Gamry Interface 5000 Quick Start Manual

Gamry Interface 5000 Quick Start Manual

Usb potentiostat
Hide thumbs Also See for Interface 5000:

Advertisement

Quick Links

Interface 5000
QUICK-START GUIDE
Guide #2

USB Potentiostat Calibration

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Interface 5000 and is the answer not in the manual?

Questions and answers

Summary of Contents for Gamry Interface 5000

  • Page 1 Interface 5000 QUICK-START GUIDE Guide #2 USB Potentiostat Calibration...
  • Page 2 Shield’s grounding post. Open Gamry Framework™. Select Experiment > Utilities > Calibrate Instrument Choose your potentiostat. Select the radio button for calibration type Both, and click the OK button. For newer Interface™ potentiostats, the AC radio button is inactive.
  • Page 3 Calibration is used to check the potentiostat’s health, and to “zero” many of the measurement circuits to your laboratory environment. A warning does not necessarily indicate a critical failure, and Gamry can use calibration information to determine the source of the warning.
  • Page 4 Gamry eQCM™. Contains a full suite of physical electrochemistry techniques. Electrochemistry Toolkit A sophisticated package for complete access to the capabilities of Gamry potentiostats in the software environment of your choice. Calibration QSG: USB Potentiostat 5000 - 988-00045 - rev. 1.11 - Gamry Instruments © 2018...