INOGEN ONE G4 OXYGEN CONCENTRATOR
TECHNICAL MANUAL
96-06917-00-01
Revision C
GUIDANCE TO AUTHORIZED SERVICE AND REPAIR CENTERS
This Manual applies to the following Inogen, Inc. products:
• Inogen One G4 Oxygen Concentrator, model #IO-400
• Inogen One AC Power Supply, model #BA-401
• DC Power Cord, model #BA-306
• Battery, model #BA-400
• Battery, model #BA-408
CONFIDENTIAL and
96-06917-00-01 rev C
DCR 20-328
INOGEN ONE G4 TECHNICAL MANUAL
Page 1 of 19
PROPRIETARY

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ONE G4 and is the answer not in the manual?

Questions and answers

Summary of Contents for Inogen ONE G4

  • Page 1 96-06917-00-01 Revision C GUIDANCE TO AUTHORIZED SERVICE AND REPAIR CENTERS This Manual applies to the following Inogen, Inc. products: • Inogen One G4 Oxygen Concentrator, model #IO-400 • Inogen One AC Power Supply, model #BA-401 • DC Power Cord, model #BA-306 •...
  • Page 2: Table Of Contents

    TABLE OF CONTENTS INFORMATION FOR PROVIDERS OF THE INOGEN ONE G4 ........... 3 1.1................. 3 AUTION AND ARNING TATEMENTS SETTING UP A PATIENT ON INOGEN ONE G4 ..............4 2.1......................4 NDICATIONS FOR 2.2....................... 4 YSTEM OMPONENTS 2.3.
  • Page 3: Information For Providers Of The Inogen One G4

    This Technical Manual will familiarize you with provider-specific information regarding the Inogen One G4 Oxygen Concentrator and its accessories. Before reading this Technical Manual, please read and review the Inogen One G4 Patient Manual for description and indications for use of the device.
  • Page 4: Setting Up A Patient On Inogen One G4

    6. Plug into closest AC outlet. 7. Turn on the Inogen One G4 by pressing the on/off button. 8. Use the + (increase flow) or – (decrease flow) buttons to adjust the Inogen One G4 to the flow setting prescribed by the physician or clinician.
  • Page 5: Selecting The Proper Flow Setting

    OCD’s efficacy at maintaining the patient’s blood oxygen saturation. As an oxygen concentrator, the Inogen One G4 does not contain a finite stored volume of oxygen, such as with compressed gas or liquid cryogenic systems. The Inogen One G4 can provide oxygen to the patient as long as a source of electricity is available.
  • Page 6: Servicing The Inogen One G4

    3.1. Maintenance by the Provider 3.1.1. Authorized Repair Centers The Inogen One G4 is only intended to be repaired by authorized repair centers. If a repair is required, please contact Inogen to locate your nearest authorized repair center. If you would like to become an authorized repair center, please contact Inogen for access to certification training, service instructions, component part lists and the necessary repair equipment.
  • Page 7 16. Remove PPE and wash hands with soap and water for a minimum of 20 seconds. Do not disassemble the Inogen One G4 or any of the accessories or attempt any maintenance other than tasks described in this Technical Manual unless you have completed a training course through Inogen.
  • Page 8 Inogen One G4 and accessories. 3.1.3. Suggested Materials for Regular Maintenance To perform regular field maintenance (by a technician) on the Inogen One G4 System, you may need: o Inogen One G4 Particle Filter (RP-405) o Output Filter Replacement Kit (RP-404) ...
  • Page 9 This filter is intended to protect the user from small particles in the product gas flow. The Inogen One G4 includes an output filter conveniently located behind the removable cannula nozzle fitting. Inogen suggests that this filter be replaced between patients.
  • Page 10: Maintenance By The Patient

    • Battery Status • Charging Status 3.1.9. Data Download To download data from the Inogen One G4, you will need a micro USB cord and the Inogen field service communication program. To collect data: a) With the Inogen One G4 off, unscrew the access panel screw on the back side of the device and remove the access panel.
  • Page 11 3.2.2. Particle Filter Cleaning and Replacement On the side of the Inogen One G4, ambient air passes through a particle filter that removes dust fragments. This particle filter must be cleaned on a weekly basis to ensure adequate air flow through the device. If the particle screen is not cleaned frequently, the life expectancy of the concentrator will be shortened due to higher internal operating temperatures.
  • Page 12 3.2.5. Battery Care and Maintenance The Inogen One G4 Lithium Ion Battery requires special care to ensure proper performance and long life. Use only Inogen One G4 Batteries with the Inogen One G4 Oxygen Concentrator. The Inogen One G4 Oxygen Concentrator adjusts its oxygen production rate to match the oxygen demand specified by the user flow setting.
  • Page 13: Expected Service Requirements

    3.2.5.5. Storage Instruct patients to remove the battery from the Inogen One G4 when it is not in use to avoid inadvertent discharge. Leaving a battery attached to an idle Inogen One G4 for prolonged periods will result in battery damage that will severely shorten the expected life of the battery.
  • Page 14: Inogen One G4 System Specifications

    4. INOGEN ONE G4 SYSTEM SPECIFICATIONS 5.91”L (15.01cm) / 2.68”W (6.8 cm) / 6.45”H (16.3 cm) Dimensions: 5.91”L (15.01 cm) / 2.68”W (6.8 cm) / 7.15”H (18.2 cm) with single battery Weight: 2.8 lbs (1.27 kg.) (includes single battery) Noise:...
  • Page 15: Inogen One G4 Errors

    5. INOGEN ONE G4 ERRORS 5.1. DEFINITIONS ⚠ 5.1.1. Alert LED: The alert LED is positioned above a warning symbol ( ) on the user interface panel and is used to indicate when an error is occurring. 5.1.2. Breath Detect LED: The breath detect LED is used to indicate proper bolus delivery.
  • Page 16: Low Priority Errors

    The concentrator should be serviced within 30 days. CONFIDENTIAL and 96-06917-00-01 rev C DCR 20-328 INOGEN ONE G4 TECHNICAL MANUAL Page 16 of 19 PROPRIETARY...
  • Page 17 LCD screen, press the mode button to confirm. This resets the system software. 4. Press the power button to resume normal use of the concentrator. CONFIDENTIAL and 96-06917-00-01 rev C DCR 20-328 INOGEN ONE G4 TECHNICAL MANUAL Page 17 of 19 PROPRIETARY...
  • Page 18: Medium Priority Errors

    Ensure air intake and outlet vents have Warning temperature limit. clear access and particle filters are clean. If this error persists, replace the concentrator. CONFIDENTIAL and 96-06917-00-01 rev C DCR 20-328 INOGEN ONE G4 TECHNICAL MANUAL Page 18 of 19 PROPRIETARY...
  • Page 19: High Priority Errors

    5.6. CONTACT for MORE INFORMATION Inogen Website: Patient Issues: http://www.inogenone.com Corporate: http://www.inogen.net Inogen : 301 Coromar Drive Goleta, CA 93117 1-866-765-2800 clientservices@inogen.net CONFIDENTIAL and 96-06917-00-01 rev C DCR 20-328 INOGEN ONE G4 TECHNICAL MANUAL Page 19 of 19 PROPRIETARY...