Sealey DBG5011 Instructions page 3

Dial bore gauge

Advertisement

1mm measurement zone
2.5mm approx
nib projection
Be aware, nib projection is
2.5mm. Aim for 0.75mm to
2.25mm as the
measurement zone.
5.3.
Accurate Measurement
5.3.1. Follow procedure 5.1.1 through 5.1.4.
5.3.2. The bore or gap machined finish must be clean with surface texture 3.2µm to 6.4µm or better than, for fiducial indication.
5.3.3. offer the anvil end into the bore slightly tilted (fig.8 and example fig.10) with the sprung loaded carriage entering just ahead of the anvil
tip. rock the dial gauge tube to the upright position (example fig.11) and beyond, observing the pointer sweep. The three point location
offered by the two guides on the spring loaded carriage and the anvil tip will centralise the head (fig.8).
5.3.4. rotate the bezel such that the "0" [zero] aligns with the largest sweep position of the pointer (fig.9).It is essential to observe the total
sweep of the dial pointer using the small dial. remember from the initial setting the pointer had rotated two full cycles. Mark the position
on the component of where the measurement was taken, if required.
5.3.5. With a micrometer, measure across the anvil and the nib (fig.8), rotate the barrel of the micrometer until the dial pointer aligns with the
same "0" [zero] as the measured bore. Take the reading from the micrometer and record.
5.3.6. With the bore dial gauge now set, the bore can now be measured for taper and ovality by reference back to the initial datum "0" [zero]
reading and counting the 0.01mm divisions between datum and new. note! the larger the deflection the smaller the bore.
fig.10 generic
"offer the anvil end into the bore slightly tilted...."
6.
MAINTENANCE
6.1.
Keep all components dry and clean with a soft micro fibre cloth.
6.2.
return all items to the presentation case after use.
6.3.
store indoors in a temperature controlled dry environment, circa 21°C.
6.4.
This is a precision instrument intended for use by engineers and engineering inspectors, keep out of reach of children.
© Jack sealey limited
fig.8
Component
example only
A support method when setting with a micrometer
Original Language Version
fig.9
fig.11 generic
Component
example only
"rock the dial gauge tube to the upright position....."
dBG5011 Issue: 1 - 23/06/15