Page 1
COMPACT 7600 Periodic Maintenance Procedure 00-878852-01 Used and Refurbished C-Arms are Available from www.SharpMedical.com - Call us at 800-969-9800 An Independent C-Arm Service Provider. This PDF provided for research / historical purposes only.
Page 2
This procedure does not contain any steps which can be performed by the system operator. The steps contained in this procedure should not be attempted by anyone who is not specifically trained or authorized by OEC Medical Systems, Inc. to work on the COMPACT 7600 System.
Page 3
COMPACT 7600 Periodic Maintenance, 00-878852-01 Revision History Dash Date Change Description Rev 1 -01 July 1996 Initial Engineering Release Rev 2 -01 Aug 1996 Added fast stop button operational check Rev A -01 Aug 1996 Manufacturing Release Used and Refurbished C-Arms are Available from www.SharpMedical.com - Call us at 800-969-9800...
Page 4
COMPACT 7600 Periodic Maintenance, 00-878852-01 Used and Refurbished C-Arms are Available from www.SharpMedical.com - Call us at 800-969-9800 An Independent C-Arm Service Provider. This PDF provided for research / historical purposes only.
COMPACT 7600 Periodic Maintenance, 00-878852-01 Table of Contents 1. General Instructions ..............................1 1.1 Tools and Test Equipment ............................1 1.2 Necessary Test Equipment............................1 1.3 COMPACT 7600 Periodic Maintenance Form......................1 2. General Safety ................................1 3. Begin PM Form................................2 4. Functional Checks..............................2 4.1 Mechanical Checks ..............................2 4.2 Inspect Drag Cable..............................2...
Page 6
COMPACT 7600 Periodic Maintenance, 00-878852-01 Used and Refurbished C-Arms are Available from www.SharpMedical.com - Call us at 800-969-9800 An Independent C-Arm Service Provider. This PDF provided for research / historical purposes only.
1.3 COMPACT 7600 Periodic Maintenance Form The COMPACT 7600 Periodic Maintenance Form must be completed as the procedures in this manual are performed. If the system does not pass a portion of the procedure, correct the problem before proceeding with the PM.
2COMPACT 7600 Pe 3. Begin PM Form 1. At the top of the COMPACT 7600 Periodic Maintenance Form check whether you are performing a Semi-Annual or Annual PM. 2. Completely fill in the customer information block, including the System Serial Number and the Field Service Report (FSR) Number.
COMPACT 7600 Periodic Maintenance, 00-878852-01 4.4 Inspect/Adjust System Power Supply Voltages WARNING Electrical circuits inside the equipment use voltages that are capable of causing serious injury or death from electrical shock. Use appropriate precautions. 1. The power supplies for the system are located in a perforated metal enclosure on the left side of the generator cabinet.
COMPACT 7600 Periodic Maintenance, 00-878852-01 4COMPACT 7600 Pe 4.6 Perform System Functional Test Steps within this procedure produces X-rays. Use appropriate precautions. WARNING: NOTE: Use the illustration below as a reference when performing the Operational Verification. Press the C-up and C-down buttons and verify that the c-arm lifts smoothly from top to bottom limits.
Page 11
COMPACT 7600 Periodic Maintenance, 00-878852-01 7. While taking x-rays in the Auto Fluoro mode press the slot collimator open and close buttons and verify that the collimator opens and closes. 8. While taking x-rays in the Auto Fluoro mode press the slot collimator rotate buttons and verify that the collimator rotates in both directions.
5. Generator Accuracy Verification/Calibration 5.1 Verify X-Ray Calibration Refer to the KV CALIBRATION and MA CALIBRATION section of the COMPACT 7600 Service Manual (00-878850-01) for instructions on checking the kV and mA. The table below lists the techniques to check and the tolerances allowed. If any portion of the Generator Accuracy Verification fails to meet specifications proceed to section 5.2 (Perform X-Ray Calibration) of the PM procedure.
6. Mail the PM form to: OEC Medical Systems, Inc. Attention Technical Support, 384 Wright Brothers Drive, Salt Lake City, Utah 84116 NOTE: OEC Regional Service offices may require the completed PM form and films to be routed through their offices prior to being forwarded to OEC Medical Systems, Inc. in Salt Lake City.
Page 14
COMPACT 7600 Periodic Maintenance, 00-878852-01 8COMPACT 7600 Pe Used and Refurbished C-Arms are Available from www.SharpMedical.com - Call us at 800-969-9800 An Independent C-Arm Service Provider. This PDF provided for research / historical purposes only.