Leonardo - Siemens SCD 1897-M Replacements Of Parts

18" color tft monitor
Table of Contents

Advertisement

10
Fig. 3:
Image1
NOTE
Select the test pattern
- Select zoom, drag the white area to the center of the screen
- Set the slider for "contrast" to the maximum.
Adjust using contrast minus so that 95% is "well" visible in the 100% field.
- Use the slider for "backlight" to adjust to 120, +/-5cd/m in the 100% field.
- Drag the black field into the center of the screen.
- Use the slider for "brightness" to adjust to 0.5, +/-0.1cd/m in the black field.
- Alternately repeat both adjustments (backlight and brightness) until the specified val-
ues are reached.
- Final Check: All steps in the test pattern must be detectable. If not, repeat the adjust-
ment.
The operating buttons must be locked again!

Leonardo

NOTE
Display
Contrast:
Detection of the 95% fields in the 100% field (limitation) is adjusted
using video gain (contrast).
Backlight:
The 100% field (maximum white) is adjusted using the backlight.
Brightness:
Adjustment of the black area (basic brightness and detection of
the 5% field).
(Fig. 3 / p.
This display is operated at 1280 x 1024 pixels / 75 Hertz.
TD00-000.841.02.04.02
05.05
System-related Adjustments
Fig. 4:
Image1
10).
Page 10 of 22
CSPS21
(Fig. 4 / p.
10).