Care Instructions

Cleaning and disinfecting are separate processes. Cleaning must precede disinfection.
Before use by a different individual: clean, disinfect, and check product for proper
functioning.
DO NOT:
- machine wash or dry the cushion
- immerse the cushion in water.
- autoclave, steam clean, or expose cushion to temperatures in excess of 150˚F (65˚C)
- use caustic, automatic dishwasher detergents
- use cleaning products containing petroleum or organic solvents such as acetone,
toluene, Methyl Ethyl Ketone (MEK), naphtha, dry cleaning fluids, or adhesive
removers
- expose the cushion to ozone gas or to ultraviolet light cleaning methods
Note: Before cleaning and disinfecting, remove the cover from the cushion. Remove hook
fastener pieces from the bottom of the cover.
To Clean the Cover: Machine wash in cold water (30°C) with mild detergent,
gentle cycle. Tumble dry low.
To Disinfect the Cover: Hand wash in warm water
(40°C) using 1 part household liquid bleach per 9
parts water, then rinse thoroughly. The cover can also
be machine washed in hot water (60°C) to disinfect.
Tumble dry low.
Follow safety guidelines on bleach container.
To Clean the Cushion: Remove the cover. Wipe gently with a cloth dampened with
water and a household detergent. Rinse cloth and re-wipe. Allow foam to air dry
thoroughly before putting the cover on.
Disinfectants are not effective on porous surfaces such as foam. If the foam
portion of the product becomes soiled or contaminated, DO NOT use with multiple
individuals. Replace soiled or contaminated foam components.