INSTRUCTIONS SIGMOIDOVIDEOSCOPE OLYMPUS OSF TYPE V60 Refer to the endoscope’s companion manual, the “OLYMPUS OSF TYPE V60 ENDOSCOPE REPROCESSING MANUAL” for reprocessing information. USA: CAUTION: Federal law restricts this device to sale by or on the order of a physician.
Attaching Accessories to the Endoscope........Preparation, Inspection and Connection of Ancillary Equipment ..Inspection of the Endoscopic System ..........Chapter 4 Operation ..............Insertion ..................Using Endo-Therapy Accessories........... Withdrawal of the Endoscope ............Transportation of the Endoscope............ OSF TYPE V60 OPERATION MANUAL...
Page 4
Contents Chapter 5 Troubleshooting ............Troubleshooting Guide..............Returning the Endoscope for Repair..........Appendix..................System Chart ................... OSF TYPE V60 OPERATION MANUAL...
The meaning(s) of the symbol(s) shown on the package with the components, the back cover of this instruction manual and/or this instrument are as follows: Refer to instructions. Endoscope. Type BF applied part. Manufacturer Authorized representative in the European Community OSF TYPE V60 OPERATION MANUAL...
Important Information — Please Read Before Use Intended Use This instrument has been designed to be used with an Olympus Video System Center, Documentation Equipment, Display Monitor, Endo-Therapy Accessories (such as a Forceps) and other ancillary equipment for endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum and sigmoid colon).
Note that the complete instruction manual set for this endoscope consists of this manual and the “OLYMPUS OSF TYPE V60 ENDOSCOPE REPROCESSING MANUAL” which also accompanied the endoscope at shipment.
Some problems that appear to be malfunctions may be correctable by referring to Chapter 5, “Troubleshooting”. If the problem cannot be resolved using the information in Chapter 5, contact Olympus. Signal Words The following signal words are used throughout this manual: Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Page 9
Bending Section may stretch or break and cause water leaks. • Turn the Video System Center power switch OFF when attaching or detaching the Electrical Connector. Failure to do this can result in equipment damage, including destruction of the CCD. OSF TYPE V60 OPERATION MANUAL...
Page 10
Patient injury may be caused by; inserting or withdrawing the endoscope, feeding air or applying suction without a clear endoscopic image; withdrawing the endoscope with the angulation controls locked; forcefully pulling, twisting or rotating the angulated Bending Section. OSF TYPE V60 OPERATION MANUAL...
Match all items in the package with the components shown below. Inspect each item for damage. If the instrument is damaged, a component is missing or you have any questions, do not use the instrument; immediately contact Olympus. This instrument was not disinfected or sterilized before shipment. Before using this instrument for the first time, reprocess it according to the instructions given in the endoscope’s companion manual, “OLYMPUS OSF TYPE V60...
Water Resistant Cap Holder Water Resistant Cap (MAJ-639) Symbol Mark ( Symbol Mark ( 12. Water Container Connector Air Pipe Light Guide 11. Endoscope Connector 10. Suction Connector 14. S-Cord Connector Mount Chain Top View OSF TYPE V60 OPERATION MANUAL...
Page 15
Control Knob Suction Cylinder 3. RIGHT/LEFT Angulation Lock Air/Water Cylinder Grip Section 6. Color Code (Yellow) Biopsy Valve (MB-358) <Non-Sterile> 8. Instrument Channel Port Boot Working Length Insertion Tube Distal End 9. Bending Section Top View OSF TYPE V60 OPERATION MANUAL...
Channel; fluid feed channel (from a syringe via the Biopsy Valve). 9. Bending Section Moves the Distal End of the endoscope when the UP/DOWN and RIGHT/LEFT Angulation Control Knobs are operated. 10. Suction Connector Connects to the Suction Tube. OSF TYPE V60 OPERATION MANUAL...
Page 17
Connect to the Video System Center. 14. S-Cord Connector Mount Connects to the S-cord to conduct leakage current from the endoscope to the Olympus Electrosurgical Unit. To connect the S-cord, refer to the instruction manual for the Olympus Electrosurgical Unit. OSF TYPE V60 OPERATION MANUAL...
(10.2 – 15.4 psia) Transportation and Ambient temperature –47 to 70°C (–52.6 to 158°F) storage Relative humidity 10 – 95% environment Atmospheric pressure 700 – 1060 hPa (0.7 – 1.1 kgf/cm (10.2 – 15.4 psia) OSF TYPE V60 OPERATION MANUAL...
Page 19
Direction from which Endo-Therapy Accessories enter and exit the endoscopic image Air Flow Rate 25 cm ° Bending Section Angulation Range UP 180 ° DOWN 180 ° RIGHT 160 ° LEFT 160 Total Length 1070 mm OSF TYPE V60 OPERATION MANUAL...
Page 20
Chapter 2 Instrument Nomenclature and Specifications Year of manufacture 2912345 The last digit of the year of manufacture is given in the second digit of the serial number. Degree of protection TYPE BF applied part against electric shock OSF TYPE V60 OPERATION MANUAL...
This instrument was not disinfected or sterilized before shipment. Before using this instrument for the first time, reprocess it according to the instructions in the endoscope’s companion manual, the “OLYMPUS OSF TYPE V60 ENDOSCOPE REPROCESSING MANUAL”. Preparation of the Equipment Refer to the “System Chart”...
Figure 3.1 Preparation and Inspection of the Endoscope Clean and disinfect or sterilize the endoscope as described in the endoscope’s companion manual, the “OLYMPUS OSF TYPE V60 ENDOSCOPE REPROCESSING MANUAL”. Inspection of the Endoscope Inspect the control section and the endoscope connector for excessive scratching, deformation, loose parts or other irregularities.
Page 23
Inspect the objective lens at the Distal End of the endoscope’s Insertion Tube for scratching, cracks, stains, gaps around the lens or other irregularities. Inspect the air/water nozzle at the Distal End of the endoscope’s Insertion Tube for abnormal swelling, bulges, dents or other irregularities. OSF TYPE V60 OPERATION MANUAL...
Page 24
Confirm that the angle of the Bending Section is roughly stabilized when the RIGHT/LEFT Angulation Control Knob is released. Confirm that the Bending Section straightens out when the RIGHT/LEFT Angulation Lock is turned in the “F ” direction and the RIGHT/LEFT Angulation Control Knob is released. OSF TYPE V60 OPERATION MANUAL...
Preparation and Inspection of Accessories Clean and disinfect or sterilize the Air/Water Valve (MB-196), Suction Valve (MB-197) and Biopsy Valve (MB-358) as described in the endoscope’s companion manual, “OLYMPUS OSF TYPE V60 ENDOSCOPE REPROCESSING MANUAL”. Inspection of the Air/Water and Suction Valves Visually inspect the Air/Water and Suction Valves (see Figures 3.4 and 3.5).
Page 26
(see Figure 3.6). Normal Abnormal Slit Discoloration Hole Main body Splits, Cracks Figure 3.6 Attach the cap to the main body as shown in Figure 3.7. Slit Main body Figure 3.7 OSF TYPE V60 OPERATION MANUAL...
Skirt. Skirt Side View Moving Surface Suction Cylinder Gutter Bottom View Metal Pin Top View Figure 3.8 The Suction Valve will make a whistling noise when it is dry; this does not indicate a malfunction. OSF TYPE V60 OPERATION MANUAL...
Page 28
Attaching the Biopsy Valve Attach the Biopsy Valve to the Instrument Channel Port of the endoscope (see Figure 3.10). Confirm that the Cap fits properly without bulging. Biopsy Valve Instrument Channel Port Assembled Figure 3.10 OSF TYPE V60 OPERATION MANUAL...
• Do not attach or remove the Endoscope Connector with your finger on the Water Resistant Cap Holder (as shown in Figure 3.11), to avoid damaging it. Figure 3.11 OSF TYPE V60 OPERATION MANUAL...
Page 30
Connector Assembled Water Container Connector Water Container’s Water Container Water Supply Tube Connection Adapter Figure 3.13 Connect the Suction Tube from the Suction Pump to the Suction Connector on the Endoscope Connector (see Figure 3.14). OSF TYPE V60 OPERATION MANUAL...
Page 31
To the Suction Connector on the endoscope Figure 3.14 Securing the Water Resistant Cap (MAJ-639) Before operating the endoscope, secure the Water Resistant Cap to the Endoscope Connector (see Figure 3.15). Water Resistant Cap (MAJ-639) Figure 3.15 OSF TYPE V60 OPERATION MANUAL...
With the Distal End immersed to a depth of 10 cm, cover the Hole in the Air/Water Valve with your finger and confirm that air bubbles are continuously emitted from the Air/Water Nozzle. OSF TYPE V60 OPERATION MANUAL...
Page 33
Suction Valve is not working properly, it may be impossible to stop suction, which could cause patient injury. If the reattached or replaced Suction Valve fails to operate smoothly, the endoscope may be malfunctioning; stop using it and contact Olympus. OSF TYPE V60 OPERATION MANUAL...
Page 34
Keep your eyes away from the Distal End when inserting Endo-Therapy Accessories. Extending the Endo-Therapy Accessory from the Distal End could cause eye injury. Insert the Endo-Therapy Accessory through the Biopsy Valve. Confirm that the Endo-Therapy Accessory extends smoothly from the Distal End. OSF TYPE V60 OPERATION MANUAL...
If the endoscope or Endo-Therapy Accessory cannot be withdrawn from the patient smoothly, do not attempt to forcibly withdraw it. If any irregularities are suspected, immediately contact Olympus. Forcibly withdrawing the endoscope or Endo-Therapy Accessory may cause patient injury, bleeding and/or perforation.
The Air/Water and Suction Valves can be operated using the left index finger. The UP/DOWN Angulation Control Knob can be operated using the left thumb. The right hand is free to manipulate the Insertion Tube and the RIGHT/LEFT Angulation Control Knob (see Figure 4.1). Figure 4.1 OSF TYPE V60 OPERATION MANUAL...
Page 37
If the endoscope cannot be withdrawn from the patient smoothly, do not attempt to forcibly withdraw it; leave it inside the patient and immediately contact Olympus. Forcibly withdrawing the endoscope may cause patient injury. OSF TYPE V60 OPERATION MANUAL...
Page 38
In the case, turn the airflow regulator switch on the Video System Center to OFF and add more sterile water to the Water Container. Otherwise, patient injury could occur. OSF TYPE V60 OPERATION MANUAL...
Page 39
During the procedure, make sure that the Suction Bottle does not fill completely or overflow. Aspirating fluids into a full bottle can damage the Suction Pump. OSF TYPE V60 OPERATION MANUAL...
The minimum visible distance depends on the type of endoscope being used. Refer to Section 2.3, “Specifications”. OSF TYPE V60 OPERATION MANUAL...
Page 41
When the cap of the Biopsy Valve is detached, it may cause patient debris or fluids to leak or spray from the endoscope, posing an infection control risk. When the valve is uncapped, place a piece of sterile gauze over it to prevent leakage. OSF TYPE V60 OPERATION MANUAL...
Page 42
If significant resistance is encountered and insertion is very difficult, straighten the Bending Section as much as possible without losing the endoscopic image. Inserting Endo-Therapy Accessories with excessive force may damage the endoscope and/or cause patient injury. OSF TYPE V60 OPERATION MANUAL...
Page 43
If the Endo-Therapy Accessory cannot be withdrawn from the endoscope, carefully withdraw both the endoscope and the Endo-Therapy Accessory together under endoscopic observation. Take care not to cause tissue trauma. OSF TYPE V60 OPERATION MANUAL...
Page 44
Prepare, inspect and connect the Electrosurgical Unit and Electrosurgical Accessories as described in their instruction manuals. The application of high frequency current may interfere with the endoscopic image. This is normal and does not indicate a malfunction. OSF TYPE V60 OPERATION MANUAL...
Light Source because it is extremely hot (see Figure 4.5). Do not lift the Endoscope Connector by the Water Resistant Cap, to avoid damage to the Chain and the Water Resistant Cap Holder (see Figure 4.4). Figure 4.4 OSF TYPE V60 OPERATION MANUAL...
Page 46
Carrying Case. Disinfect or sterilize the endoscope again before use. • Do not attach the Water Resistant Cap when transporting the endoscope, to avoid damage to the endoscope caused by changes in the air pressure. OSF TYPE V60 OPERATION MANUAL...
Section 5.1, “Troubleshooting Guide”. Should you fail to correct the problem even after taking the described remedial action, stop using the endoscope and send it to Olympus for repair. Olympus does not repair accessory parts. If an accessory part becomes damaged, contact Olympus to purchase a replacement. •...
Page 48
Air/Water Valve is Replace it with a new Air/Water damaged. Valve. Air/Water Valve cannot Improper Air/Water Use a proper Air/Water Valve. be attached. Valve is used. Air/Water Valve is Replace it with a new Air/Water damaged. Valve. OSF TYPE V60 OPERATION MANUAL...
Page 49
Image is not clear. Dirty Objective Lens. Feed water to remove mucous, etc. Excessively dark or Improper Video System Adjust the Video System Center’s bright images. Center setting. setting as described in its instruction manual. OSF TYPE V60 OPERATION MANUAL...
Olympus. Olympus is not liable for any injury or damage which occurs as a result of repairs attempted by non-Olympus personnel. Before returning the endoscope for repair, contact Olympus. With the...
Page 51
Chapter 5 Troubleshooting OSF TYPE V60 OPERATION MANUAL...
New products released after the introduction of this instrument may also be compatible for use in combination with this instrument. For further details, contact Olympus. If combinations of equipment other than those shown below are used, the full responsibility is assumed by the medical treatment facility.
Page 53
Leakage Tester Channel Cleaning (MB-155) Adapter (MB-19) All-Channel Irrigator with 30 cm (30 ml) Syringe (CW-3) AW Channel Cleaning Channel Cleaning Adapter (MB-107) Brush (BW-20T) Channel-opening Maintenance Unit Ultrasonic Cleaner Cleaning Brush (MU-1) (KS-2) (MH-507) OSF TYPE V60 OPERATION MANUAL...
Page 54
FB-25KR-1 ROTATABLE CYTOLOGY BRUSH BIOPSY FORCEPS Rat Tooth with Fenestrated with Alligator Jaws Standard Type With Sheath Needle (Swing Type) Endoscope OSF-V60 FB-53KR-1 to FB-24KR-1 BC-2J BC-9L 55KR-1 GRASPING FORCEPS Alligator Jaws Endoscope OSF-V60 FG-6L-1 OSF TYPE V60 OPERATION MANUAL...
Page 55
FS-1K WASHING PIPE SURGICAL LOOP MAGNETIC SCISSORS CUTTER EXTRACTOR Standard Type Endoscope OSF-V60 FS-3L FS-5L-1 1E-1L PW-1H-1 WASHING PIPE MEASURE ROTATABLE CLIP LIGATING DEVICE FIXING DEVICE Spray Type Straight Endoscope OSF-V60 PW-5L-1 HX-5LR-1 HX-20L-1 M1-2K OSF TYPE V60 OPERATION MANUAL...
Page 56
Oval with Thorn Oval Mini Oval Thorn Endoscope OSF-V60 SD-16L-1 SD-17U-1 SD-210L-25 SD-210L-15 DISPOSABLE ELECTRO- COAGULATION ELECTRODE HOT BIOPSY SURGICAL SNARE FORCEPS Extra Mini Oval Ball Point Suction Type Endoscope OSF-V60 SD-210L-10 CD-1L/2L CD-3L FD-1L-1 OSF TYPE V60 OPERATION MANUAL...
Page 57
Appendix ELECTRO- DISPOSABLE HOT SURGICAL KNIFE BIOPSY FORCEPS Needle Type Endoscope OSF-V60 FD-5L KD-1L-1 OSF TYPE V60 OPERATION MANUAL...