of Contents The Models 5503 Chamber: Item Description The Model 5503 is a clear acrylic chamber, dimensions 24”W x 15”D x 18”H, and includes a 12” x 12” access door with a ¼-turn latch and non- Chamber setting gasket.
  • Page 5 Model 5482 Humidifier Model 5461 Desiccant Dehumidifier Model 5473 Thermoelectric Cooler Model 5478 Regenerating Dehumidifier Model 5503 with Ports & Gloves Model 5503 Compact Chamber – Operating Manual D01307 Revision B - 2021-03-16 - Page 5 of 20...
  • Page 6: Setup Guide

    Connections for humidity and gas are on the left side of the chamber to the left of the door. Place component systems near the connections. Model 5503 Compact Chamber – Operating Manual D01307 Revision B - 2021-03-16 - Page 6 of 20...
  • Page 7 • Position the sensor as shown and tighten the fitting nut by hand only - Do not use tools Model 5503 Compact Chamber – Operating Manual D01307 Revision B - 2021-03-16 - Page 7 of 20...
  • Page 8 HEAT on the Controller. • Route this cord to the controller temperature section and plug it into the receptacle labeled HEAT. Model 5503 Compact Chamber – Operating Manual D01307 Revision B - 2021-03-16 - Page 8 of 20...
  • Page 9 • LED will turn GREEN and mist will begin. • After 3 seconds, mist output will increase to maximum level. • Disconnect the AC cord. Model 5503 Compact Chamber – Operating Manual D01307 Revision B - 2021-03-16 - Page 9 of 20...
  • Page 10 Place the desiccator column, lid down, onto the 4 mounting studs of the base. The side marked “IN” should face forward. Marking is on the bottom of the column. Model 5503 Compact Chamber – Operating Manual D01307 Revision B - 2021-03-16 - Page 10 of 20...
  • Page 11 Output fitting to the Dry Air In fitting on the Chamber. • Check that all tubing is pushed firmly into the fittings for a good seal. Model 5503 Compact Chamber – Operating Manual D01307 Revision B - 2021-03-16 - Page 11 of 20...
  • Page 12 Switch the front panel Dehumidify switch to ON. The controller will request dehumidification as needed. Consult the Operating Manual for the 5461 Dehumidification System for more details. Model 5503 Compact Chamber – Operating Manual D01307 Revision B - 2021-03-16 - Page 12 of 20...
  • Page 13: Quick Start Guide

    (I = on) except those operating systems not needed to achieve the setpoint (e.g. no cooling is needed if a high temperature is selected). Model 5503 Compact Chamber – Operating Manual D01307 Revision B - 2021-03-16 - Page 13 of 20...
  • Page 14: Functionality

    The humidity controller includes a * button along with down and up arrows (▲//▼) to allow selecting and CONTROLLER Buttons changing the setpoint and other settings. Model 5503 Compact Chamber – Operating Manual D01307 Revision B - 2021-03-16 - Page 14 of 20...
  • Page 15 Provides power to humidification system when needed. (RH INCREASE) FUSE Fuse for internal controller electronics, .375A COMM PORT DB9 connector for communications to a computer Model 5503 Compact Chamber – Operating Manual D01307 Revision B - 2021-03-16 - Page 15 of 20...
  • Page 16: System Performance

    ±3.0% RH at 68° F (20° C) and 90-100% RH Sensor Accuracy RH Sensor Accuracy, Temp Temp: ±0.4° F (±0.2° C) at 20° C Model 5503 Compact Chamber – Operating Manual D01307 Revision B - 2021-03-16 - Page 16 of 20...
  • Page 17 Specify 115 VAC or 230 VAC Power (For full heat/cool/humidify/dehumidify system) 115 VAC @ 8 Amps, or 230 VAC @ 4 Amps, 50 / 60Hz, Single Phase Model 5503 Compact Chamber – Operating Manual D01307 Revision B - 2021-03-16 - Page 17 of 20...
  • Page 18: Maintenance And Calibration

    ETS recommends the Model 5503 system be calibrated annually. Only the controller and sensor need to be returned for calibration. Do NOT return the chamber. To return equipment to ETS for calibration or repair it is first necessary to obtain a RMA number, please call 215-887-2196 or email service@ets2.com...
  • Page 19 Guide” to return the humidifier back into operation again. Repair To return equipment to ETS for repair it is first necessary to obtain a RMA number, please call 215-887-2196 or email service@ets2.com Model 5503 Compact Chamber – Operating Manual D01307 Revision B - 2021-03-16 - Page 19 of 20...
  • Page 20: Warranty

    (such as gloves, iris ports and desiccants), sixty (60) days from date of shipment and (ii) for all standard equipment and otherwise nonrenewable items, one year from date of shipment. Model 5503 Compact Chamber – Operating Manual D01307 Revision B - 2021-03-16 - Page 20 of 20...