AGT PSG BCR01 Installation Manual, Operating Manual, Maintenance Manual, Spare Part List page 8

Sample gas cooler
Table of Contents

Advertisement

Installation-, Operating-, Maintenance-Manual, Spare Part List
Assembly and Installation
Feed Lines
All operations mentioned below to be performed by qualified personnel only.
As a standard, medium connections of the sample gas cooler have been designed for a 6/4mm hose. The
sample gas cooler is equipped with exactly one heat exchanger (PVDF, glass or SS316) and provides one
gaspath. The following feed lines must be connected:
Sample gas pipeline to the gas inlet connection
Analyzer to the gas outlet connection
Condensate pump/tank to the condensate outlet connection
Marking
Gas inlet
Gas inlet
The feed lines of the numerous possible heat-exchanger types (material PVDF/glass/SS316,
condensate connection top/bottom- see chapter "Technical Data") may have different designs.
Coolers without condensate pump are equipped with heat-exchangers with condensate outlet at
bottom (pipe, 12mm outer diameter, without thread).
The picture above shows a PVDF dual-HE with condensate connection on the top that is
connected with a condensate pump.
AGT-PSG GmbH & Co.KG
Errors and printing mistakes excepted
Marking
Gas outlet
Condensate outlet
Gas outlet
Translation - Manual BCR01 Rev. 6/2017
BCR01
Page 8