H&O Equipments Accurett 16g User Manual page 7

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
• Edemas
• Bleeding
• Pain and syncope
• Insufflation of subcutaneous tissue
• Retraction caused by freezing next to orifices
• Cartilage defect during treatment of the nose or the ear
• Infection
• Conjunctival ectopy due to treatments of the eyelid
• Milia
• Depressed or atrophic scars
• Post-inflammatory hyperpigmentation (PIH)
is a frequent concern when treating SL
in darker skin types.
Cryotherapy using Accurett is a simple and practical therapy for SL considering
the following instructions:
1- Select your candidate: A typical candidate for treatment has multiple,
brown, well defined, small and homogenous color lesions with no history
of significant post-inflammatory hyperpigmentation.
2- Perform a spot test by treating one lesion, and evaluate the result
in three weeks.
3- Apply sunscreen and use whitening cream for darker skin types
for 3 weeks before and after treatment.
Empty loading cavity
Blistered cartridge and assembled
filter/opener device
Step 1
I
Empty the loading cavity
Before loading a new cartridge, release any unused gas from
the used cartridge by pressing the button or unscrewing
the applicator. Discard the used cartridge. Please ensure that the
used filter does not remain inside the cavity.
Step 2 I Load a new cartridge
The cartridges are packed in peel-pack blisters with an assembled
filter and cartridge opener included. The filters are used
for protection of the applicator. The opener punctures the cartridge.
Accurett 8g is used with 8g CO
with 16g CO
cartridges. Indications are clearly applied on the
2
cartridge's packaging.
Hold the Accurett's tip end pointed downward. Ensure that the
loading cavity is empty of all residue.
Open the peel-pack. Take out the filter/opener with the tweezers.
Place the filter in the instrument and make sure that it is seated flat
at the bottom of the loading cavity. The trocar of the opener should
be pointed upward. Take the cartridge and put it into the cavity with
the narrow side down, on top of the filter/opener device.
NO
Installation of the assembled filter opener device
cartridges, Accurett 16g is used
2
YES
7
7