INDEX PAGE 3 Index ..............Getting started ............ 4 - 5 Connections ............Controls .............. 6 - 7 Status icons ............PurgeNet™ interface ..........Care and maintenance ......... Warning .............. Other languages ..........Customer support and worldwide care ....Warranty ............. Address and contact details ........
gETTINg STarTED PAGE 4 Your PurgEye 500 Desk Weld Purge Monitor is an instrument ® ® to read from 0.1% (1000 ppm) down to 0.0001% (1 ppm) oxygen. Your PurgEye 500 Desk Weld Purge Monitor delivery and ® ® storage box contains: •...
coNNEcTIoNS PAGE 6 PurgeNet™ The PurgEye 500 Desk Weld Purge Monitor is fitted with a ® ® 3-pin PurgNet™ connector at the rear as indicated above. This is used for connection to a PC via USB to utilise the PurgeLog™ software, charging and ability to attach other accessories.
Page 7
coNTrolS PAGE 7 Display Holding the toggle button will change the display units from either ppm or %, pressing the display button will cycle the display through the various menu options. By default these are as follows: current oxygen sensor reading. NoTE: Dashes indicate an over-range reading or the unit is warming up.
Page 8
coNTrolS continued... PAGE 8 Adjust + Pressing this button will increase the currently displayed set-point by 10 ppm. Holding this button will cause the set-point to increase in 100 ppm steps. The set-point will not increase beyond the maximum range of the sensor. Adjust - Pressing this button will decrease the currently displayed set-point by 10 ppm.
STaTUS IcoNS PAGE 9 When necessary, the following three status icons will be displayed on the right hand side of the display. This will be either ‘PPM’ or ‘%’ for oxygen concentration values. Press and hold ‘display’ button to change the display units. Pump When the sample pump is enabled, you will see an arrow icon scrolling to the right, if the PurgEye...
PUrgENET™ INTErfacE PAGE 10 Smart accessories: When PurgeNet™ accessories are connected to the PurgeNet™ interface connection, they will be automatically detected by your PurgEye 500 Desk and extra menu options will appear. ® Language Support current support: The PurgEye 500 Desk product range currently ®...
carE and MaINTENaNcE PAGE 11 Zirconium Oxide Sensor Care As a general guideline, the gases to be measured should be free from dust, organic vapours, condensation and products capable of condensing. Such components can block the gas conduits in the sensor and cause permanent damage.
MUST ALWAYS be used, or your PurgEye may read incorrectly ® and/ or be permanently damaged. Spare filters available from your HFT Distributor. ® Internal Pump The internal pump should not be force fed with any pressure as it may be damaged.
Other laNgUagES in ProgrESS PAGE 13 Les autres langues sont actuellement dans le progrès. Pour le moment, s’il vous plaît contacter votre Distributeur local. Andere Sprachen sind momentan im Gange. Kontaktieren Sie vorläufig, bitte Ihren örtlichen Verteiler. Otros idiomas son actualmente en progreso. Para ahora, contacta por favor su Distribuidor local.
For further information and support, please contact us at: Internet: www.huntingdonfusion.com Email: hft@huntingdonfusion.com Tel: +44 (0) 1554 836 836 Worldwide Offices, Partners and Distributors are listed on our website or contact us and we will direct you to the correct location.
Should you encounter a problem with your product, please notify us immediately upon receipt. Huntingdon Fusion Techniques HFT warrants this product to be ® free of defects in workmanship and material, with exceptions stated below.
Need help?
Do you have a question about the PurgEye 500 Desk WELD PURGE MONITOR and is the answer not in the manual?
Questions and answers