Page 2
USER’s MANUAL - March 2010 MTS-2000 METER TEST SYSTEM Pay special attention to the warnings and safety instructions that accompany the above symbol wherever it is found within this manual! CU R001 01B MANTA TEST SYSTEMS...
MTS-2000 User’s Manual All rights reserved by Manta Test Systems Inc. No part of this publication may be reproduced or distributed in any form or by any means without the permission of Manta Test Systems Inc. The information and specifications contained within from Manta Test Systems are believed to be accurate and reliable at the time of printing.
Testing a Voltage Transducer with Channel 1 Testing a Voltage Transducer with Channel 2 Testing a Watt Transducer Testing a Voltage Device (0 – 300V) 9.10 Outputting Transducer Current (0-20mA) 9.11 Outputting DC Current (0 – 2500 mA DC) CU R001 01B MANTA TEST SYSTEMS...
PRODUCT OVERVIEW The MTS-2000 is a meter test system designed to test panel meters, transducers and similar metering hardware. All applicable quantities (voltage, current, power, frequency, phase, etc) are displayed as required. The MTS-2000 is designed to be easily operated from the controls on the front panel.
The user must be qualified to work safely in the intended application environment of this instrument. Non-adherence to the following minimum requirements constitutes misuse of the MTS-2000, and the manufacturer accepts no liability for damages arising from such misuse: The instrument case must always be effectively grounded. The integrity of the power supply cord ground should always be verified before use.
Provided that Manta Test Systems receives notice of any defects in materials or workmanship of its hardware products, or hardware components of its products, within a one-year period, Manta shall, at its option, either repair or replace the defective hardware product or hardware component, if proven to be defective.
Extensi on of Warrant y At the discretion of Manta Test Systems, the warranty may be extended for a product which has been returned for service shortly after its warranty period has expired.
In voltage mode it generates up to 150V (AC or DC). In current mode it generates up to 5Aac or up to ±2.5Adc. Power suppl y The mains input is auto-ranging, power factor corrected and can operate from either 105-130 or 208-250Vac (50-60Hz). MANTA TEST SYSTEMS CU R001 01B...
Power DC: 15W dc max per channel, 30W dc with 2 channels in series Noise & distortion at maximum power: <1% Bandwidth: 3kHz (-3dB) Protection: Overload, short circuit, over-temperature, transient overvoltage, clip indication Surge Current (DC mode): 1A maximum (limited) Connector: 4mm Banana CU R001 01B MANTA TEST SYSTEMS...
AC Outputs Frequency/Phase Angle Frequency range: DC, 10-1000 Hz Frequency resolution: 0.001Hz Frequency accuracy: 50ppm Capable of generating 2 simultaneous frequencies Phase angle range: 0-359.9°, resolution 0.1° Phase accuracy: 0.1° (43-65Hz), 0.5° (10-43Hz and 65-120Hz), 1.0° (120Hz-1000Hz) MANTA TEST SYSTEMS CU R001 01B...
14.9”W x 6.9”H x 12.1”D (37.9cm W x 17.5cm H x 30.8cm D) Weight: 15 lbs (6.8 kg) Display: 20 character by 4 line VFD Operating temperature: 32° to 122°F (0° to 50°C) Storage temperature: -13° to 158°F (-25° to 70°C) CU R001 01B MANTA TEST SYSTEMS...
8 function keys, positioned to the left and right of each line on the display device MENU key ENTER key CE (clear entry) key 0-9 digit keys decimal point key +/- key OUTPUTS ON key OUTPUTS OFF key LEDs: OUTPUTS ENABLED VOLTAGE MODE (2nd Channel) MANTA TEST SYSTEMS CU R001 01B...
Page 14
Whenever the amplifier outputs are on, the ENABLED LED is on. Some tests configure channel 2 to output current, others configure it to outpt voltage. The VOLTAGE MODE LED is on when the convertible amplifier is configured to output voltage. CU R001 01B MANTA TEST SYSTEMS...
OPERATING MODES The MTS-2000 offers various operating modes that are designed to provide an easy to use interface for the particular category of testing to be performed. The basic modes are: Manual test Meter test Transducer test Setup Manual Test Mode The manual test mode is used for basic meter and transducer testing.
0%, 25%, 50%, 75% and 100% of full scale. Setup Mode The “Setup” selection in the main menu allows the user to configure the device, including: Calibrate the amplifiers Calibrate the current loop input and output CU R001 01B MANTA TEST SYSTEMS...
Testi ng a Vol tage Devi ce (0 – 300V) 9.10 Outputti ng Transducer Current (0-20mA) 9.11 Outputti ng DC Current (0 – 2500 mA DC) (note: Channel 2 must be set to DC) MANTA TEST SYSTEMS CU R001 01B...
Page 23
ISO 9001 Registered www.mantatest.com Manta Test Systems Inc 4060B Sladeview Crescent, #1, Mississauga, ON L5L 5Y5, Canada e-mail: sales@mantatest.com Phone: 905-828-6469 Toll Free: 800-233-8031 Fax: 905-828-6850...
Need help?
Do you have a question about the MTS-2000 and is the answer not in the manual?