CAS EXI-200AD Manual

CAS EXI-200AD Manual

Explosion proof indicator

Advertisement

`
이미지 추가
2

Advertisement

Table of Contents
loading

Summary of Contents for CAS EXI-200AD

  • Page 1 이미지 추가...
  • Page 2: Table Of Contents

    Contents Cautions for Safety................... 6 1. Normal Requirements ................. 9 1.1 Application Range.......................9 1.2 Transport, assembly and installation.................9 1.3 Function, Operation and Use..................13 1.4 Maintenance and Cautions..................... 15 2. Technical Specifications ..............16 2.1 Product Specifications ..................... 16 2.2 Product Labelling ......................16 2.3 Product Structure and Composition ................
  • Page 3 6. Weight Calibration Mode..............35 6.1 What is weight calibration?....................35 6.2 How to enter the weight calibration mode..............35 6.3 Weight calibration menu (CAL1 – CAL9)...............36 7. Weighing Mode ...................45 7.1 Zero-set Function ......................45 7.2 Tare function........................46 7.3 Item number changing .....................46 7.4 Set Point changing ......................47 7.5 Subtotal Print........................47 7.6 Total Print...........................48...
  • Page 4 9.5 Hardware Set Function ....................69 9.6 Relay Setting ........................70 9.7 TCP IP ..........................76 9.8 Print Function Setting ...................... 79 10. Error Message...................84 10.1 Errors that can occur in weight calibration mode ............84 10.2 Errors that can occur in weighing mode ..............85 Appendix 1>...
  • Page 5: Cautions For Safety

    Cautions for Your Safety ‘Please comply with 'Cautions for Your Safety', which will lead you to use the product safely and properly to prevent any dangerous situations. ■ Cautions are divided into 'Warning' and 'Alert', which mean as follows. ■ Keep this manual in a place where product users can find out, after finish reading it. 'Warning' means a great possibility led to the death or heavy injury when instructions are violated.
  • Page 6 Warning Never disassemble, repair or retrofit the Ensure the power plug to be fully Ensure the grounding of the product. product. inserted to prevent shaking. Poor grounding might cause failure or It might exclude the product Any instable connection might cause electric shock upon electric leak..
  • Page 7 Attention Find a proper place to attach the rubber Check the weighing error anytime for Avoid any sudden shock to the product. pad at the bottom of the indicator, the accurate weighing. It might damage the product to fail the which was shipped together.
  • Page 8: Normal Requirements

    1. Normal Requirements 1.1 Application Range This product is an explosion-proof and dust ignition-proof product for locations of gas, steam or chemical use and other hazardous areas. ■ Explosion-proof Structure and Class Explosion proof Dust ignition proof Explosion- Ex d IIB +H T6 IP66 Ex tD A21 85℃...
  • Page 9 1.2.2. Assembly and Installation (1) Assembly This product is fully assembled when shipped and there is no need of separate assembly by the user. (2) Installation When installing electrical instruments in an explosion-proof area, compare the below requirements with the technical specifications beforehand to confirm correspondence. (a) Normal Requirements.
  • Page 10 (e) Wiring Method during Installation (power and load cell) ① Refer to "4. External Explanation" and "5. Internal Explanation" of this manual. ② Use an Allen wrench to loosen the front bolts and open the front cover ③ Connect an appropriate cable gland (joint screw diameter(M20*1.5)) to the power line. (Refer to 4.2 Front Surface Explanation) ④...
  • Page 11 < AC Type> < DC Type> ⑤ Connect an appropriate cable gland (joint screw diameter (M20*1.5)) to the load cell line. (Refer to 4.2 Front Surface Explanation) ⑥ Refer to this manual's "5.1.2 Load Cell Connector" and the wiring blueprint of the load cell for the wiring order.
  • Page 12: Function, Operation And Use

    1.3 Function, Operation and Use 1.3.1. Inspection and Adjustment (1) For the explosion-proof instrument, check for following suitability list (a) Terminal voltage, pole number (b) The type and specification of replacement parts from each part (c) Whether the overall control system (control, manipulation, display, notification etc) is working (d) Operation and display of auxiliary instruments (e) Presence of vibration and its intensity.
  • Page 13 < Basic Inspection Items > Item Method Inspection Content Notes Container cleaning, (explosion-proof By eye There must be no external damage (rust, dust) corrosion box) protection Display window By eye There must be no damage by scraping etc cleaning By eye, Screws There must be no loosening, dust or rust tighten, cleaning...
  • Page 14: Maintenance And Cautions

    1.4 Maintenance and Cautions 1.4.1. Maintenance (1) This product requires continued management through periodic inspection (1 time or more per month recommended). (2) When maintaining electrical instruments the following items must be checked. (a) There must be no damage to the container connection surface (screw). (b) There must be no damage or cracks from shock to the container exterior.
  • Page 15: Technical Specifications

    2. Technical Specifications 2.1 Product Specifications Indicator Product Name EXI-200AD Mode Name Explosion proof Ex d IIB +H T6 IP66 Explosion proof structure Dust ignition proof Ex tD A21 85℃ IP66 Aluminum (AL) Material container 100 V ~ 240 V , 50/60 Hz...
  • Page 16: Product Structure And Composition

    2.3 Product Structure and Composition 2.3.1. To form the body, an aluminum composite material suitable for the shaping and usage requirement of the product was selected. 2.3.2. The protective galss is tempered, explosion-proof structure formed by heat treatment and metallic molding to be capable of maintaining sufficient strength. 2.3.3.
  • Page 17: Features

    3. Features 3.1 Features ■ High speed, High accuracy ■ High speed micro processor adoption ■ A/D conversion speed : Maximum 200 times/sed ■ Appropriate for weight and measurement system ■ Easy operation and various options. ■ Simple and prompt Full Digital Calibration (SPAC : Single pass automatic span Calibration) ■...
  • Page 18: Product Specifications

    3.3 Product Specifications Power supply Unit Power supply voltage 10 ~ 24(DC), 100V~240V(AC) Max consumption 70 with 350 Ω Temperature range Storage temperature -25 ~ 85 ℃ Operating temperature -20 ~ 40 ℃ Loadcell Impedance(complete bridge) > 80 Ω Connection 4 or 6 wires Loadcell power supply 5 ±...
  • Page 19 Low pass, band-stop and self-adaptive digital filters Set points managements Checkweigher functioning mode Peak detection functioning mode Option Display 7 Segment LED, 4 key -10 V~10 V A-Out V-out, I-out 0~24 mA Relay I/O 2input, 4output 50,000ea Save Alibi memory Real time clock Ethernet 10/100 M...
  • Page 20: External Explanation

    4. External Explanation 4.1 External Specifications...
  • Page 21: Frontal Surface Explanation

    4.2 Frontal Surface Explanation 4.2.1 Detailed Explanation Name Function ① Display Displays the load weight, status, various messages etc. ② Key (touch key) Used for function and data input. Wiring inlet (3 openings total), 2 attachments standard, internal pressure packaging * Use a cable gland which satisfies the KCs explosion proof certification IP66..
  • Page 22 4.2.2 Wiring entrance and ground connection explanation ① Load cell signal line wiring inlet ② Power line wiring inlet ③ Wiring inlet for communication line or other optional lines ④ Ground wiring connector (select based on environment) * Ensure wiring is installed in the designated positions as it may affect the capability of the product.
  • Page 23: Display Explanantion

    4.3 Display Explanantion Main function (F5-05 OFF Set) Sub function (F5-05 ON Set) ○ Stable LED LED ON : output 4 Closed TARE Tare LED LED ON : output 3 Closed On = Net , Off = Gross LED ON : output 2 Closed Zero LED LED ON : output 1 Closed HOLD...
  • Page 24: Key Manual

    4.4 Key manual 4.4.1 EXI-200 KEY PAD 4.4.2 Function Key - Functions of keys used in Normal Mode. * Adjust the weight display near the zero-set to 0. (2 %, 5 %, 10 %, 20 %, 100 % ranges can be selected.) * For every press, alternate between total/net load and display the weight with the lamp indicator.
  • Page 25 * Use to change the product number. * Use to input the item code. * Use to change the Set Point. * Use when weighing a moving object. * Use when deleting all input. * Use to enter the menu mode. * Can be customized to suit the needs.
  • Page 26: Internal Explanation

    5. Internal Explanation 5.1 Internal Explanation. Explanation Notes ① RS-232 connector(COM1) Refer to 5.1.3 ② RS-232 connector (COM2) Refer to 5.1.3 ③ RS-485 connector Refer to 5.1.3 ④ RELAY connector Refer to 5.1.4 ⑤ A-OUT connector ⑥ Load cell connector Refer to 5.1.2 ⑦...
  • Page 27: Power Connector (Ac, Dc)

    5.1.1 Power Connector (AC, DC) - Before connecting always check the power requirements of the product. - For fine measurement, measure for 10 minutes after power is turned on. (1) AC Type...
  • Page 28 (2) DC Type...
  • Page 29: Load Cell Connector

    5.1.2 Load Cell Connector...
  • Page 30 Note 1. In case of 4 wires load cell, connect EX+ with SEN+, and connect EX- with SEN-. Note 2. Max 8 loadcells can connect to EXI-200(when loadcell impeadance 350 Ω ) * Relationship between the load cell output and input sensitivity. The input sensitivity of this product is maximum 0.3uV/digit or more.
  • Page 31: Rs-232, Rs-422/485 Connector

    5.1.3 RS-232, RS-485 Connector RS-485 RS-232(COM2) RS-232(COM1)
  • Page 32 Connection Note Explanation PC D-SUB No.2 RS-232 PC D-SUB No.3 RS-232 PC D-SUB No.5 RS-232 485 Converter T+ RS-485/422 485 Converter T- RS-485/422 OUT+ 485 Converter R+ RS-485/422 OUT- 485 Converter R- RS-485/422 Reference 1. For the computer RS-422 and RS-485 line PIN numbers, please refer to the PCI card or converter (RS-422) manual.
  • Page 33: Relay Input(Option)

    5.1.4 Relay Input (Option) * If 5~24V voltage is applied to terminals IN1, IN2, signal is input into each Input unit.
  • Page 34: Weight Calibration Mode

    6. Weight Calibration Mode 6.1 What is weight calibration? Calibration for adjusting the weight display so the displayed value corresponds with the actual weight. 6.2 How to enter the weight calibration mode To enter the weight calibration mode, open the front of the indicator and turn ON the power when the CAL S/W of the S/W depicted in the figure below is ON.
  • Page 35: Weight Calibration Menu (Cal1 - Cal9)

    6.3 Weight calibration menu (CAL1 – CAL9) CAL 1 : Maximum capacity calibration CAL 2 : Minimum division calibration CAL 3 : Zero-set and span calibration CAL 4 : Hopper weight setting CAL 5 : Direct calibration CAL 6 : Zero adjustment CAL 7 : Factor calibration CAL 8 : Gravity adjustment CAL 9 : Setting dual range...
  • Page 36 6.3.1 CAL 1 (Maximum capacity calibration) Function: Setting Maximum Value Range of set value: 1 ~ 99,999 Used key Display Descriptions C= 10000 Max value = 10000 kg Using numeric keys Enter maximum weight Max value = 10 kg =Set, =Cancel Reference 1.
  • Page 37 6.3.3 CAL 3 (Zero-set and span calibration) CAL 3-1 Function: Setting Multi Calibration Step Range of set value: 1 ~ 5 Used key Display Descriptions Setting multi calibration for step 1 STEP- 1 (CAL3-3 and CAL 3-4 are carried out Using numeric keys once) Set the multi step,...
  • Page 38 CAL 3-3 Function: Setting Weight Range of set value: 1 ~ 99,999 Used key Display Descriptions It means the weight setting mode. LOAD 1 (Number = multi calibration number) Using numeric keys Set the counterweight values W=100.00 100.00 (unit: Kg or Ton) =Set, =Cancel 0.10...
  • Page 39 6.3.4 CAL 4 (Hopper weight setting) CAL 4-1 Function: Setting Multi Calibration Step Range of set value: 2 ~ 5 Used key Display Descriptions Setting multi calibration for step 2 STEP- 2 (for hopper system) Using numeric keys Set the range. Setting multi calibration for step 5 STEP- 5 (for hopper system)
  • Page 40 6.3.5 CAL 5 (Direct Calibration)) CAL 5-1 Function: Direct input about the zero value of loadcell Range of set value: 1 ~ 99,999 Used key Display Descriptions ZE-CAL Direct zero input mode Using numeric keys 0.0000 Enter the output value. Loadcell zero = 0.0000 mV/V =Set, =Cancel...
  • Page 41 6.3.6 CAL 6 (Zero-set adjustment) Function: Zero adjustment - calibration when any zeroing error occurs. Used key Display Descriptions Empty the load tray and press the setup 2-CAL key. The current weight value is displayed. 1234 =Set, =Cancel Confirm 'Stable' and press the setup key. - - - Zero adjustment in progress...
  • Page 42 6.3.8 CAL 8 (Gravity correction) Function: Gravity Adjustment Used key Display Descriptions It means you accessed to the G-CAL menu for the gravity adjustment. Using numeric keys Enter an initial gravity value. Gr-CAL Set the gravity for the production =Set, =Cancel 9.XXXX place.
  • Page 43 CAL 9-2 Function: Setting the applied section for the Dual Range Range of set value: 0 ~ 99999 Used key Display Descriptions Dual range is applied to less than M 1000 1000kg. Using numeric keys Dual range is applied to less than M 5000 Enter dual values.
  • Page 44: Weighing Mode

    7. Weighing Mode 7.1 Zero-set Function (used when changing the zero-set) ■ Zero-set range: within the range set in F1.09 Display Part or Used Keys Load Plate Description 0.100 Step 1 Empty State with zero changed Step 2 Push the zero key State after performing zero 0.000 function.
  • Page 45: Tare Function

    7.2 Tare function (used when weighing using a container) ■ Maximum tare setting range: maximum capacity * Caution: The weight value including the container weight cannot exceed the maximum capacity. Display Part or Used Keys Load Plate Description State with tare 1.000 Tare Step 1...
  • Page 46: Set Point Changing

    7.4 Set point changing Display Part or Used Keys Load Plate Description Step 1 Press the Item key SP-1 Displayed current setpoint Step 2 number. 00000 Step 2 Change the set point value Step 3 Set point is saved Repeat setp1~3 until end Step 4 point of max set point 7.5 Subtotal Print...
  • Page 47: Total Print

    -------------------------------------------- SUB-TOTAL --------------------------------------------- DATE 2012/ 1/ 1 TIME 09:30 COUNT TOTAL 350.0 kg Reference 2. The subtotal DATA is automactically or manually deleted based on the menu number[F7-03]. 7.6 Total Print Display Part or Used Keys Load Plate Description Push No.5(Total)key “No.5 key pushed”...
  • Page 48: Test Mode

    8. Test Mode How to enter test mode Press the key whilst in normal mode and select test mode by pressing After testing in test mode, press key to enter weighing mode. Test Menu (1 – 9) 1 : Key Test 2 : Display Test 3 : AD Test 4 : Communication Test...
  • Page 49: Key Test

    8.1 TEST 1 (Key Test) Function: Key test Used key Display Descriptions When you press any key to test, the KEY 01 number and code for the key are : Next Menu displayed on the screen. Other keys : Test <...
  • Page 50: Communication Test

    8.3 TEST 3 (AD Test) Function: Load cell test and A/D conversion test Used key Display Descriptions : Cancel The internal value for the current XXXXXX weight value is displayed. : Next Menu Reference 1. Press to display the load cell output as m/V units. 8.4 TEST 4 (Communication Test) Function: Serial Communication Test Used key...
  • Page 51: 1/10 Division Test

    8.6 TEST 6 (A-OUT Test _OPTION) Function : A-Out Test Used key Display Descriptions The output level of Aout is raised by 25% each : Next Menu time the ▲ key is pushed. Other keys : Test Reference 1. This test operates only if Analog out Option Card is mounted. 8.7 TEST 7 (1/10 division Test) Function: 1/10 division Test Used key...
  • Page 52: Set Mode

    9. Set Mode 9.1 How to enter set mode Press whilst in normal mode and select Set Mode by pressing key. After testing in set mode, press to enter weighing mode. Classification Menu SubMenu General Setting F1.01 Set AD speed F1.02 Set average filter F1.03...
  • Page 53 Classification Menu SubMenu 2. Communication F2.01 Set Device ID Setting F2.02 Set Data Transmission Rate F2.03 Set COM1 Port Setting F2.04 Set COM1 Baud Rate F2.05 Set COM1 Out Data F2.06 Set COM1 Output Format F2.07 Set COM1 Output mode F2.08 Set COM2 Port Setting F2.09...
  • Page 54 Classification Menu SubMenu 5. Relay Setting F5.01 Relay Out mode F5.02 Ext input 1 F5.03 Ext input 2 F5.04 Relay Reverse On/Off F5.05 Relay Display Classification Menu SubMenu 6. TCP.IP F6.01 Set DHCP F6.02 Set Server /Client F6.03 EXI-200’s IP Server mode F6.04 Subnet Mask F6.05...
  • Page 55: General Function

    9.2 General Function Menu-F1.01: Set AD Speed Function AD Speed Setting Display Part Meaning 1-01. Converting speed 5 times per second Converting speed 10 times per second 1-01. 1-01. Converting speed 20 times per second 1-01. Converting speed 40 times per second Set Range (0~8) 1-01.
  • Page 56 Menu-F1.04: Set Band Stop Filter Function Band Stop Filter Setting Display Part Meaning Set Range 1-04 : XX Band Stop Filter use setting. (0: Not use, 1: Use) (0 ~ 1) Initial Value : 0 Set Range H-FrEq Band Stop Filter High Frequency setting.. (1 ~ 1600) Initial Value : 60 Set Range...
  • Page 57 Menu-F1.08: Set Weight Back-up Function Weight Back-up Setting Display Part Meaning Set Range 1-08. 0 Weight back up function is not used (0, 1) 1-08. 1 Weight back up is used (based on operation) Menu-F1.09: Set Zero Key Range Function Zero Key Range Setting Display Part Meaning...
  • Page 58 Menu-F1.13: Set F1 Key Use Type Function F1 Key Use Type Setting Display Part Meaning F1 Key used as the tare/tare cancelling key 1-13.00 1-13.01 F1 Key used as the total/net weight key 1-13.02 F1 Key used as the Hold key Set Range 1-13.03 F1 Key used as the Holdless key...
  • Page 59 Menu-F1.16: Set F4 Key Use Type Function F4 Key Use Type Setting Display Part Meaning 1-16.00 F4 Key used as the tare/tare cancelling key 1-16.01 F4 Key used as the total/net weight key 1-16.02 F4 Key used as the Hold key Set Range F4 Key used as the Holdless key 1-16.03...
  • Page 60 Menu-F1.21: Set Automatic Hold Canceling Conditions Function Automatic Hold Canceling Conditions Setting Display Part Meaning Set Range Hold is canceled when the value is changed by more 1-21. XX (00 ~ 99) Initial Value : 10 Than 00 % of the hold value Menu-F1.22: Set Key Operating Conditions Function Key Operating Conditions(Zero, Tare) Setting...
  • Page 61: Communication Setting

    9.3 Communication Setting Menu-F2.01: Equipment Number Function Equipment Number Display Part Meaning Set Range 2-01. XX (0 ~ 99) Desired device ID may be entered. Initial Value : 0 Reference 1. This function can be used as the indicator unique ID when using COMMAND mode. Menu-F2.02: Data Transfer Speed Setting Function Data Transfer Speed Setting...
  • Page 62 (0 ~ 3) 2-06. 2 18 byte Format(AND, FINE) 2-06. 3 22 byte of CAS with relay status Reference 1. Refer to <Appendix 1> for the communication format Menu-F2.07: COM1 Communication Method Function COM1 Communication Method (RS-232C - Output mode)
  • Page 63 (0 ~ 3) 2-11. 2 18 byte Format(AND, FINE) 2-11. 3 22 byte of CAS with relay status Reference 1. Refer to <Appendix 1> for the communication format Menu-F2.12: COM2 Communication Method Function COM2 Communication Method (RS-232C - Output mode)
  • Page 64 22 byte of CAS 2-16. 0 Set Range 2-16. 1 10 byte of CAS (0 ~ 3) 2-16. 2 18 byte Format(AND, FINE) 2-16. 3 22 byte of CAS with relay status Reference 1. Refer to <Appendix 1> for the communication format...
  • Page 65 Menu-F2.17: RS-422/485 Communication Method Function RS-422/485 Communication Method (Output mode) Display Part Meaning Data is not transmitted 2-17. 0 2-17. 1 Transmitted only if the print key is pushed Transmitted in both stable/unstable cases 2-17. 2 (Stream Mode) 2-17. 3 Transmitted only if the weight is stable Set Range (0 ~ 8)
  • Page 66: Analogue Output Setting

    9.4 Analogue Output Setting Menu-F3.01: A-out activation range setting Function Set A-Out range Display Part Meaning Set Range 3-01. 0 Unipolar(operating in +weight) (0, 1) 3-01. 1 Bipolar(operating in ±weight) Menu-F3.02: V-out range setting Function V-out range Display Part Meaning 3-02.
  • Page 67 Menu-F3.06: Maximum weight of A-out Function Maximum Output Weight Value upon Using Analog Out option Display Part Meaning Set Range 1000 Maximum output at 1000 kg (0 ~ Max weight) 2000 Maximum output at 2000 kg Initial value : Max weight Menu-F3.07: Adjust zero of A-out Function Adjust the Zero Output upon Using Analog Out option...
  • Page 68: Hardware Set Function

    9.5 Hardware Set Function Menu-F4.01: Set Date Function Set Date Number Key Display Part Meaning : Data Designation 10.08.17 August 17th, 2010 Menu-F4.02: Set Time Function Set Time Number Key Display Part Meaning : Data Designation 11.30.10 30 minutes and 10seconds past 11 o'clock Menu-F4.03: Set Alibi memory Function Set Alibi memory...
  • Page 69: Relay Setting

    9.6 Relay Setting Menu-F5.01: Set Relay mode Function Set Relay mode Display Part Meaning Limit Mode 5-01. 0 Set Range 5-01. 1 Limit type Checker Mode (0 ~ 3) 5-01. 2 Checker Mode 5-01. 3 Programmable Mode <Limit Mode> Note. 1.
  • Page 70 <Checker Mode> Note. 1. Set value input requirement: SP2 > SP1 2. Relay Output LOW : SP1 ≥ W  ON HIGH : W ≥ SP2  ON Finish(OK) : SP1 < W < SP2  ON Near zero: Set value ≥ 0 range output...
  • Page 71 <Limit type Checker Mode> Note. 1. Set value input requirement: SP2 > SP 1 2. Relay Output LOW : SP1 ≥ W  ON HIGH : W ≥ SP2  ON Finish(OK) : SP1 < W < SP2  ON Near zero: Set value ≥...
  • Page 72 <Programmable Mode> Note. 1. Set value input requirement: SP4 > SP3 > SP2 > SP1, SP5 > SP6 > SP7 > SP8 2. Relay Output Step 1 : W ≥ SP1  ON, W ≤ SP8  OFF Step 2 : W ≥ SP2  ON, W ≤ SP7  OFF Step 3 : W ≥...
  • Page 73 Relay Mode OUT 1 OUT 2 OUT 3 OUT 4 Step 1 Step 2 ᆷ 1_Limit Mode Zero Finish(Stable) SP1 ≤ W SP2 ≤ W HIGH Finish(OK) ᆷ 2_Limit type Checker Mode Zero W ≤ SP1 SP2 ≤ W SP1< W<SP2 HIGH Finish(OK) ᆷ...
  • Page 74 Menu-F5.02: Set Ext Input 1 Function Set Ext Input 1 Display Part Meaning External In1 is zero key 5-02. 00 Set Range 5-02. 01 External In1 is tare/tareless key (0~4) 5-02. 02 External In1 is print key 5-02. 03 External In1 is hold key 5-02.
  • Page 75: Tcp Ip

    9.7 TCP IP Menu-F6.01: Set use DHCP Function Set use DHCP Display Part Meaning Set Range 6-01. 0 Use DHCP (0, 1) 6-01. 1 Not use DHCP(Static) Menu-F6.02: Set TCP mode Function Set TCP mode Display Part Meaning Set Range 6-02.
  • Page 76 Menu-F6.06: Set EXI-200(Server) TCP Port Function Set EXI-200(Server) TCP Port Display Part Meaning 5000 TCP port number = 5000 Set Range (0 ~ 65535) 20000 TCP port number = 20000 Initial value : 20306 Menu-F6.07: Server TCP Output mode Function Server TCP Output mode Display Part Meaning...
  • Page 77 Menu-F6.10: Client TCP Output mode Function Client TCP Output mode Display Part Meaning 6-10. 0 Data is not transmitted 6-10. 1 Transmitted only if the print key is pushed 6-10. 2 Transmitted in both stable/unstable cases (Stream Mode) 6-10. 3 Transmitted only if the weight is stable Set Range (0 ~ 8)
  • Page 78: Print Function Setting

    Display Part Meaning Printer is not used 7-01. 0 Set Range 7-01. 1 CAS DEP Ticket Print Standard Type (0 ~ 3) 7-01. 2 CAS DLP Label Print Standard Type 7-01. 3 CAS BP Label Printer Menu-F7.02: Set Print Form...
  • Page 79 【 Form 1 】 【 Form 2 】 【 Form 3 】 Date, Time, Date, Time, Date, Time, Serial No., Item No., Net Weighing No., Net Weight Gross Weight, Tare, Net Weight Weight 2009.07.07[TUE] 12:30:46 2009.07.07[TUE]12:30:46 2009.07.07[TUE] 12:30:46 ID_11 , 50.0 kg 50.0 kg Gross :...
  • Page 80 Item Number 2 byte Item Name 10 byte Print count 3 byte Date 10 byte Time 8 byte  CAS BP Series Printer Protocol Parameter Description Data Length Gross Weight 7 byte Tare Value 7 byte Net Weight 7 byte Net (‘.’...
  • Page 81 (032: print, all others: do not print) and everything is printed from data 1 until data 255. Reference 3. To add company name "CAS" to the existing print format, designate as follows P00-032 (ASCII code 32: data starts),...
  • Page 82 Menu-F7.06: Set Printing Delay Time Function Set Printing Delay Time Display Part Meaning Set Range 7-06. 1 (0 ~ 200) Issue print after 00 x 10ms Initial Value: 1 Menu-F7.07: Set Print Condition Function Set Print Condition Display Part Meaning Set Range 7-07.
  • Page 83: Error Message

    10. Error Message 10.1 Errors that can occur in weight calibration mode Error Cause Solution Lower resolution. Resolution = maximum allowed weight/value of 1 increment. The set resolution has exceeded the Err 20 Adjust maximum allowed weight in CAL 1 of weight calibration allowance limit of 1/30,000.
  • Page 84: Errors That Can Occur In Weighing Mode

    10.2 Errors that can occur in weighing mode Error Cause Solution Due to unstable load the weighing scale cannot Move the weighing scale to a flat, vibration-free Err 01 be reset. surface and turn on the power. Load cell is not properly connected or there is a Check if the main body and the load plate are Err 02 problem with the A/D convertor.
  • Page 85: Appendix 1> Data Format

    Data (8 bytes) └──┘ └──┘ └──┘ US (Unstable) GS (Gross Weight) Unit (kg/t) ST (Stable) NT (Net Weight) OL (Overload) * 22 Bytes for CAS (Relay status) Data (8 bytes) CR LF └──┘ └──┘ │ │ │ └──┘ US (Unstable)
  • Page 86 *Weight Date (8 byte) Example Byte No weight 13.5kg ‘ ‘ ‘ ‘ ‘ ‘ ‘ ‘ ‘1’ ‘3’ ‘.’ ‘5’ 135kg ‘ ‘ ‘ ‘ ‘ ‘ ‘ ‘ ‘1’ ‘3’ ‘5’ ‘ ‘ -135kg ‘-‘ ‘ ‘ ‘ ‘ ‘...
  • Page 87: Appendix 2> Command Mode 1 Description

    Appendix 2> Command Mode 1 Description CAS <NT-500 Command> Indicator Reception Function Indicator Response dd RW CR LF Request for Weight Data Transmit the data in the set format upon command input Execute the zero and retransmit dd MZ CR LF to PC...
  • Page 88: Appendix 3> Command Mode 2 Description

    Appendix 3> Command Mode 2 Description CAS <NT-570 Command> Reference 1. Command Mode Table Command data to NT-570A Command description NT-570A Respond ZERO key Z CR TARE key Return the received T CR GROSS key Return the received K G CR...
  • Page 89: Appendix 4> Command Mode 3 Description

    Appendix 4> Command mode 3 Description CI-5000 : Transmission only if data is requested ( 1 byte communication) Appenix 5> ASCII Table CHA CODE CHA CODE CHA CODE CHA CODE CHA CODE CHA CODE Space “ & ‘ < >...
  • Page 90: Appendix 6>Modbus-Rtu Protocol

    Appendix 6>MODBUS-RTU PROTOCOL The registry read and write shown below of MODBUS-RTU protocol can be managed based on the requirements included in the reference document regarding the Modicon PI-MBUS-300 standard. For selection of communication with Modbus-RTU, the serial communication settings paragraph has been included.
  • Page 91 FUNCTION 3: Read holding registers Request Register1 Address Funcion No. register 2 bytes Address 0x03 0x0000 0x0002 Total. bytes = 8 Reply Address Funcion No.bytes Register1 Register2 2 bytes 0x03 0x04 0x0064 0x00C8 Total. bytes = 3 + 2*No. register + 2 -number of register = number of modbus register to be read, start at address 1 register.
  • Page 92 Communication Error Management The management of communication data error must be controlled by CRC (cycle redundancy check). If a communication error occurs, the slave does not reply to any request. The master must consider the time-out before receiving the reply. If no reply is received, it can be inferred that a communication error has occurred.
  • Page 93 Register List The Modbus-RTU protocol registers capable of being executed in this instrument are as follows. R = Register that can be used as read-only W= Register that can be used as write-only R/W = Register that can be used as for both read and write H = Higher word of the Double word composing the register L = Lower word of the Double word composing the register Saving to...
  • Page 94 40038 Ext_Output 40042 Analog out Span Weight H 0~99999 40043 Analog out Span Weight L 0~99999 40044 Analog out Zero Adjust H 0~99999 40045 Analog out Zero Adjust L 0~99999 40046 Analog out Span Adjust H 0~99999 40047 Analog out Span Adjust L 0~99999 Analog Out V-Out range 40050...
  • Page 95 40066 ADC reserved 40067 ADC reserved 40068 ADC reserved 40081 Stable range 0~99 40082 Zero tracking range 0 : OFF 40083 Weight back up 1: ON 40084 Zero key range 0~99 Tare key range 0~99 40085 Initial zero range 0~99 40086 Overload range 40087...
  • Page 96 40173 Set Date 40174 Set Hour 40175 Set Minute 40176 Set Second 40177 Set use Alibi memory 40178 reserved 40199 Local IP1 40200 0~255 (000. XXX. XXX. XXX) Local IP2 40201 0~255 (XXX. 000. XXX. XXX) Local IP3 40202 0~255 (XXX.
  • Page 97 Gate way3 40214 0~255 (XXX. XXX. 000. XXX) Gate way4 40215 0~255 (XXX. XXX. XXX. 000) 0 : OFF 40216 Set DHCP 1: ON Set TCP/IP Mode 0 : Server 40217 (Server mode, Client mode) 1: Client 40218 Set Local Port Number 0~65535 40219 Set Server Port Number...
  • Page 98 MEMO...
  • Page 99 MEMO...

Table of Contents