INTRODUCTION The MPS Medical Pharm/Pump III is an easy to use, microprocessor controlled, dual mode, quadruple rotor peristaltic pump. Key Features The Pharm/Pump III has the following key features: Variable Volume The pump can be programmed to dispense samples with a volume of between 1.0 and 4095mL.
III allowing a hands free start, stop and resume of all pumping functions. This switch directly mimics the operation of the START/STOP front panel keys. The MPS Medical Pharm/Pump III has an Non-Volatile Real Time Clock hardware real time clock which will maintain the...
When the key is released the display will show the following message: MPS MEDICAL EQUIP SETUP ENTER PASSWORD: Enter the system password (default 1212) and the display will show the following message: ERASE MEMORY ACCEPT NEXT QUIT MPS Medical Pharm/Pump III Operator Manual V5.01...
Hours:Minutes:Seconds format and the date in either American Month/Date/Year format or European Date/Month/Year format. The MPS Medical Pharm Pump will automatically compensate for the number of days in each month, including during a leap...
Care should be taken when setting the date. Printing a Test Label The final menu allows the printing of a test label to confirm the correct function of the label printer and its interface. Print Test Label ACCEPT NEXT QUIT MPS Medical Pharm/Pump III Operator Manual V5.01...
BASIC OPERATING PRINCIPLES The PHARM/Pump III is switched on by pressing the rear panel rocker switch, located above the main socket. When the PHARM/Pump III has initialized the display will show: MPS MEDICAL PHARM/Pump III SOFTWARE VERSION 5.01 After approximately 1 second, assuming the Erase Memory function above has just...
The volume of fluid in the container which is being dispensed from can be entered into the Pharm/Pump III. Every time a dispense is made the dispensed volume will be subtracted from this original volume, giving a running total of volume remaining. If...
From the main menu, pressing the central SETUP soft function key moves the user from the main top level operating menu into the SETUP menu. The following display will typically be visible: Named Syringe 10ml TYPE NEXT QUIT MPS Medical Pharm/Pump III Operator Manual V5.01...
1.0mL and 4095mL. When the first digit is pressed the volume field will be cleared and a cursor will be placed over the number being entered. Enter the appropriate digits, then press the MPS Medical Pharm/Pump III Operator Manual V5.01...
By pressing the Up/Down arrows the letter can be changed through number, upper and lower alphabet and some symbols. This is repeated until the top line has your label (20 characters max). By pressing the key the display shows: CONC: mg/ml NEXT MPS Medical Pharm/Pump III Operator Manual V5.01...
The top line shows the lot number from the last time this memory was used. In this example no lot number has previously been used for this memory, so the default ‘0000000000’ lot number is displayed. MPS Medical Pharm/Pump III Operator Manual V5.01...
This priming will continue indefinitely, until either the START/STOP key or the foot switch are pressed again. MPS Medical Pharm/Pump III Operator Manual V5.01...
Calibrating the Pharm/Pump III Tube Flow Constants The PHARM/Pump III is supplied set up in a default state with flow constants for a standard MPS Medical single lead pumping chamber set, based on measurements...
Page 18
20C. If the PHARM/Pump III is used with different tubing assemblies, or to pump liquids that are at a different temperature and/or have different viscosity then the PHARM/Pump III can be re-calibrated to take account of this.
Setting the Operating Mode The PHARM/Pump III can pump in either manual or auto mode. With manual mode a single dispense is made each time the pump is activated. In automatic mode a number of dispenses can be made automatically, each of the same size, with a pre-programmed time delay between each dispense.
4.10 Setting the Auto Mode Count (Auto Mode Only) When the PHARM/Pump III is set to automatic mode the operator can enter the number of times to repeat the dispense between 1 and 999. If the PHARM/Pump III is set to manual mode this menu is not displayed.
USE NEXT QUIT The letter in the top right hand corner of the display denotes the memory location into which the current set of operating parameters will be stored. Use the UP/DOWN MPS Medical Pharm/Pump III Operator Manual V5.01...
MENU menu. The following display will typically be visible: Vol Remain 1000ml NEXT QUIT Type in the new volume remaining, between 0 and 6000mL, using the numeric keys and press ENTER to accept. MPS Medical Pharm/Pump III Operator Manual V5.01...
14:55:03 11/03/06 NEXT QUIT A dynamic display of the current date and time is given, with the time in standard Hours:Minutes:Seconds format and the date in either American Month/Date/Year or European Date/Month/Year format. MPS Medical Pharm/Pump III Operator Manual V5.01...
Manual dispense A manual dispense can be initiated from the main menu by pressing either the START/STOP key or the foot switch. The PHARM/Pump III will dispense its programmed volume, using the programmed speed profile & direction. During a manual dispense the display shows the following information: A: Syringe 25.0ml...
An automatic dispense can be initiated from the main menu in an identical manner to a manual dispense, by pressing either the START/STOP key or the foot switch. The PHARM/Pump III will dispense its programmed volume, using the programmed speed profile & direction. During an automatic dispense the display typically shows the...
CAL soft key. The following screen will be displayed: Calibrate DEFAULT QUIT Pressing the DEFAULT soft key will restore the standard factory calibration. Enter the measured value of fluid dispensed using the keypad with up to 2 decimal places MPS Medical Pharm/Pump III Operator Manual V5.01...
RESUME the dispense, or ABORT. The resume operation can be performed by pressing the START/STOP key or the foot switch, as well as with the soft key. The abort operation can only be performed by pressing the middle soft key. MPS Medical Pharm/Pump III Operator Manual V5.01...
Dispense option above. LABEL PRINTER The PHARM/Pump III hardware and software has been designed to operate with one specific make and model of label printer, and with one specific size and format of label medium. No other combinations are recommended.
Turn the printer off for at least 5 seconds before proceeding. Next it is necessary to allow the PHARM/Pump III to download the correct operating parameters to the printer. This is done automatically every time the pump is turned on but the printer must be correctly connected and turned on first to allow it to receive them.
Safety door over the rotor is open RELEASE FOOTSWITCH Foot switch permanently held down In Drawback Portion An attempt was made to calibrate on Cannot Calibrate the fly during the drawback portion of a dispense MPS Medical Pharm/Pump III Operator Manual V5.01...
Page 33
100mg/mL OXACILLIN 1g 100mg/mL OXACILLIN 2g 200mg/mL PENICILLIN G 1 MIL ------- PENICILLIN G 1.5 MIL ------- PENICILLIN G 2 MIL ------- PENICILLIN G 2.5 MIL ------- PIPERACILLIN 1g 100mg/mL PIPERACILLIN 2g 200mg/mL MPS Medical Pharm/Pump III Operator Manual V5.01...