User’s Manual Blood Pressure Monitor Model BP2, BP2A...
Page 2
1. The Basics This manual contains the instructions necessary to operate the product safely and in accordance with its function and intended use. Observance of this manual is a prerequisite for proper product performance and correct operation and ensures patient and operator safety. 1.1 Safety Warnings and Cautionary Advices Before using the product, please ensure that you have read this manual thoroughly and fully...
Page 3
themselves. Do not store the product in the following locations: locations in which the product is exposed to direct sunlight, high temperatures or levels of moisture, or heavy contamination; locations near to sources of water or fire; or locations that are subject to strong electromagnetic influences. This product displays changes in the heart rhythm and blood pressure etc.
Page 4
of the blood of the patient. Do not apply the cuff on an arm with another medical electrical equipment attached. The equipment may not function properly. People who have a severe circulatory deficit in the arm must consult a doctor before using the ...
This product is contraindicated for use in ambulatory environments. This product is contraindicated for use on aircraft. 2.3 About the product product name: Blood Pressure Monitor Product model: BP2 (include NIBP+ECG), BP2A (only NIBP) 1. LED screen Display date, time and power status, etc. ...
Power On/ Off Power On:Press the button to power on. Power off:Press and hold the button to power off. Press to power on the product and press again to start measuring blood pressure. Press to power on the product and touch the electrodes to start measuring ECG. ...
The blood pressure can differ between the right arm and the left arm, and the measured blood pressure readings can be different. The Viatom recommends to always use the same arm for measurement. If the blood pressure readings between both arms differ substantially, check with your physician to determine as to which arm to use for your measurements.
Page 8
measurement takes about 30s. 3. The blood pressure readings will scrolling appear in the product when the measurement finished. 4. The product will automatically release the cuff gas after the measurement is over. 5. Press the button to turn off the power after the measurement, then remove the cuff. 6.
Page 9
3.3 Measure ECG 3.3.1 Before using ECG Before using ECG function, pay attention to the following points in order to obtain precise measurements. The ECG electrode must be positioned directly against the skin. If your skin or hands are dry, moisten them using a damp cloth before taking the measurement. ...
Page 10
3.When the bar if fully filled, the product will show the measurement result. 4.Press the memory button to review historical data. Note: Do not press the product too firmly against your skin, which may result in EMG (electromyograph) interference. The device can store maximum 10 records for ECG data.
The product can pair and transmit with the phone under the wireless coexistence environment (e.g. microwaves, cell phones, routers, radios, electromagnetic anti-theft systems, and metal detectors), but other wireless product may still interface with pairing and transmission between the phone and the product under uncertain environment. If the phone and the product display inconsistent, you may need to change the environment.
6. Specifications Classifications Degree of protection against Type BF electrical shock Protection against electric shock Internally powered equipment Electro-magnetic Compatibility Group I, Class B Apply part Cuff and ECG electrode Environmental Item Operating Storage Temperature 5°C to 45°C -25°C to 70°C Relative humidity (non-condensing) 10% to 95% 10% to 95%...
30 to 250 /min ±2 /min or ±2%, whichever is greater. Accuracy Memory BP2: up to 100 readings for Blood Pressure data Capacity + 10 records for ECG data. BP2A: up to 100 readings for Blood Pressure data Bluetooth RF 2.402-2.480 GHz...
• Do not drop the instrument or treat it roughly in any way. Avoid strong vibrations. • Never open the product! Otherwise the manufacturer calibration becomes invalid! 7.2 Cleaning The product can be repeatedly used. Please clean before reuse as follow: •...
The device has been evaluated to meet general RF exposure requirement. The device can be used in portable exposure condition without restriction. 9. Electromagnetic Compatibility The product meets the requirements of EN 60601-1-2. Warnings and Cautionary Advices Using accessories other than those specified in this manual may result in increased ...
Page 16
The Blood Pressure Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the Blood Pressure Monitor should assure that it is used in such an environment. Compliance Electromagnetic Immunity test IEC60601 test level level environment - guidance Floors should be wood,...
Page 17
6Vrms 150 kHz to 80 should be used no closer to MHz outside ISM bands any part of the system, including cables, than the recommended separation distance calculated from the equation appropriate for the frequency of the transmitter. Recommended separation distances: ...
Page 18
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Blood Pressure Monitor is used exceeds the applicable RF compliance level above, the Blood Pressure Monitor should be observed to verify normal operation.