1. Neon ® Neon ® is a product by Dr. A. Kuntze GmbH which offer high quality and long-term reliability, made in Germany. With its modular design, the Neon Gas warning instrument can be tailored to your application. ® The entry-level version contains two inputs for measurement, two digital inputs, and an alarm relay.
We have to point out that the warranties specified in our trading conditions are valid only if the following conditions are met: ˃ Installation and start-up by Dr. A. Kuntze personnel or trained and authorized technicians ˃ Maintenance of instrument and peripherical equipment according to the instructions of this manual ˃...
Intended use Use these instruments only for the monitoring of ambient air. Use only sensors, assemblies, and accessories made by Dr. A. Kuntze, and make sure that they fit your application. Install and operate the instrument according to this manual. Carry out all steps described, and check all measurements and settings before you activate the controller.
Eventlog Stores up to 100 events Event help Provides help for current events Add-on Functions can be added via code Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Page 10
Interval adustable between 1 second and 24 hours Ring or Stop mode Digital interface Modbus RTU, 19200 bps, 8 Bit, 1 Stopp-Bit and parity even Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
250 V, 6A, 550 VA Operation temperature 0.. 50°C Storage temperature -20.. +65°C Rel. Humidity max. 90% rH at 40°C (non-condensing) Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Attention To preserve the protection class, the flat gasket of the housing and the sealing ring of the frame must be used! Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Or you can open the instrument and put the screws through the holes at the back of the instrument. Mind to replace the covers before closing the instrument. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60...
Mind that the sensor is not subject to strong wind, heat, or direct sunlight. The sensing element must never get wet! Connect the M12 cable. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60...
22 Ohm 2 Watt 150 mA 100 nF 260 V 47 Ohm 2 Watt 1,0 mA 220 nF 260 V 47 Ohm 2 Watt Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
30 = + , 31 = -, max. load 500 Ohm Analog output 2 32+33 32 = + , 33 = -, max. load 500 Ohm Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Page 18
A = + B = - ┴ ┴ = Screen Add-ons (marked in grey) Up to 2 mA outputs, data logging, and RS 485 Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Never remove the SD card while the logging is running! Always set the mode to OFF (not stop!) before removing the card. Otherwise data loss might occur, and an error message will appear. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60...
The key pad version is operated via four keys: DOWN, UP, ESC (= one step back), and OK (= enter). Starting from the desktop (display of the measured values), enter the menu with key DOWN. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
DOWN, the scroll bar follows automatically. Select a parameter with the OK key to enter a value or select a setting from a drop-down list. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60...
Home – Touching this symbol takes you back to the desktop from any position ESC – Touching ESC takes you back to the previous menu. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60...
With a touch screen instrument open the drop-down list by touching the triangle. A list of possible settings appears. Select the desired setting via touch. Press ESC if you want to cancel and leave without changing the setting. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60...
Page 24
The „C“ (clear) key erases the last input ˃ The „AC“ (all clear) key erases the complete input ˃ The „OK“ acknowledges the selection Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Page 25
00:00:01 (hours:minutes: seconds). Unfortunately, for date, code, add-on codes, text, and contact data, this support is not feasible. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60...
Settings for analog output 1. mA out mA out 2 Settings for analog output 2. Add-ons (marked in grey) mA output, Datalogger and Modbus RTU. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Page 27
Settings for storage mode and interval and display of status. network Setting of bus address and bus terminator. Add-ons (marked in grey) mA output, Datalogger and Modbus RTU. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Page 28
Restore to factory settings. addon Activating of add-on codes. code Entering a code. Add-ons (marked in grey) mA output, Datalogger and Modbus RTU. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
/ out system code zero alarm in / out system code With code 0202, access is restricted, and the number of symbols decreases. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
For more information, see the chapter “System functions”. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60...
Here you can set the slope written on the sensor element. Sensor test You can adjust the interval between subsequents tests. Next test Shows when the next test will take place. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
˃ With the SAVE button you store the results. With DISCARD you reject the results and keep the previous calibration values. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Note The zeropoint calibration results of sensor 1 is shown in the left column and of sensor 2 in the right column. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
For both inputs you can define whether the input should work as a normally open (NO) or normally closed (NC) contact. in / out digital 1 assignment relay 1 contact Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Note When you leave the menu, all relays and all analog outputs are reset to their previous settings. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
PC and then load them into the Neon instrument via SD card. ® Note Settings can be duplicated only to instruments of the same type. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
The display shows the name of the created file. The name consists of Paddmmyyyy_hhmm.set successful! the parameter and the date. Acknowledge with OK. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
After installation, the instrument initializes. During that process, the time is displayed on the desktop. Note You can only load settings files for the same parameter. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Progress is indicated by a status bar. 18:03:35 After installation, the instrument initializes. During that process, the time is displayed on the desktop. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Acknowledge with OK. successful! Note The .dia file is a binary file and can only be read with an appropriate configuration program. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
LANGUAGE. A drop-down list shows event info reset option all languages that are available on your instrument. system language language Deutsch Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
If you enter 6, for example, the time will be event info reset option set to 06:00:00. system time time 05:21:10 date 31.07.2011 Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
0000, and the display shows the desktop with the measured values. If you set the interval to 0, the autolock function is deactivated. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60...
TOUCH CAL to start the calibration. The touch screen calibration texts are available in English only. system display display touch cal Calibrating Touch Screen Touch the target (upper left) Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
/ out system code event messages system language time display contact event info reset option system event help no active event Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Note The analog outputs are add-ons, and are shown only if the instrument is equipped with these features. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
All settings are reset, and the instrument initializes. Restoring... Note With touch screen instruments, the touch screen has to be calibrated after reset. The calibration routine starts automatically. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
If the entered code was incorrect, no message appears, and the previous settings are maintained. Note We recommend to reset all settings to factory settings after activating add-ons. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
1 relay 1 is switches, if the measured value is higher than limit 2 relay 2 switches. Addtional alarm can be issued via alarm relay - if you have specified the limits as alarm actions. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
The following events always cause the alarm relay to switch: ˃ Communication error ˃ Unknown measuring module Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
00:01:00 In the submenu LOG SET you can make the following settings: status Logging on stop SD card Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Page 53
The instrument fills in zeros automatically. Example: 1 is interpreted as 00:00:01 / 1: is interpreted as 00:01:00 / 1:1: is interpreted as 01:01:00 Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60...
Power was turned off ˃ The measuring parameter changed ˃ The logging was stopped and restarted ˃ Other parameters for the datalogging were selected Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
2 in the ANALOG 2 menu to “No sensor” the readout of the assigned mA output is automatically reassigned to measuring input 1. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60...
Request Function code 1 Byte 0x03 Start register 2 Bytes 0x0000 to 0xFFFF Quantity of registers 2 Bytes 1 to 125 (0x7D) Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Status of digital inputs and outputs: digital inputs 1 and 2, relays 1-3 ˃ Event messages ˃ Calibration results as in the Cal/Info menu ˃ Instrument data incl. activated add-ons Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
The first register gives the value 250. This represents a temperature of 25.0 and from the third register the unit °C. Registers four and five show the range of the temperature measurement, which is -30.0…140.0°C. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60...
HOLD. We recommend the use the HOLD-value if you have set alarm values in the PLC that you do not want to be activated during maintenance. Note The operation mode can be found in register 2240, refer to table “parameters”. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Minute+second contains the minutes in the upper 8 bits and the seconds in the lower 8 bits. For event messages, a code indicates the nature of the event – refer to table „event codes“. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Page 61
Memory fault Start 10, 11, 13, 14 - 18 and 45 Internal error No communication with the measurement module Unknown measurement module Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Page 62
SD card: load error SD card: logger stopped Lower limit M1 limit value 1 (Gas) M1 limit value 2 (Gas) Lower limit Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Page 63
Cleaning Cleaning – base load dosing Note For events listed as „going“ in the eventlog, the numerical value is increased by 126. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Add-on First analog output Add-on Second analog output Add-on Controller Add-on volume based dosing Add-on Modbus RTU Add-on datalog Add-on ASR ® Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Code, Mode, IN/OUT, System, Alarm, and Add-ons. Mind that not all parameters listed are available in your instrument, according to settings and activated add-ons. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60...
Page 66
4640 Parameter: LF (IL and CM) Measurement 4464 u_16 Measuring range 4536 u_16 Averaging 4328 u_16 C-value 4664 u_16 TDS unit 4616 Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Page 67
Temperature 4472 u_16 coefficient (DES / LF / STD) Digital inputs Dig. input 1 event 6176 Dig. input 1 contact 6168 Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Page 68
Gas: delay relay 1 7520 u_16 Gas: delay relay 2 7528 u_16 Gas: delay relay 3 7512 u_16 Alarm action 6336 u_32 Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Page 69
Gas: measured value 1 Logparameter 6512 Temperature / Gas: measured value 2 Logparameter Raw value 6520 Logparameter S1 7456 Logparameter S2 7464 Logparameter Flow 7472 Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Page 70
ORP: 7440 S1 integral time 6744 i_16 S1 differential time 6752 i_16 S1 dosage check 6784 u_16 S1 pulse type 6864 Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Page 71
Add-on: volume based dosing Dosage l/l (pR / LF) 7256 i_32 (higher bits) 7257 (lower bits) Pulses 7264 u_16 Pump output 7280 u_16 Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
The following table contains the codes of the units and texts of register 3 of the variables: Code Unit 1001 °C 1054 1058 1059 1211 1243 1283 MOhm 1302 mS/cm 1342 1353 1422 1423 1552 µS/cm Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Page 73
2 / S1 / S2 / Relay / Flow 3013 Stop / Ring / OFF 3014 ON/OFF / P / PI / PID / 3point 3015 Pulse-pause / pulse-frequency Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
S will poison the cell, further exposure to Cl will reactivate the sensor. Nitrogen dioxide 10 ppm Ozone 0,25 ppm Sulfur dioxide 20 ppm Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Page 76
Check the settings for alarm and the assignment of the digital inputs in the menü EIN/AUS – > digital. Check the wiring and the sensor elements in the menu IN/OUT – > TEST. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
If the zeropoint calibration is done with an extant concentration or with a reading caused by cross references the measurement is distort and a reliable measurement is not possible. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60...
In the test menu you can operate all relays manually, and set fix values for each available mA output. The test menu can only be used in the Manual mode. The test menu is a submenu of the menu IN/OUT. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60...
To that purpose save the diagnosis file on your SD card. It contains all current values and settings and the eventlog. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60...
Not all events will cause the alarm relay to switch. For many events you can decide for yourself if the relay should switch by defining the event as an alarm action in the alarm action list – see Alarm relay. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
IN/OUT->ANALOG->measurement. If necessary change the parameter. Note For Chlorine and Chlorine dioxide the message only occurs if you connect an Ozone sensor element. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Page 82
If the LED is red go on like decribed by “no sensor”. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60...
Page 83
Check the settings for the analog outputs, and change the settings if necessary. Finally, there are some event messages indicating fundamental problems that you cannot solve on site. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60...
SD card, check if there is enough storage space left on the card. This message appears also if you remove the card without deactivating the logger first. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60...
With the Reset function, the system is set to a defined condition. User settings that might severely impede an evaluation of the measured values, such as a calibration gone wrong, are erased. Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
............32 Dimensions ............. 12 Calibration level code ........30 Display ............43 Check measuring input ........82 Disposal ............78 Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Page 87
Interval (datalog) ..........52 Operating hours ..........47 Key operation ..........20 Operation ............20 via keys ............20 via touch screen .......... 22 Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Page 88
Touch calibration ..........44 M1/ M2 ............82 Touch screen operation ........22 manual ............77 Trouble-Shooting ..........79 Serial number ..........47 Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Page 89
Index Unknown measurement module ..... 84 View ..............80 raw signals ..........47 Warranty ............7 Zoom (mA output) ........... 55 Robert-Bosch-Str. 7a Dr. A. Kuntze Gutes Wasser mit System 40668 Meerbusch +49 (0) 2150-7066-0 / Fax -60 www.kuntze.com...
Page 92
Robert-Bosch-Str. 7a I D-40668 Meerbusch I Germany Fon +49(0)2157066-0 I Fax +49(0)21507066-60 info@kuntze.com I www.kuntze.com...
Need help?
Do you have a question about the Neon Gas and is the answer not in the manual?
Questions and answers