Built-In Calibration Sensor And Selfcalibration Problems - Eizo CG247 User Manual

Caibration color lcd monitor
Hide thumbs Also See for CG247:
Table of Contents

Advertisement

7-4. Built-in calibration sensor and SelfCalibration
problems
Problem
1. The built-in calibration sensor does not
come out/stays out.
2. SelfCalibration cannot be executed.
3. SelfCalibration failure
Error Code Table
Error Code
0014
0034
0050
0061
• Execute SelfCalibration again.
• Adjust using ColorNavigator.
• The built-in calibration sensor may not have come out during calibration, or light may
have entered the sensor during measurement.
• Turn off the main power, wait for a few minutes before turning the power back on, and
then execute SelfCalibration again.
• Execute SelfCalibration again.
• The built-in calibration sensor may not have worked correctly.
• Check whether there is any foreign object near the built-in calibration sensor.
• Execute SelfCalibration again.
Possible cause and remedy
• Turn off the main power, and then turn it on again a few
minutes later.
• Check whether the date and time are set correctly (see
2. Setting Monitor Date and Time" (page
• Check whether the mode is set (see
Calibration Schedule" (page
• Check whether the calibration targets are set correctly (see
"4-4. Setting the calibration targets" (page
• Set the SelfCalibration target using ColorNavigator.
• Refer to the error code table. If an error code that does
not appear in the error code table is displayed, contact
your local EIZO representative.
Description
36)).
"4-3. Setting
36)).
38)).
Chapter 7 Troubleshooting
"4-
49

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CG247 and is the answer not in the manual?

Table of Contents