1. Overview
Main unit:
A
Automatic Blood
Pressure Monitor
B
Model HEM-7120
Instruction Manual
A. Display
B. START/STOP button
5342182-7A
Introduction
Arm cuff:
Thank you for purchasing the OMRON HEM-7120 Automatic Blood Pressure
Monitor.
I
F
The OMRON HEM-7120 is a compact, fully automatic blood pressure monitor,
operating on the oscillometric principle. It measures your blood pressure and
pulse rate simply and quickly. For comfortable controlled inflation without the
need of pressure pre-setting or re-inflation the device uses its advanced
G
"IntelliSense" technology.
H
Intended Use
This product is designed to measure the blood pressure and pulse rate of
Display:
people within the range of the designated arm cuff, following the instructions in
this instruction manual.
It is mainly designed for general household use. Please read the Important
J
Safety Information in this instruction manual before using the unit.
K
Please read this instruction manual thoroughly before using the unit.
Please keep for future reference.
For specific information about your own blood pressure, CONSULT YOUR
L
DOCTOR.
M
N
Important Safety Information
Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia
J. Memory symbol
or arteriosclerosis.
K. Systolic blood pressure
Please read this section carefully before using the unit.
L. Diastolic blood pressure
M. Low battery symbol
Warning: Indicates a potentially hazardous situation which, if not avoided,
N. Heartbeat symbol
could result in death or serious injury.
1. Flashes during measurement.
2. If flashing after measurement
(General Usage)
completed or when viewing
• Do not use the device on the injured arm or the arm under medical treatment.
results stored in the memory,
• Do not wrap the arm cuff on the arm while being on a intravenous drip or
indicates blood pressure out of
blood transfusion.
recommended range*.
(AC Adapter (optional) Usage)
• Do not use the AC adapter if the unit or the power cord is damaged. Turn off
the power and unplug the power cord immediately.
2. Preparation
• Plug the AC adapter into the appropriate voltage outlet. Do not use a
multiple-tap.
2.1 Installing/Replacing the Batteries
• Never plug in or unplug the power cord from the electric outlet with wet
hands.
1.
Remove the battery cover.
Caution: Indicates a potentially hazardous situation which, if not avoided,
2.
Insert four "AA" batteries
may result in minor or moderate injury to the user or patient or
damage to the equipment or other property.
as indicated in the battery
compartment and then
(General Usage)
replace the battery cover.
• Always consult your doctor. Self-diagnosis of measurement results and
self-treatment are dangerous.
Notes:
• People with severe blood flow problems, or blood disorders, should consult a
• If the low battery symbol (
doctor before using the unit, as cuff inflation can cause internal bleeding.
• If there are any abnormalities during the measurement, remove the arm cuff.
monitor off then replace all batteries at the same time.
• The measurement values continue to be stored in memory even after
• Do not use this device on infants or persons who cannot express their
the batteries are replaced.
intentions.
• The supplied batteries may have a shorter life.
• Do not inflate the arm cuff more than necessary.
• Do not use the unit for any purpose other than measuring blood pressure.
Disposal of used batteries should be carried out in accordance with the
• Use only the approved arm cuff for this unit. Use of other arm cuffs may result
national regulations for the disposal of batteries.
in incorrect measurement results.
• Do not use a mobile phone or other devices that emit electromagnetic fields,
near the unit. This may result in incorrect operation of the unit.
• Do not disassemble the unit or arm cuff.
3. Using the Unit
(Battery Usage)
• Do not insert the batteries with their polarities incorrectly aligned.
3.1 Applying the Arm Cuff
• Use only four "AA" alkaline or manganese batteries with this unit. Do not use
Remove tight-fitting clothing or tight rolled up sleeve from your
other types of batteries. Do not use new and used batteries together.
• Remove the batteries if the unit will not be used for 3 months or more.
upper arm. Do not place the arm cuff over thick clothes.
(AC Adapter (optional) Usage)
1.
Insert the air plug into the air jack
• Fully insert the power plug.
securely.
• When disconnecting the power plug, do not pull the power cord. Be sure to
hold the power plug.
• When handling the power cord, observe the following:
Do not damage.
Do not break it.
Do not tamper with it.
Do not forcibly bend or pull.
Do not twist.
Do not bundle during use.
Do not pinch.
Do not place under heavy objects.
• Wipe the dust of the power plug.
• Disconnect the power plug if the product will not be used for a long period of
time.
2.
Put your arm through the cuff
• Disconnect the power plug before starting maintenance.
loop.
• Use only the original AC adapter designed for this unit. Use of unsupported
adapters may damage and/or may be hazardous to the unit.
General Precautions
• Do not forcibly bend the arm cuff or bend the air tube excessively.
• To unplug the air plug, pull on the air plug at the connection with the main unit
not the tube itself.
• Do not apply strong shocks and vibrations to or drop the unit and arm cuff.
• Do not inflate the arm cuff when it is not wrapped around your arm.
The bottom edge of the arm
• Read and follow the "Important information regarding Electro Magnetic
cuff should be 1 to 2 cm
Compatibility (EMC)" in the Technical Data Section.
above the elbow. Marker
• Read and follow the "Correct Disposal of This Product" in the Technical Data
(arrow under the air tube) is
Section when disposing of the device and any used accessories or optional
centred on the middle of
parts.
your inner arm.
3.
Close the fabric fastener
FIRMLY.
C
D
E
Notes:
• When you take a measurement on the right
arm, the air tube will be at the side of your
C. Battery compartment
elbow. Be careful not to rest your arm on
D. AC adapter jack (for optional
AC adapter)
the air tube.
E. Air Jack
• The blood pressure can differ between the right arm and the left arm,
and therefore also the measured blood pressure values can be
different. OMRON recommends to always use the same arm for
measurement. If the values between both arms differ substantially,
please check with your doctor which arm to use for your
F. Arm Cuff (Arm circumference
measurement.
22 - 32 cm)
G. Air plug
H. Air tube
I. Marker
3.2 How to Sit Correctly
To take a measurement, you need to be
O
relaxed and comfortably seated, under
comfortable room temperature. No
P
bathing, drinking alcohol or caffeine,
smoking, exercising or eating 30 minutes
before taking a measurement.
• Sit on a chair with your feet flat on the floor.
Q
• Sit upright with your back straight.
• The arm cuff should be at the same level as
R
your heart.
O. Irregular heartbeat symbol
P. Movement error symbol
Q. Pulse display
R. Deflation symbol
* Note: If your systolic or diastolic
pressure is outside the standard
range (above 135/85 mmHg)
the Heartbeat symbol ( ) will
3.3 Taking a Reading
blink. Please refer to Section
3.3.
Notes:
• To cancel a measurement, press the START/STOP button to release
the air in the arm cuff.
• Remain still while taking a measurement.
1.
Press the START/STOP button.
The arm cuff will start to inflate automatically.
START
INFLATE
DEFLATE
COMPLETE
) appears on the display, turn the
If your systolic pressure is more than 210 mmHg
After the arm cuff starts to inflate, press and hold the START/STOP
button until the monitor inflates 30 to 40 mmHg higher than your
expected systolic pressure.
Notes:
• The monitor will not inflate above 299 mmHg.
• Do not apply more pressure than necessary.
Note: Wait 2-3 minutes before taking another blood pressure
measurement. Waiting between readings allows the arteries to
return to the condition prior to taking the blood pressure
measurement.
Always consult your doctor. Self-diagnosis of measurement
results and self-treatment are dangerous.
2.
Remove the arm cuff.
3.
Press the START/STOP button to turn the monitor off.
The monitor automatically stores the measurement in its memory.
It will automatically turn off after 2 minutes.
1 to 2 cm
Important:
4. Troubleshooting and Maintenance
• If your systolic or diastolic pressure is outside
4.1 The Icons and Error Messages
the standard range, the heartbeat symbol will
blink when the measurement result is displayed.
Error Display
Cause
Recent research suggests that the following
values can be used as a guide to high blood
pressure for measurements taken at home.
Irregular heartbeats are
Systolic Blood Pressure
Above 135 mmHg
detected.
Diastolic Blood Pressure
Above 85 mmHg
This criteria is for home blood pressure measurement.
Movement during
measurement.
• Your blood pressure monitor includes an
irregular heartbeat feature. Irregular heartbeats
can influence the results of the measurement.
The irregular heartbeat algorithm automatically
The batteries are low.
determines if the measurement is usable or
needs to be repeated. If the measurement
results are affected by irregular heartbeats but the result is valid, the
The batteries are
result is shown together with the irregular heartbeat symbol (
). If
exhausted.
the irregular heartbeats cause the measurement to be invalid, no
result is shown. If the irregular heartbeat symbol (
) is shown after
you have taken a measurement, repeat the measurement. If the
Air plug disconnected.
irregular heartbeat symbol (
) is shown frequently, please make
your doctor aware of it.
Arm cuff not applied
correctly.
• If you move during measurement, the movement
error symbol (
) will appear on the display.
Keep still and repeat the measurement.
Air is leaking from the arm
cuff.
Movement during
measurement and the arm
cuff has not been inflated
sufficiently.
3.4 Using the Memory Function
The arm cuff was inflated
The monitor automatically stores the last measurement values (blood
above 299 mmHg
pressure and pulse rate).
when inflating the cuff
manually.
To View the Readings Stored in Memory
1.
Press and hold the START/STOP button for more than
Movement during
measurement.
5 seconds.
The last measurement value is displayed along with the memory
symbol.
Clothing is interfering with
the arm cuff.
Device error.
Notes:
• If your systolic or diastolic pressure is outside the
4.2 Troubleshooting
standard range, the heartbeat symbol will blink
when the measurement result is displayed.
• If there are no measurements results stored in
Problem
Cause
the memory, the screen to the right is displayed.
Arm cuff not applied
correctly.
2.
Press the START/STOP button to turn the monitor off.
The reading is
It will automatically turn off after 2 minutes.
Movement or talking
extremely low (or
during measurement.
high).
Clothing is interfering
To Delete the Values Stored in Memory
with the arm cuff.
Press and hold the START/STOP button for more than
15 seconds.
The air tube is not
securely connected
Arm cuff pressure
into the main unit.
does not rise.
Air is leaking from the
arm cuff.
Arm cuff deflates too
The arm cuff is loose.
soon.
Cannot measure or
The arm cuff has not
readings are too low
been inflated
or too high.
sufficiently.
The batteries are
empty.
Nothing happens
when you press the
The batteries have
buttons.
been inserted
incorrectly.
• Press the START/STOP button and repeat
measurement.
• If the problem continues, try replacing the
Other problems.
batteries with new ones.
If this still does not solve the problem, contact
your local OMRON representative.
4.3 Maintenance
To protect your unit from damage, please observe the following:
• Do not subject the main unit and the cuff to extreme
Remedy
temperatures, humidity, moisture or direct sunlight.
Remove the arm cuff. Wait
• Do not fold the cuff or tubing tightly.
2 - 3 minutes and then take
• Do not inflate the arm cuff over 299 mmHg.
another measurement.
• Do not disassemble the unit.
Repeat the steps in section
• Do not subject the unit to strong shocks or vibrations
3.3. If this error continues to
appear, contact your doctor.
(for example, dropping the unit on the floor).
• Do not use volatile liquids to clean the main unit.
Carefully read and repeat
• Do not wash the arm cuff or immerse it in water.
the steps in section 3.3.
• Do not use petrol, thinners or similar solvents to clean
You should replace them
the arm cuff.
with new ones ahead of
time.
• Do not carry out repairs of any kind yourself. If a defect
Refer to section 2.1.
occurs, consult your local OMRON representative.
You should replace them
with new ones at once.
Refer to section 2.1.
Insert the plug securely.
Refer to section 3.1.
Apply the arm cuff correctly.
Refer to section 3.1.
Replace the cuff with the
new one.
Refer to Chapter 5.
Repeat measurement.
Remain still and do not talk
• The unit should be cleaned with a soft, dry cloth.
during measurement.
• Use a soft, moistened cloth and neutral soap to clean
Refer to section 3.3.
the arm cuff.
If "E2" appears repeatedly,
inflate the cuff manually
Note: Read and follow the "Correct Disposal of This Product" in the
until it is 30 to 40 mmHg
Technical Data Section when disposing of the device and any
above your previous
used accessories or optional parts.
measurement result.
Refer to section 3.3.
Calibration and Service
Do not inflate the cuff above
• The accuracy of this blood pressure monitor has been
299 mmHg.
carefully tested and is designed for a long service life.
Refer to section 3.3.
• It is generally recommended to have the unit inspected
every two years to ensure correct functioning and
Repeat measurement.
accuracy. Please consult your local OMRON
Remain still and do not talk
during measurement.
representative.
Refer to section 3.3.
Remove any clothing
4.4 Storage
interfering with the arm cuff.
Refer to section 3.1.
1.
Unplug the air plug from the air jack.
Contact your local OMRON
representative .
2.
Gently fold the air tube into the arm cuff.
Notes:
• Do not bend the air tube excessively.
• Do not store the unit in the following situations:
- If the unit is wet.
- Locations exposed to extreme temperatures, humidity, direct
sunlight, dust or corrosive vapours.
Remedy
- Locations exposed to vibrations, shocks or where it will be at an
angle.
Apply the arm cuff
correctly.
Refer to section 3.1.
5. Optional Parts
Remain still and do not
talk during
Medium Arm Cuff
Large Arm Cuff
AC Adapter S
measurement.
Arm circumference
Arm circumference
Refer to section 3.3.
22 - 32 cm
32 - 42 cm
Remove any clothing
interfering with the arm
cuff.
Refer to section 3.1.
Make sure that the air
tube is connected
securely.
Note: Please check with your local OMRON representatives for the
Refer to section 3.1.
appropriate optional part models.
Replace the arm cuff
with a new one.
Using the Optional AC Adapter
Refer to Chapter 5.
1.
Insert the AC adapter plug
Apply the cuff correctly
so that it is firmly
into the AC adapter jack on
wrapped around the arm.
the rear side of the main
Refer to section 3.1.
unit.
Inflate the cuff so that it
2.
Plug the AC adapter into an
is 30 to 40 mmHg above
electrical outlet.
your previous
measurement result.
To disconnect the AC adapter, unplug the AC adapter from the
Refer to section 3.3.
electrical outlet first and then remove the AC adapter plug from the
Replace the batteries
main unit.
with new ones.
Refer to section 2.1.
Insert the batteries with
the correct (+/-) polarity.
Refer to section 2.1.
HEM-7120-IN_A_M03_130621.pdf
6. Technical Data
Product Description
Automatic Blood Pressure Monitor
Model
HEM-7120
Display
LCD Digital Display
Measurement Method
Oscillometric method
Measurement Range
Pressure: 0 to 299 mmHg
Pulse: 40 to 180 beats/ min.
Accuracy
Pressure: ±3 mmHg
Pulse: ±5% of display reading
Inflation
Fuzzy-logic controlled by electric pump
Deflation
Automatic pressure release valve
Memory
Last Measurement
Rating
DC6V 4W
Power Source
4 "AA" batteries 1.5V
or AC adapter (optional, INPUT AC100-240V
50/60Hz 0.12A)
Battery Life
Approx. 1000 measurements
(using new alkaline batteries)
Applied Part
= Type BF
Protection Against
Internally powered ME equipment (When using only
Electric Shock
the batteries)
= Class II ME equipment (Optional AC adapter)
Operating
+10 to +40°C / 30 to 85% RH
temperature/ Humidity
Storage temperature/
-20 to +60°C / 10 to 95% RH / 700-1060 hPa
Humidity/ Air pressure
Console Weight
Approx. 250g without batteries
Cuff Weight
Approx. 130g
Outer Dimensions
Approx. 103 (w) mm × 80 (h) mm × 129(l) mm
Cuff Dimensions
Approx. 145 mm × 466 mm
(Cuff: arm circumference 22 to 32 cm)
Cuff/ Tube Material
Nylon, polyester, polyvinyl chloride
Package Content
Main unit, arm cuff, instruction manual, battery set
Note: Subject to technical modification without prior notice.
• This device fulfils the provisions of EC directive 93/42/EEC (Medical Device
Directive).
• This blood pressure monitor is designed according to the European Standard
EN1060, Non-invasive sphygmomanometers Part 1: General Requirements
and Part 3: Supplementary requirements for electromechanical blood
pressure measuring systems.
• This OMRON product is produced under the strict quality system of OMRON
HEALTHCARE Co., Ltd., Japan. The core component for OMRON blood
pressure monitors, which is the Pressure Sensor, is produced in Japan.
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC's and mobile
(cellular) telephones, medical devices in use may be susceptible to
electromagnetic interference from other devices. Electromagnetic interference
may result in incorrect operation of the medical device and create a potentially
unsafe situation.
Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility)
with the aim to prevent unsafe product situations, the EN60601-1-2:2007
standard has been implemented. This standard defines the levels of immunity to
electromagnetic interferences as well as maximum levels of electromagnetic
emissions for medical devices.
This medical device manufactured by OMRON HEALTHCARE conforms to this
EN60601-1-2:2007 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
• Do not use mobile (cellular) telephones and other devices, which generate
strong electrical or electromagnetic fields, near the medical device. This may
result in incorrect operation of the unit and create a potentially unsafe situation.
Recommendation is to keep a minimum distance of 7 m. Verify correct operation
of the device in case the distance is shorter.
Further documentation in accordance with EN60601-1-2:2007 is available at
OMRON HEALTHCARE EUROPE at the address mentioned in this instruction
manual.
Documentation is also available at www.omron-healthcare.com.
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
This marking shown on the product or its literature, indicates that it
should not be disposed of, with other household wastes at the end
of its working life. To prevent possible harm to the environment or
human health from uncontrolled waste disposal, please separate
this product from other types of wastes and recycle it responsibly to
promote the sustainable reuse of material resources.
Household users should contact either the retailer where they purchased this
product, or their local government office, for details of where and how they can
return this item for environmentally safe recycling.
Business users should contact their supplier and check the terms and conditions
of the purchase contract. This product should not be mixed with other commercial
wastes for disposal.
Manufacturer
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto
617-0002 JAPAN
OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp THE NETHERLANDS
www.omron-healthcare.com
OMRON HEALTHCARE SINGAPORE PTE LTD.
Asia Pacific HQ
438A Alexandra Road, #05-05/08
Alexandra Technopark, Singapore 119967
www.omron-healthcare.com.sg
Production facility OMRON HEALTHCARE MANUFACTURING VIETNAM
CO., LTD.
Binh Duong Province, VIETNAM
Made in Vietnam