Summary of Contents for Daihan Scientific ThermoStable OV-20
Page 1
DAIHAN-Sci.com or ALL for LAB.com Shopping Mall Precise Vacuum Ovens Certifications by - - - ² English Page : P. 11 ~ 26 ISO 9001 Certified. ISO 13485 Certified for Medical Instruments ² Korean Page : P. 27 ~ 52 UL/CUL Certified CE Certified PL Insurance.
Page 2
DAIHAN-Sci.com or ALL for LAB.com Shopping Mall DAIHAN Scientific, Precise Vacuum Ovens 2/52...
This symbol indicates power “OFF” on the power on/off switch. This symbol indicates power “ON” on the power on/off switch. This symbol indicates the connection point of the neutral conductor on the installed unit. This symbol indicates alternating current Voltage. DAIHAN Scientific, Precise Vacuum Ovens 4/52...
10. This product is remotely controlled by connecting to PC through RS232 communication port, and can be transmitted data. 11. CE Certified Products. 12. PL(Product Liability) Insurance DAIHAN Scientific, Precise Vacuum Ovens 5/52...
2 people and by lifting units up holding the indicated positions in the figure below. 4. The product should be positioned in a location where it has enough space for ventilation and prevention from fire. The recommended distance of each side is minimum 20cm from the back DAIHAN Scientific, Precise Vacuum Ovens 7/52...
Page 8
Product Liability (PL) Insurance Admission ¨ DAIHAN Scientific (Co., Ltd.) is a member of Product Liability Insurance (PL Insurance), so that its customers can trust our products, (believe and use its products,) by preparing against the accident which can be occurred during using the products by self- production.
Page 9
In addition, please install using the vibration-proof and rubber, etc. so that the vibration is not transferred to the vacuum pump. The Cold Trap Bath, and Vacuum Pump connected to this product, including the Vacuum Oven can be identified at http://mall.allforlab.com/. DAIHAN Scientific, Precise Vacuum Ovens 9/52...
67 kg 100 kg Power Consumption 600 W 600 W 1.6 kW Power Supply & Cord Plug* 1 Phase, AC 120V, 60Hz or AC 230V, 50/60 Hz * Other Specifications are available upon Customer's Request. DAIHAN Scientific, Precise Vacuum Ovens 10/52...
Page 12
If the regulator SAFETY is set to 0, the heater does not operate. If unexpected sound, smell or smog is generated, pull the main plug out from the wall socket and contact the technical assets of DAIHAN Scientific If irradiated or contaminated samples are placed in the chamber, the warranty is voided.
Page 13
If unusual noise or signs of abnormal behavior are found during use of this product, please immediately contact the supplier or DAIHAN Scientific (Co., Ltd.) after turning off the power supply and disconnecting the power line.
DAIHAN-Sci.com or ALL for LAB.com Shopping Mall technical assets of DAIHAN Scientific (Co., Ltd.). Only technicians of the technical support team at DAIHAN Scientific (Co., Ltd.) or technicians approved through the education from DAIHAN Scientific should be allowed to repair this product.
Page 15
DAIHAN-Sci.com or ALL for LAB.com Shopping Mall ⑤ ⑥ ⑦ ⑧ ⑨ ThermoStable OV-20 (120V,230V) : 250V F 10AH ThermoStable OV-30 (120V,230V) : 250V F 10AH ThermoStable OV-70 (120V,230V) : 250V F 20AH <Fig.3.2.4> Right Side of Precise Vacuum Oven " ThermoStable OV " ⑩...
Page 16
2. Put Samples into the Chamber before performing the operation of the product. 3. Check the Vent Control Valve in the vacuum control unit of the oven, and close it by completely turning clockwise. 4. Open the Vacuum Control Valve by turning horizontally. DAIHAN Scientific, Precise Vacuum Ovens 16/52...
Page 17
The flask used in the vacuum state should be safely used after being fixed using a joint clip and hanging in the Clamp. * RS232C Connector Port for cable connection to PC. Monitoring and Controlling with PC by connecting RS232C cable. <Fig.3.2.8> Precise Vacuum Oven " ThermoStable OV " RS232C Connect DAIHAN Scientific, Precise Vacuum Ovens 17/52...
Page 18
Convert into mode of delay timer and offset function. ④ Control that Reduces or Adds Values, and Does Start/Stop Jog Dial knob ⑤ Operation. PUSH Button Operate Start/Stop. ⑥ <Table 3.2.3> Controller for " ThermoStable OV " DAIHAN Scientific, Precise Vacuum Ovens 18/52...
The range of setting time is from 00:00 to 99hr and 59min(it is possible to adjust at 1min.) and 00:00 indicates continuous operation. If the timer operation is completed, the alarm will sound and heating will be stopped. The timer setting mode can be operated by referring to the jog-shuttle operation method. DAIHAN Scientific, Precise Vacuum Ovens 19/52...
Page 20
You can change or save the setting by using the operation method of the Jog-Dial, and Start/Stop of operation can be performed. If pressing MAIN button, it is switched to the Set Operation Time Mode. DAIHAN Scientific, Precise Vacuum Ovens 20/52...
Page 21
The scope of the compensative value is -20.0℃ ~ +20.0℃, and it is possible to adjust at 0.1℃ unit. After setting the compensation value, the indication of the current temperature is displayed as the temperature including the compensation value. DAIHAN Scientific, Precise Vacuum Ovens 21/52...
Page 22
Informs the time to be displayed the temperature of Please check the temperature of abnormal value due to the abnormality of Error 1 the heated top plate. Temperature Sensor. <Table 3.3.1 > ERROR Message Table DAIHAN Scientific, Precise Vacuum Ovens 22/52...
Do not clean the unit aggressively. Otherwise, the coating or painting could be damaged. - Cleaning Method of the Parts Clean the parts using a neutral detergent and soft cloth. Do not clean the unit aggressively. Otherwise, the coating or painting could be damaged. DAIHAN Scientific, Precise Vacuum Ovens 24/52...
Check the closed state of the Door. Check the degree of adhesion between the Door gasket and the glass. ※ If you have any problem other than the above, please call technical service of manufacturer or an official agent. DAIHAN Scientific, Precise Vacuum Ovens 25/52...
Need help?
Do you have a question about the ThermoStable OV-20 and is the answer not in the manual?
Questions and answers