iHealth PT1 User Manual

iHealth PT1 User Manual

Clinical digital thermometer

Advertisement

Advertisement

Table of Contents
loading

Summary of Contents for iHealth PT1

  • Page 1: Table Of Contents

    Clinical Digital Thermometer Model: PT1 User’s Manual Version 1.1 Contents Indication for Use Product Introduction Structure and Components Speci cations Contraindications Safety Warnings Instruction for Use Unit Switching Replacing Battery Troubleshooting Care & Maintenance Box Contents Signs & Symbols Electronic Compatibility Information...
  • Page 2: Indication For Use

    It is intended for use on adults and children ages 4 and up. Product Introduction Thank you for choosing iHealth Clinical Digital Thermometer, a reliable instrument providing a safe, accurate, and fast temperature measurement. You can easily and quickly develop full awareness of your family’s health condition, without the...
  • Page 3: Contraindications

    9. Waterproofness: probe portion only. 10. Operating conditions: Temperature: 41℉ - 104℉(5℃-40℃) Humidity: 15% ~ 95% RH, non-condensing Atmospheric Pressure: 70kPa ~ 106kPa 11. Transportation and storage conditions: Temperature: -4℉ ~ +131℉(-20℃-55℃) Humidity: 15% ~ 95% RH, non-condensing Atmospheric Pressure: 50kPa ~ 106kPa Contraindications DO NOT use on the injured area.
  • Page 4: Instruction For Use

    (2) About product 1. The thermometer is a precise instrument. Please properly store in the protective case after use to prevent unnecessary damage. 2. DO NOT tread, or bend the thermometer, which may cause product damage. 3. Make sure the thermometer is not damaged from shock. If broken, DO NOT attempt to re-assemble.
  • Page 5: Unit Switching

    (3) Monitor Mode The thermometer could be used orally, rectally or axillary. To enter Monitor Mode, make sure the thermometer is in the OFF mode, press and hold the button for 5 seconds, beep will sound when activated. Then follow the measurement procedures as usual, the results will display in 3 minutes.
  • Page 6: Replacing Battery

    Reminder: If you do not press and hold the On/O button for 8 seconds to con rm the unit selected before the screen switches o , the system will consider the switch as a failure and retain the original unit for temperature measurements. Replacing Battery (1) Installing and replacing battery 1.
  • Page 7: Troubleshooting

    Troubleshooting Problem Item to detect Solution Battery depleted. Replace the battery. Battery have been installed Take out the battery with the wrong polarity. and re-install them Battery are not installed correctly. properly. Unable to carry out measurement as current Replace the battery. battery capacity is too low.
  • Page 8: Box Contents

    8. If you suspect any product issues during the warranty period, please contact customer services for subsequent handling. Box Contents 1 Clinical Digital Thermometer 1 CR1632 battery (installed) 1 thermometer protective case 1 User’s Manual * Use only accessories provided by the original manufacturer, and check for any missing accessories.
  • Page 9: Electronic Compatibility Information

    Electronic Compatibility Information Table 1 - Emission limits per environment Phenomenon Compliance Electromagnetic environment Conducted and CISPR 11 The device is intended radiated RF Group 1, Class B to be used in home emissions healthcare environment. Harmonic IEC 61000-3-2 The device is distortion powered by battery.
  • Page 10: Safety Precautions

    4. Liability for direct or indirect consequential losses caused by the unit is excluded even if the damage to the unit is accepted as a warranty claim. The device is not FDA approved or cleared. Manufactured for iHealth Labs, Inc. USA: iHealth Labs, Inc. www.ihealthlabs.com 120 San Lucar Ct., Sunnyvale, CA 94086, USA...