Temperature-Monitoring and Error-Diagnosis
Schaltplan
Circle-1 – Sensor-F1 – alarm: Monitoring and notification of temperature alarms.
Circle-2 – Sensor-F1 – power regulation: Refrigerant circuit.
Error-Diagnosis / Checklist
Fault
Failure of sample gas cooler
Sample gas flow reduced or
blocked
Condensate in gas outlet
Temperature > +10°C
Temperature < 0°C
AGT-PSG GmbH & Co.KG
Errors and printing mistakes excepted
Cause
Interrupted power supply
Sample gas cooler switched off
(optional equipment)
Defective fuse
(optional equipment)
Defective refrigerant circuit
Gas paths contaminated
Gas paths frozen
Condensate drainage leaking,
not or incorrectly connected
Gas inlet/outlet connected incorrectly
Condensate outlet blocked
Sample gas cooler overloaded
Defective refrigerant circuit
Too low ambient temperature
Defective performance regulation
Remedy
Re-establish the power supply
Check power supply,
switch on the sample gas cooler
Check power supply,
replace fuse
Contact AGT Service
Contamination may occur due to dust or
sublimate. Ensure pre-separation
Adhere to operation data
Check installation and condition of
condensate pumps and tanks
Connect gas inlet/outlet correctly
Clear condensate outlet,
check installation
Check operating conditions,
ensure proper cooling air circulation,
condensor cleaning
Contact AGT Service
Adhere to operation data
Contact AGT Service