Instructions for Assembly and Operating SONIX 10D Contents GENERAL INFORMATIONS ON SAFETY....................3 Introduction............................3 Symbols..............................3 Disclaimer.............................. 3 INTRODUCTION............................4 Component parts ..........................4 List of supplied elements........................5 Purpose ..............................5 Basic technical data ..........................5 ASSEMBLY............................6 Mechanical assembly of the flow sensor ....................6...
- SONIX reserves the right to change the product in order to increase its utility values or adjust it to current regulations. Such changes may not be taken into account in the current version of the manual.
Instructions for Assembly and Operating SONIX 10D INTRODUCTION The SONIX 10D flowmeter consists of a flow sensor and a measuring Component transducer connected together with probes cables. The flow sensor is parts available in three versions: 1. An U-shaped section ended with flanges or a screw thread (DN15...40)
- a flow sensor with two or four ultrasound probes including cables - Instructions for Assembly and Operating and a Guarantee Certificate - Calibration certificate for the flowmeters calibrated in the SONIX company flow stand Measurement of flow intensity and summing the volume of flowing liquids in Purpose closed, fully filled pipelines.
Access eye for removing solid impurities In the majority of cases, straight sections of the following lengths are Length of sufficient: straight - Inlet: 10D sections of the - Outlet: 3D Two- or more surface spatial configurations of the pipeline, the pump, not pipeline entirely closed valves, side inlets etc.
Instructions for Assembly and Operating SONIX 10D Recommendations Choose the longest available straight section of the pipeline. Install the ➢ sensor with probes on the 3/4 of its length starting from the inflow. for the flow sensor Make sure that in the selected spot the liquid will not be aerated and air or ➢...
Choice of the measuring spot on the pipeline and choice of measuring ➢ transducer location. The standard length of probes cables is 4 x 5m. In preparation for case of doubts, it is recommended to consult a SONIX representative. assembly Preparation of the pipeline for assembly: ➢...
Instructions for Assembly and Operating SONIX 10D Power supply conection - Protective earth conductor up to 2.5mm - Protective earth conductor up to 2.5mm - Protective earth conductor up to 2.5mm (13 AWG) has to be connected to terminal terminal...
Instructions for Assembly and Operating SONIX 10D Probe's cable correct 25 mm ending 55 mm OPERATING General The user communicates with the device via the LCD 2 x 1 6 characters display using two push buttons for choice of functions or parameters. The...
Time of correct work means the number of hours without signaling errors of Working time group I or II. Time of correct work Logo of the SONIX company. Factory number of the flowmeter. Phone number of SONIX’s service. SONIX Using the user’s functions does not influence flow measurement and volume User’s functions counting done at the same time by the flowmeter.
The counter may also be operated by the RS485 series link (see Protocol of series transmission). Servicing Service of the flowmeter is performed only by the SONIX servicing team.
Instructions for Assembly and Operating SONIX 10D ERRORS Operating flowmeters errors are divided into two groups: General I. Errors causing incorrect measurement and requiring immediate servicing. information The occurrence of such error is signaled by periodical blanking of the display every 0.3s. All outputs return to non-active states. Volume counting and the counter of correct work are blocked.
2.5V voltage. (see electrical connections of the flowmeter). The flowmeter is supplied with jumpers attached. The SONIX 10D flowmeter enables readings of temporary values of the flow Unit change and the content of counters in mass units (t/h, t and kg). In order to achieve...
Instructions for Assembly and Operating SONIX 10D List of registers Register Variable Name of the variable type float LO Flow m3/h ( t/h) long int HI Positive volume m3(t) Positive volume l(kg) float Liquid speed m/s errors 1... 16 reserved...
Instructions for Assembly and Operating SONIX 10D DECLARATION OF CONFORMITY The manufacturer: SONIX ® Industrial Electronic Devices ul. Leopolda Lisa-Kuli 12, 05-270 Marki declares that the following described product: Ultrasonic Flowmeter type SONIX10D is compliant with the requirements of the European Counsil directives no...
(e.g. the probes’ surface covering with deposit, presence of air in the pipeline, exceeding the allowed temperature limit for the ultrasound probes etc.). Name of the product: ULTRASOUND FLOWMETER, type: SONIX 10D Purchasing company:..................