Immunity Test
IEC 60601-1-2 Test Level
Conducted disturbances
3 V
induced by RF fields
0,15 MHz – 80 MHz
IEC 61000-4-6
6 V in ISM bands
between 0,15 MHz and
80 MHz
80 % AM at 1 kHz
Radiated RF EM
3 V/m
fields
80 MHz – 2,7 GHz
IEC 61000-4-3
80 % AM at 1 kHz
6
Compliance Level
3 V
Portable and mobile RF communications equipment should be used no closer to any part of the EUT, including cables, than the
recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
0,15 MHz – 80 MHz
6 V in ISM bands
Recommended separation distance:
between 0,15 MHz and
80 MHz
80 % AM at 1 kHz
3 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Where P is the maximum output power rating of the transmitter in watts(W) according to the transmitter manufacturer and d
is the recommended separation distance in meters(M).
Field strengths from fixed RF transmitters as determined by an electromagnetic site survey, should be less than the compliance
level in each frequency range.
Interference may occur in the vicinity of equipment marked with the following symbol:
Electromagnetic Environment – Guidance
80 MHz to 800 MHz
800 MHz to 2,7 GHz