Instruction Manual
5342190-BG
Introduction
Thank you for purchasing the OMRON HEM-7120 Automatic Blood Pressure
Monitor.
The OMRON HEM-7120 is a compact, fully automatic blood pressure monitor,
operating on the oscillometric principle. It measures your blood pressure and
pulse rate simply and quickly. For comfortable controlled inflation without the
need of pressure pre-setting or re-inflation the device uses its advanced
"lntelliSense" technology.
Intended Use
This product is designed to measure the blood pressure and pulse rate of
people within the range of the designated arm cuff, following the instructions in
this instruction manual.
It is mainly designed for general household use.
Please follow this instruction manual thoroughly.
Please keep for future reference.
For specific information about your own blood pressure, CONSULT YOUR
DOCTOR.
Important Safety Information
A Warning: Indicates a potentially hazardous situation which, if not avoided,
could result in death or serious injury.
(General Usage)
• DO NOT adjust medication based on measurement results from this blood
pressure monitor. Take medication as prescribed by your doctor. Only a
doctor is qualified to diagnose and treat High Blood Pressure.
• Consult your doctor before using this monitor for any of the following
conditions: common arrhythmias such as atrial or ventricular premature beats
or atrial fibrillation, arterial sclerosis, poor perfusion, diabetes, age,
pregnancy, pre-eclampsia, renal diseases.
Note that PATIENT motion, trembling, shivering may affect the measurement
reading.
• Do not use this monitor on the injured arm or the arm under medical
treatment.
• Do not apply the arm cuff on the arm while being on an intravenous drip or
blood transfusion.
• Consult your doctor before using this monitor on the arm with an arterio­
venous (A-V) shunt.
• Do not use this monitor with other medical electrical (ME) equipment
simultaneously. This may result in incorrect operation of this monitor and/or
cause an inaccurate reading.
• Do not use this monitor in the area the HF surgical equipment, MRI, or CT
scanner exists, or in the oxygen rich environment. This may result in incorrect
operation of the monitor and/or cause an inaccurate reading.
• The air tube or the AC adapter cable may cause accidental strangulation in
infants.
• Contains small parts that may cause a choking hazard if swallowed by
infants.
• Stop using this monitor and consult your doctor if you experience skin
irritation or other troubles.
(AC Adapter (optional) Usage)
• Do not use the AC adapter if this monitor or the power cord is damaged. Turn
off the power and unplug the power cord immediately.
• Plug the AC adapter into the appropriate voltage outlet. Do not use a
multiple-tap.
• Never plug in or unplug the power cord from the electric outlet with wet
hands.
LL Caution: Indicates a potentially hazardous situation which, if not avoided,
may result in minor or moderate injury to the user or patient or
damage to the equipment or other property.
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results and
self-treatment are dangerous.
• Consult your doctor before using the monitor for any of the following
conditions:
- If you have had a mastectomy.
- People with severe blood flow problems or blood disorders as cuff inflation
can cause bruising.
• Do not take measurements more than necessary. It may cause bruising due
to blood flow interference.
• If there are any abnormalities during the measurement, remove the arm cuff.
• Do not use this monitor on infants or persons who cannot express their
intentions.
• Do not inflate the arm cuff more than necessary.
• Do not use this monitor for any purpose other than measuring blood
pressure.
• Use only the approved arm cuff for this monitor. Use of other arm cuffs may
result in incorrect measurement results.
• During measurement, make sure that no mobile phone or any other electrical
devices that emit electromagnetic fields is within 30cm of this monitor. This
may result in incorrect operation of the monitor and/or cause an inaccurate
reading.
• Do not disassemble the monitor or arm cuff. This may cause an inaccurate
reading.
• Do not use in a location with moisture, or a location where water may splash
on this monitor. This may damage the monitor.
• Do not use this monitor in a moving vehicle (car, airplane).
(Battery Usage)
• Do not insert the batteries with their polarities incorrectly aligned.
• Use only four "AA" alkaline or manganese batteries with this monitor. Do not
use other types of batteries. Do not use new and used batteries together.
• Remove the batteries if this monitor will not be used for 3 months or more.
• Use the battery within recommended period mentioned to it.
(AC Adapter (optional) Usage)
• Fully insert the power plug.
• When disconnecting the power plug, do not pull the power cord. Be sure to
hold the power plug.
• When handling the power cord, observe the following:
Do not damage.
Do not break it.
Do not tamper with it.
Do not forcibly bend or pull.
Do not twist.
Do not bundle during use.
Do not pinch.
Do not place under heavy objects.
• Wipe the dust off from the power plug.
• Disconnect the power plug if the product will not be used for a long period of
time.
• Disconnect the power plug before starting maintenance.
• Use only the original AC adapter designed for this monitor. Use of
unsupported adapters may damage and/or may be hazardous to this monitor.
General Precautions
• Do not forcibly bend the arm cuff or the air tube excessively.
• Do not fold or kink the air tube while taking a measurement. This may cause
harmful injury by interrupting blood flow.
• To unplug the air plug, pull on the air plug at the connection with the monitor,
not the tube itself.
• Do not apply strong shocks and vibrations to or drop the monitor and arm
cuff.
• Do not inflate the arm cuff when it is not wrapped around your arm.
• Read and follow the "Important information regarding Electromagnetic
Compatibility (EMC)" in the "6. Technical Data".
• Do not use this monitor outside the specified environment. It may cause an
inaccurate reading.
• Dispose of this monitor, components and optional accessories according to
applicable local regulations. Unlawful disposal may cause environmental
pollution.
• Please check (for example, by observation of the limb concerned) if this
monitor is not causing a prolonged impairment of PATIENT blood circulation.
• If this monitor is stored at the maximum or minimum storage and transport
temperature and is moved to an environment with a temperature of 20
recommend waiting for approx. 2 hours before using the monitor.
1. Overview
Monitor:
A. Display
C. Battery compartment
B. START/STOP button
D. AC adapter jack (for optional
AC adapter)
E. Air jack
Arm cuff:
F. Arm cuff (Arm circumference
22 - 32 cm)
G. Air plug
H. Air tube
I. Marker
Display:
J. Memory symbol
P. Irregular heartbeat symbol
K. Systolic blood pressure
Q. Movement error symbol
L. Diastolic blood pressure
R. Pulse display
M. Low battery symbol
S. Deflation symbol
N. Cuff wrapping guide
• Note: If your systolic or diastolic
0. Heartbeat symbol
pressure is outside the standard
1. Flashes during measurement.
range (above 135/85 mmHg)
2. If flashing after measurement
the Heartbeat symbol (•) will
completed or when viewing
blink. Please refer to Section
results stored in the memory,
3.3.
indicates blood pressure out of
recommended range•.
Before Taking a Measurement
To help ensure an accurate reading, follow these directions:
1. Avoid bathing, drinking alcohol or caffeine, smoking, exercising and
eating for 30 minutes before taking a measurement.
2. Rest for at least 5 minutes before taking the measurement.
3. Stress raises blood pressure. Avoid taking measurements during
stressful times.
4. Measurements should be taken in a quiet place.
5. Remove tight-fitting clothing from your arm.
2. Preparation
2.1 Installing/Replacing the Batteries
1.
Remove the battery cover.
2.
Insert four "AA" batteries
as indicated in the battery
compartment and then
replace the battery cover.
Notes:
□)
• If the low battery symbol (
appears on the display, tum the
monitor off then replace all batteries at the same time.
• The measurement values continue to be stored in memory even after
the batteries are replaced.
• The supplied batteries may have a shorter life.
• Dispose of this monitor, components and optional accessories
according to applicable local regulations. Unlawful disposal may
cause environmental pollution.
3. Using Your Monitor
3.1 Applying the Arm Cuff
Remove tight-fitting clothing or tight rolled up sleeve from your
upper arm. Do not place the arm cuff over thick clothes.
1.
Insert the air plug into the air jack
securely.
2.
Put your arm through the cuff
loop.
°
c, we
The bottom edge of the arm
cuff should be 1 to 2 cm
above the elbow. Marker
(arrow under the air tube) is
centred on the middle of
your inner arm.
3.
Close the fabric fastener
FIRMLY.
Notes:
• When you take a measurement on the right
arm, the air tube will be at the side of your
elbow. Be careful not to rest your arm on
the air tube.
• The blood pressure can differ between the right arm and the left arm,
and therefore also the measured blood pressure values can be
different. OMRON recommends to always use the same arm for
measurement. If the values between both arms differ substantially,
please check with your doctor which arm to use for your
measurement.
3.2 How to Sit Correctly
To take a measurement, you need to be
relaxed and comfortably seated at
comfortable room temperature.
• Sit on a chair with your legs uncrossed and
your feet flat on the floor.
• Sit with your back and arm being supported.
• The arm cuff should be placed on your arm
at the same level as your heart.
3.3 Taking a Measurement
Notes:
• To cancel a measurement, press the START/STOP button to release
the air in the arm cuff.
• Remain still and do not talk while taking a measurement.
1.
Press the START/STOP button.
The arm cuff will start to inflate automatically.
TSTART
TINFLATE
TDEFLATE
t
t
n
'18
u
0t
Cuff Wrapping Guide
If your systolic pressure is more than 210 mmHg
After the arm cuff starts to inflate, press and hold the START/STOP
button until the monitor inflates 30 to 40 mm Hg higher than your
expected systolic pressure.
Notes:
• The monitor will not inflate above 299 mmHg.
• Do not apply more pressure than necessary.
Note: Wait 2-3 minutes before taking another blood pressure
measurement. Waiting between readings allows the arteries to
return to the condition prior to taking the blood pressure
measurement.
LL Always consult your doctor. Self-diagnosis of measurement
results and self-treatment are dangerous.
Cuff Wrapping Guide:
The Cuff Wrapping Guide is a unique feature that indicates if the cuff
is not wrapped tightly enough around the arm. Even when theO)
appears on the display, a blood pressure reading will be taken.
Note: This reading is NOT reliable due to the incorrect wrapping of the
cuff. Please wrap the cuff again, taking care to wrap it correctly
and take the measurement again. When the@ is displayed, the
cuff is correctly wrapped tightly enough on the arm and the
reading is accurate and reliable.
2.
Remove the arm cuff.
3.
Press the START/STOP button to tum the monitor off.
The monitor automatically stores the measurement in its memory.
It will automatically turn off after 2 minutes.
Important:
• If your systolic or diastolic pressure is outside
the standard range, the heartbeat symbol will
blink when the measurement result is displayed.
Recent research suggests that the following
values can be used as a guide to high blood
pressure for measurements taken at home.
Systolic Blood Pressure
Diastolic Blood Pressure
This criteria is for home blood pressure measurement.
• Your blood pressure monitor includes an
irregular heartbeat feature. Irregular heartbeats
can influence the results of the measurement.
The irregular heartbeat algorithm automatically
determines if the measurement is usable or
needs to be repeated. If the measurement
results are affected by irregular heartbeats but the result is valid, the
result is shown together with the irregular heartbeat symbol (■). If
the irregular heartbeats cause the measurement to be invalid, no
result is shown. If the irregular heartbeat symbol (■) is shown after
you have taken a measurement, repeat the measurement. If the
irregular heartbeat symbol (Ill is shown frequently, please make
your doctor aware of it.
• If you move during measurement, the movement
error symbol (I.I) will appear on the display.
Keep still and repeat the measurement.
_b__J\__
3.4 Using the Memory Function
The monitor automatically stores the last measurement values (blood
pressure and pulse rate).
To View the Readings Stored in Memory
1.
Press and hold the START/STOP button for more than
5 seconds.
The last measurement value is displayed along with the memory
symbol.
Notes:
• If your systolic or diastolic pressure is outside the
standard range, the heartbeat symbol will blink
when the measurement result is displayed.
• The cuff wrapping guide result appears on the
display with the measurement values.
• If there are no measurements results stored in
the memory, the screen to the right is displayed.
2.
Press the START/STOP button to turn the monitor off.
It will automatically turn off after 2 minutes.
To Delete All Values Stored in Memory
Press and hold the START/STOP button for more than
15 seconds.
TCOMPLETE
t
'
152
18
'18
'l
0
Above 85 mmH