Pressure probe

Advertisement

* unimportend if constant tempered and
sales@GlobalTestSupply.com
Probe
Meas.range
during long-term meas. *
0638.1345
± 100 Pa
0638.1445
±10 hPa
0638.1545
± 100 hPa
0638.1645
2 bar
reset before each measurement.
www.
Zero t.c. value
± (0.3 Pa + 0.5% of m.v.) 100 hPa
type 0.3% of f.v./°C
in range >10°C to 50°C
type 1% of f.v./°C
in range 0°C to 10°C
type 0.1% of f.v./°C
± 10 Pa (0 to 20 hPa)
type 0,05% of f.v./°C
in range 0°C to 20°C
Operating temperature...............................0 to +50 °C
Storage and transport temp. ..................-20 to +70 °C
The pressure probe is maintenance-free and is not
subject to a particular maintenance interval.
Clean the housing with a damp cloth.
Strong detergents or solvents should
never be used for cleaning the measuring
probe! Make sure that no liquid gets into the connec-
tion nipple!
Pressure sensor ±100 Pa ............................0638.1345
Pressure sensor ±10 hPa ............................0638.1445
Pressure sensor ±100 hPa ..........................0638.1545
Pressure probe 2bar abs .............................0638.1645
Magnetic holding fixture...............................0554.0225
Silicone hose ...............................................0554.0440
Pressure probe ............................................24 months
If the instrument is opened, improperly handled or
if force is applied, no warranty will be granted!
0973.2710/0402/T/wh/13.04.2004
Testo-Direct
Handling instructions
Technical data
Accuracy
Max.
overload
± 3 Pa
50 hPa
200 hPa
± 0.5% of m.v.
(20 to 100 hPa)
± 5 hPa
3200 hPa
Care and maintenance
Ordering data
Warranty