1. General The company Analytical Control Instruments (hereinafter the ‘manufacturer’) provides with the Stationary Photoionization Detector SPID2 (hereinafter the ‘device’) a continuously measuring detector for volatile organic compounds (VOCs) with a high stable hollow cathode lamp with Ceramic Discharge Channel with an ionization potential < 10.6 eV. The use of new technologies for the excitation source and the sensor allows a high stability of measurement and longer maintenance intervals.
Significant dust deposits on the gas inlets will increase the response time of the device. Checks for dust deposits must be done at regular intervals. SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 8 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
(including but not limited to negligence), by operation of law, or otherwise. SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 9 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
The device must be installed vertical with the filter port unit downward to avoid plugging of the gas inlets by particles or liquids. SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 11 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
7. Replace electronic unit and secure the captive screw. 8. Replace enclosure lid and secure the locking screw. SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 12 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
Page 13
1 2 3 The terminal plugs accepts AWG 26 (0.14 mm²) to AWG 16 (1.5 mm²) wire. SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 13 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
3. Screw the filter cartridge into the filter port unit. Only light force is required. 4. Open the filter cartridge after installation by using the opener. SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 14 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
3. Replace sample port – protection cap with gas intake hole downward. Calibrate the device after installation of a new dust filter. SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 15 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
3. Select the desired menu item. After a period of 120 seconds of inactivity, the navigation bar disappears automatically. SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 16 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
2. Enter the required password and confirm the process with “accept” 3. Confirm access to the instrument menu SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 17 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
300 – 500 mm. (see following connection scheme) For calibration, it is recommended to use the Calibration Kit from accessories. (see chapter 14 Accessories) SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 18 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
2. Select “run” to go to calibration screen and select START (Start new calibration) 3. Press “accept” to start new calibration and follow the instructions on the screen SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 19 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
Page 20
Remove SPAN Gas [Isobutene] [Isobutene] [User gas] [User Gas] Calibration Successful Calibration Successful Calibration Successful Calibration Successful SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 20 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
The device takes ambient air through the Filter Port • The current measurement value will be displayed for 2 minutes SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 21 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
Current Loop. 6.2.4 Current Loop • (4 ... 20 mA or 0 … 5 mA) OUTPUT and SINGAL SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 22 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
1. Go to the Instrument Menu and select MEAS CONFIG 2. Select “run” to start measurement configuration 3. Select Calibration, “change” calibration data setting and “accept” SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 23 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
4. Select Response and modify response setting Response Screen List Item Gas Name CAS Number Response Factor (Identification) SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 24 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
PURGE TIME ( including recovery time ( Tr ) ) Recovery time ( fixed to 10 seconds ) SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 28 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
Page 29
Adjustable between 10 seconds and 4 weeks The Automatic Purge mode is made by default but can be switched off. SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 29 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
Page 30
Adjustable between 1 min and 5 hours After purging the device independently returns again to the normal operation. SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 30 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
Default Settings: • Slave ID: • Baud rate: 115200 • Mode: • Termination: Enable • Parity: Even (fixed) SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 31 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
Page 32
Alarm LO level [ppm] 30112 Float 4 byte Alarm HI level [ppm] *see Status Information Bit Definition SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 32 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
1. Go to the Instrument Menu and select RESET. 2. Select “run” resets the configuration to factory defaults. Customized response factors that are generated via PC software SPID2 Control Center will be deleted after reset. SPID2 - User Manual (en) - 1.11.docx Version 1.11...
The remote control mode allows connecting the instrument to a PC via RS485 interface for advanced configuration. PC software SPID2 Control Center must be installed before you can connect the device to the PC. A special adaptor is required. How To – Activate Service Mode 1.
2 mA 0 ... 5 energized flash In case of purging • Status LED is flashing green SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 38 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
2 years after continuous 0002673 Pump every 2 years operation. 0002671 Sensor Block every 2 years SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 39 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
It is not allowed to unscrew the safety screws of the Filter Port Unit and Sample Port Unit. SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 40 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
Swing Out Electronic Unit 4. Pull on the top of the board to swing out the upper electronic unit. SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 41 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
Don´t remove the locking screw! This screw is glued in and may be very tight. Remove Electrical Connection 4. Unplug the connecting cable. SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 42 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
8. Replace the spring plate. Do not fasten the two red marked screws. 9. Calibrate the device after installation of a new PID lamp. SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 43 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
4. Switching on the device and after 15 min the calibration procedure can be started. Do not touch the lamp window during installation. SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 44 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
For detailed information, see chapter 4.3. Calibrate the device after installation of a new filter cartridge. SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 45 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
2. Replace the dust filter by a new one. 3. Replace the protection cap. Calibrate the device after installation of a new dust filter. SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 46 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
Ingress Protection IP64 *The indicated values were obtained under standardized conditions. Test gas was isobutene in synthetic air. SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 47 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
-40 °C ≤ Ta ≤ +60 °C Type key: SPID2 - R* - D* - L* - S* - C* Position of thread hole for cable gland: 0 – left; 1 – right 0 – without alarm relays; 1 – with alarm relays 0 –...
Page 49
M20 x 1,5; Torque 8 – 12 Nm EMC Conformance according to the Directive 2014/30/EU EN 50270:2006 Type 2, EN 61000-6-3:2007 SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 49 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
-40 °C ≤ Ta ≤ +60 °C Type key: SPID2 - R* - D* - L* - S* - C* Position of thread hole for cable gland: 0 – left; 1 – right 0 – without alarm relays; 1 – with alarm relays 0 –...
Page 51
The housing of the detector has to be equipped with flameproof cable gland Ex d IIC certified according to Directive 2014/34/EU (ATEX). • M20 x 1,5; Torque 8 – 12 Nm SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 51 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
If the error code is still shown after a restart and troubleshooting, the device could be defective. For additional support, please contact the manufacturer. SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 52 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
Flow Controller with pressure indicator 0002735 Calibration Gas 100 ppm Isobutene in air 34 l 0002770 SPID2 USB-RS485 communication kit SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 57 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
• 1 x User manual • 1 x Declaration of conformity • 1 x Certificate of inspection SPID2 - User Manual (en) - 1.11.docx Version 1.11 Page 58 of 60 Analytical Control Instruments GmbH, Volmerstraße 9A, D-12489 BERLIN, Germany, www.aci-berlin.de...
Page 59