Calibration procedure
Table of Contents

Advertisement

3.2 Square Wave Adjustment

a. Referring to Fig. 8, connect output BNC to the oscilloscope has good grounding.
b. Connect the red input line to a 20 Vp-p, 10 kHz square wave signal source and the black input
line to ground.
c. Put SW1 at 1/100.
d. Set the VOLTS/DIV switch on oscilloscope's panel to 0.1V for 1/100 and the TIME/DIV
switch to 20 us.
e. Adjust VC1 to make the displayed waveform become as Fig. 9(b).
f. Connect the black input line to 20 Vp-p, 10 kHz square wave signal source and the red input
line to ground.
g. Adjust VC2 to make the displayed waveform become as Fig. 9(b).
Fig. 8
Fig. 9
5