Agilent Technologies 85103F Installation Manual page 31

Network analyzer performance upgrade
Table of Contents

Advertisement

5. Turn off the 8510.
6. On the 85102:
a. Remove the top cover.
b. Disconnect W24 from the A17J4 Ref X connector.
c. Connect the BNC-to-SMB cable provided in the upgrade package from the rear panel
LO Phase Lock Out jack to the A17J4 Ref X connector.
7. Turn on the 8510.
8. Press
INSTRUMENT STATE RECALL
9. Press
PARAMETER MENU
10. Press
FORMAT MENU
11. Press
RESPONSE AUTO
12. The signal displayed should ramp up as shown in
Figure 6-3
LO Phase Lock Out and Main Phase Lock Signals
13. On the 85102:
a. Remove the BNC-to-SMB cable.
b. Reconnect W24 to A17J4.
c. Leave the top cover off.
85103F Performance Upgrade Installation Manual
and select
and select
[USER 1 a1]
and select
.
[REAL]
.
The 85103F Upgrade Package (8510A/B to 8510C)
[MORE], [FACTORY PRESET]
.
Figure
6-3.
Checking the 8510C
.
21