ELZAB Care SP1 UV-C User Manual page 10

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 8
Replacement of the lower filter (
replacement every 4 - 6 weeks (depending on environmental pollution)
1
carefully place the device on its side, on a flat surface
(manually) remove 3 bolts fixing the filter cassette to the base
2
slide out the filter cassette
3
(sliding in the inclined position is necessary due to protruding elements)
4
slide out the filter
replace it with a new filter
5
make sure that the power cord is correctly connected and slide
6
in the filter cassette
7
(manually) insert the 3 fixing bolts
8
carefully put the device in the proper position
Fig. 2
6
Replacement of the upper filter (
replacement once a year
1
carefully place the device on its side, on a flat surface
remove the lower filter cassette
2
(points 2 - 3 in the lower filter replacement, Fig. 2)
remove the power cord
3
remove 4 screws of the upper cover
4
(use the Torx driver provided with the device)
5
carefully slide out the upper part of the internal unit
6
slide out the filter
l
care.elzab.p
Fig.
(check 2 guiding pins)
1
Fig.
2)
3
4
5
3)
EN