Download Print this page
Nakanishi NSK V10 Instruction Manual

Nakanishi NSK V10 Instruction Manual

Plastic tip holder/maintenance plastic tip

Advertisement

Quick Links

Thank you very much for purchasing our V10 / V10-E / V10-S (Plastic tip holder) and V-P10 (plastic Tip) products. Read this
Instruction Manual carefully before use.
Precautions for Use
• This tip is designed specifically for the Ultrasonic System "Varios" and "VarioSurg" manufactured by NAKANISHI Inc.
Be sure to use it in a dedicated unit.
• This product is designed for dental use. Do not use it for any purpose other than dental care.
• Be sure to use it under sufficient quantities of water supply.
• Be sure to use the Tip power setting from a lower level according to the case.
• Be sure to check that the power level is set appropriately before use.
• Be sure to check that the Tip is fixed securely before use.
• The Tip is a consumable. Check it for the state of wear and replace it with a new one in good time. (For standard
replacement time, refer to the descriptions in the "5. Maintenance")
• Do not grind the Tip or change the angle by bending it. It may be broken or may not vibrate during use.
• Do not use a flawed, bent or rusted Tip. It may be broken during use.
• When the Tip is damaged such as by breaking during use, stop use immediately and replace it with a new one. The
damaged portion may injure the gums or periodontal tissues.
• Do not use the Tip for persons who have a cardiac pacemaker.
• Sterilization is required, after you purchase and each patient.
1. Product Names and Numbers
V10 : Z217041
Plastic tip holder for Varios
2. Features of V-P10 (Plastic Tip)
• This Tip is made by special plastic material and maintenance can be performed without damaging the surface of
a prosthetic appliance such as metal.
• Superior operability, with a curette type shape that easily fits between teeth.
• Economical Tip, easy to replace in case of breakage or wear.
• This Tip made of plastic material and it is different from conventional metal Tips. Be sure to use it at a specified
or lower power level.
3. Application
Removal of plaque built up on the surface of a prosthetic appliance such as metal.
4. Directions for Use
4-1 Mounting the Tip
When V-P10 (plastic Tip) is used, V10, V10-E or V10-S (Plastic tip
holder) and a wrench (E-Tip Wrench and CR-10 Tip Wrench) are
required.
1) Screw V10 / V10-E / V10-S into a handpiece by hand until it is
tightened lightly. Then, tighten it securely with the wrench (E-Tip
Wrench). (Fig. 1)
Plastic Tip Holder/Maintenance Plastic Tip
V10 / V10-E / V10-S / V-P10
Instruction Manual
V10-E : Z291020
Plastic tip holder for
Varios2 SHIN-HP/SHIN-LUX-HP
V10-S : Z305117
Plastic tip holder for VarioSurg
E-Tip Wrench
E-Tip Wrench
Holder
(V10 / V10-E / V10-S)
OM-Z0078E 005
V-P10 : Y900184
Plastic Tip
Tighten
Fig. 1
Fig. 1

Advertisement

loading

Summary of Contents for Nakanishi NSK V10

  • Page 1 Instruction Manual carefully before use. Precautions for Use • This tip is designed specifically for the Ultrasonic System “Varios” and “VarioSurg” manufactured by NAKANISHI Inc. Be sure to use it in a dedicated unit. • This product is designed for dental use. Do not use it for any purpose other than dental care.
  • Page 2 2) Insert V-P10 into holder ( V10 / V10-E / V10-S). Insert it into Holder holder until holder ( V10 / V10-E / V10-S) aligns with the (V10 / V10-E / V10-S) V-P10 indentation of the metal core of V-P10 (plastic Tip). (Fig. 2) CAUTION Appropriate vibration is not provided unless tip comes in to a dent part.

This manual is also suitable for:

Nsk v10-eNsk v10-sNsk v-p10Nsk z217041Nsk z291020Nsk z305117 ... Show all