User Manual www.orpc.com Revision 4.2c2 (October 2019) Translated Version: English Original version: German ACL Medical Display Series OR-MD® Made in Germany...
Operation Panel Servicing Tips for Care and Hygiene Maintenance Intervals and Wear Parts Faulty Signals and Alarm Signals Repairs Instructions for Transporting ACL Units Environmentally Correct Disposal Getting in Contact with the Manufacturer Imprint Warranty and Guarantee The Warranty The Guarantee...
General Instructions ID 1 Congratulations on purchasing your new ACL OR-MD . You have decided on a state-of-the-art quality product from a ® next-generation manufacturer. These instructions will give you a general idea of the steps you have to take for starting up and operating this display system.
Classification and Intended Use ID 4 The unit is designed for displaying data such as patient information inside and outside of operating theaters, intensi- ve-care wards and similar areas of application. It is intended to assist a signal source device without a display or as its alternative display.
www.orpc.com Setting Up and Starting Up ID 16 The unit has been cleaned before leaving our production facilities. However, the units are neither sterile nor disinfected on arrival due to transport conditions and packing. Advices for initial cleaning can be found at section 5.1. ID 17 Make sure that the unit and its accessories are complete with the attached delivery note.
Page 6
ID 22 Furthermore, it is necessary to ensure appropriate unit ventilation to protect it from overheating and for reliable and continuous operation. This is the reason why the unit should not be put in a place or under conditions where it is not possible to o er su cient ventilation for the unit’s housing surfaces.
www.orpc.com Connecting the Cables and other Units ID 25 Only use the supplied cables. ID 26 Start o by connecting the potential equalization cable (special greenish-yellow marking) to ensure a su cient potential equalization between the unit and the user. The unit should not be operated in medical areas without connecting up potential equalization.
Safety Instructions ID 32 Please comply with the safety instructions compiled below to prevent fire, electrical shocks, injuries and damage to the unit. ID 85 Protect the unit from exposure to strong infrared radiation like i.e. direct impact of sunlight. ID 86 Protect the unit from strong magnetic fields like the field close to a MRI.
www.orpc.com Electromagnetic Compatibility (EMC) ID 46 Emissions Test Compliance Level Electromagnetic Environment - Guidance RF emissions Group 1 The unit uses RF energy only for is internal function. CISPR 11 Therefore, is RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
IEC 60417-5007 and IEC 60417-5008 disconnection from the mains protective earth IEC 60417-5019 potential equalization connection IEC 60417-5021 manufacturer ISO 15223-1 date of manufacturing ISO 15223-1 2002/96/EG WEEE Annex IV barcode scanner RFID scanner ACL driver set...
www.orpc.com Operation ID 52 You can use the Standby button to turn the unit on and o . This does not disconnect the power supply from the mains though. ID 53 Your unit is equipped with a capacitive operating panel. You can simply touch the glass pane above the symbol to trigger the key.
Website of the “Robert Koch-Institut “with the title “Liste der vom Robert Koch-Institut geprüften und anerkannten Desinfektionsmittel und –verfahren“ (source: Bundesgesundheitsblatt 12/2013) or from ACL in German. The active substances in the following table are expected to be harmless for the unit as long as the listed limits are observed.
www.orpc.com Maintenance Intervals and Wear Parts ID 61 All units are designed for 24/7-operation. This is resulting in highest requirements to components and materials. To ensure the reliable operation under these harsh conditions we recommend to replace the following components prophylactically: Component Exchange After...
Diligent testing of the unit is guaranteed by the manufacturer. He is responsible for safety-related characteristics within the legal limits only if repairs and modifications are performed by personnel authorized and trained or supervised by ACL and if the unit and its accessory are used according to the rules of the intended use. Therefore, there are warranty seals on or in your unit.
Instructions for Transporting ACL Units ID 69 Please use ACL transport systems only to avoid any damages due to transportation conditions. If you need an appropriate transport system, we can send you all required materials for your unit at net costs.
Please get in contact with the manufacturer at the address given in Section 6 or with an authorized disposal company for returning your old units. We would be glad to advise you as to the options for disposing of your ACL unit.
No part of this manual may be reproduced in any form (printing, photocopy, microfilm or any other process) without the prior written permission of the manufacturer or processed, copied or disseminated using electronic systems. We reserve ourselves the right to make modifications in terms of technical progress. All copyrights shall be reserved to ACL GmbH. All rights reserved.
The term “service” is used through this section to state generally applicable items of information on both, guarantee and warranty. You can find the precise type and term of guarantee or warranty for your ACL product in the description on your original invoice and original delivery note.
Please do not hesitate to get in contact with the address specified in Section 6 if you have any questions. guarantee with on-site service The customer reports a defective unit to ACL GmbH by email or phone. If ACL GmbH cannot correct the malfunction on the phone, an ACL service partner will repair the fault at the customer’s location.
It shall be su cient that the product a ected complied with the test requirements in conformity with ACL's specifications for the original product in conformity with the product data sheet. ACL GmbH’s service does not constitute restoring customer data or software.
Afterwards you will get a reply via eMail from our service department containing your RMA process number as well as queries from our service team if a problem needs to be specified. Disinfect the unit. Send it back to ACL GmbH conside- ring the hints from section 5.5.
ACL GmbH. The guarantee claim shall expire - if ACL products have or are used with parts, components or peripheral units not approved by ACL GmbH for the specific application - if services/repairs or other modifications to the product are carried out by persons not authorized by ACL GmbH (warranty seal broken) unless the defect can be proven not to have been caused by these events or be their consequence.