Page 1
No. 169 Mujin Road, Hi-tech Park Ningbo, China Made in China Importato da / Imported by / Importé de / Importiert von / Importado de / Importado de: Gima S.p.A. Via Marconi, 1 - 20060 Gessate (MI) Italy gima@gimaitaly.com - export@gimaitaly.com www.gimaitaly.com...
6. Check the input voltage: be sure the input voltage which is signed in the back of the microscope is consistent with the power supply voltage, or it will bring a serious damage to the instrument. 7. Always use the power cord provided by Gima. 8. The electrical equipment of the microscope should be discard as electronic waste.
ENGLISH 7. Do not disassemble any parts of the microscope, as this will affect the function or reduce the perfor- mance of the microscope. 8. Place the instrument in a cool, dry position. When not using the microscope, keep it covered with a dust cover.
ENGLISH 2. Assembly 2.1 Assembly Diagram The following figure shows the installation sequence of the components. The number in the figure shows the assembly steps. • Before installing, be sure every components is clean, do not score any parts or glass surface. •...
ENGLISH 2.2 Assembly procedure Installing binocular viewing head (fig. 1-2) Insert the binocular viewing head into the microscope head and turn it to a proper position, then tight it. fig. 1 fig. 2 Installing the eyepiece (fig. 3-4) Insert the eyepiece into the eyepiece tube until they are against each other (fig. 4). fig.
Page 6
ENGLISH fig. 5 fig. 6 Connecting the Power Cord (fig. 7-9) The cable and cords are vulnerable when bent or twisted, never subject the power cord to ex- cessive force. 1. Set the main switch 1 to “0” (off) state before connecting the power cord. 2.
ENGLISH 3. Adjustment and Operation 3.1 Adjustment set diagram Portrait Adjustment Knob Aperture Iris Diaphragm adjustment Lateral Adjustment Knob Right Coarse Focus Knob Main Switch Tension Adjustment Collar Right Fine Focus Knob fig. 10...
Page 8
ENGLISH Interpupillar Distance Indicator Condenser Focusing Knob Left Fine Focus Knob Left Coarse Focus Knob Brightness Clocking Set Adjustment Knob fig. 11...
ENGLISH 3.2 Adjustment and Operation Brightness Adjustment (fig. 12-13) 1. Connect the power cord and set the main switch 1 to “ – ”state (ON). 2. Turning the brightness adjustment knob 2. fig. 12 fig. 13 Placing the Specimen (fig. 14) 1.
Page 10
ENGLISH Condenser Adjustment (fig. 16) Turn the condenser focus knob 1 to move the condenser up and down. Raise the condenser when using the high magnification objective, and descend it when using the low magnifica- tion one. 1. Focus the specimen with 10x objective. 2.
ENGLISH Adjusting the Diopter (fig. 19) 1. Turn the eyepiece 1 to adjust the diopter while looking through it. • The diopter range of the eyepiece is ±5 di- opter. The number aligned to the line on the viewing head is the diopter in use. fig.
ENGLISH 5. Troubleshooting Guide 5.1 Optical System TROUBLE CAUSE SOLUTION 1. The edge of the field of The nosepiece is not in the locat- Locate the nosepiece properly view is dark or the bright- ed position (objective and light where it clicks ness is not uniform path not coaxial) The surface of the lamp becomes...
Disposal: The product must not be disposed of along with other domestic waste. The users must dispose of this equipment by bringing it to a specific recycling point for electric and elec- tronic equipment. GIMA WARRANTY TERMS The Gima 12-month standard B2B warranty applies.