Bubble point (BP) of the soil ....................14 Soil Tensiometer ........................15 Infiltration Measurement with the tension infiltrometer............15 Calculation of hydraulic conductivity ................... 16 Micrologger IL-2700 ......................17 Features ..........................17 Technical Data ........................17 General guidelines for operation ..................18 Measurement ........................
The effective increase in pressure on the soil surface can be freely selected between zero and a vacuum up to the bubble point of the soil. There are several ways to contact UGT if you need a quote or product information please contact Email: info@ugt-online.de...
2. Introduction The Hood Infiltrometer IL-2700 (Fig. 1) is used for measuring the soil hydraulic conductivity near saturation in field tests. Infiltration starts from a closed hood filled with water and stand- ing on the ground. The circular surface under the hood is directly covered with water and acts as the source area of infiltration flow.
However, infiltration from different source areas makes sense only in largely homogeneous soils. Umwelt-Geräte-Technik GmbH | Eberswalder Str. 58 | D-15374 Müncheberg | Tel.: +49 (0) 33 43 2 - 89 575 | www.ugt-online.de Page 7 of 32...
In such experiments, equation (11) should be used only for representative means of the infiltration rates of the infiltration hoods under review. Umwelt-Geräte-Technik GmbH | Eberswalder Str. 58 | D-15374 Müncheberg | Tel.: +49 (0) 33 43 2 - 89 575 | www.ugt-online.de Page 9 of 32...
MARIOTTE water supply system from where the reading can be taken of a scale (filling height). Umwelt-Geräte-Technik GmbH | Eberswalder Str. 58 | D-15374 Müncheberg | Tel.: +49 (0) 33 43 2 - 89 575 | www.ugt-online.de Page 10 of 32...
Page 11
Valve for pressure adjustment Vent pipe Filling height of the infiltration reservoir Hose connection 2 Stand Umwelt-Geräte-Technik GmbH | Eberswalder Str. 58 | D-15374 Müncheberg | Tel.: +49 (0) 33 43 2 - 89 575 | www.ugt-online.de Page 11 of 32...
“I” following close it with the plug. If using a Hood Infiltrometer IL-2700 with pressure sensor and capturing the data with the handheld logger it is advisable to calibrate the pressure sensor before the test run. Therefor filling the infiltration reservoir should be carried out as last step and the calibration process at the handheld device needs to be started before filling the infiltration reservoir.
The scale values at the hood (Hs) and at both brackets of the U-tube manometer “Us” should be recorded. Umwelt-Geräte-Technik GmbH | Eberswalder Str. 58 | D-15374 Müncheberg | Tel.: +49 (0) 33 43 2 - 89 575 | www.ugt-online.de Page 13 of 32...
1) Us = Us(le) + Us(ri) (left/right limb of the U-tube manometer) In addition to the manual recording the data for the leakage rate can also be recorded dig- ital with the Micrologger IL-2700 ( Chapter 8 Each test has to be run up to a steady-state infiltration rate. Depending on the type of soil involved, infiltration to mid-column will be required.
Herefore up to three soil tensiom- eters can additionally be installed at a shallow depth. The soil tensiometers as well as the appropriate outer ring are available at UGT-GmbH. 7. Infiltration Measurement with the tension infiltrometer To determine a representative ku-function a hood Infiltrometer test from tension “zero”(kf-...
Infiltration area at Table 2 describes the area within the ring, because the sand between ring and hood has usually a high conductivity. Umwelt-Geräte-Technik GmbH | Eberswalder Str. 58 | D-15374 Müncheberg | Tel.: +49 (0) 33 43 2 - 89 575 | www.ugt-online.de Page 16 of 32...
9. Micrologger IL-2700 9.1 Features The UGT-Micrologger IL-2700 captures the discharge rate out of the infiltration reservoir „ΔZ/Δt“at different tensions at one test site (series of measurement). It is possible to rec- ord up to 8 tension steps. 9.2 Technical Data...
RS 232 (to PC) Fig. 5: Micrologger IL-2700 9.3 General guidelines for operation A complete specification of the menu structure and the functions of this handheld device at the chart at subsection “4.8 menu structure” as well as on the appositional laminated leaflets.
If the moving average is set to 1 the flow rate is not checked for constant conditions and the maximum number of values (100) is recorded. Umwelt-Geräte-Technik GmbH | Eberswalder Str. 58 | D-15374 Müncheberg | Tel.: +49 (0) 33 43 2 - 89 575 | www.ugt-online.de Page 19 of 32...
The measurement interval is defined before every measurement. To get to this point proceed as shown in the scheme on page 18 to 22. First of all the handheld device IL-2700 is switched on pushing the “On”-button. The measurement site is chosen by pushing the “Step”-button accordant times and confirming it pushing the “Start”-button.
PC and the diagnostics menu can be found here. 9.9 Menu structure Umwelt-Geräte-Technik GmbH | Eberswalder Str. 58 | D-15374 Müncheberg | Tel.: +49 (0) 33 43 2 - 89 575 | www.ugt-online.de Page 21 of 32...
The diagnostics menu allows troubleshooting while ongoing operation. It is particularly in- tended for maintenance work by UGT-staff and is hence not further explained in this manual. Umwelt-Geräte-Technik GmbH | Eberswalder Str. 58 | D-15374 Müncheberg | Tel.: +49 (0) 33 43 2 - 89 575 | www.ugt-online.de...
“IL2700””Options (COM Port)” and try again to establish a con- nection. The handheld device remains in communication mode all the time. Umwelt-Geräte-Technik GmbH | Eberswalder Str. 58 | D-15374 Müncheberg | Tel.: +49 (0) 33 43 2 - 89 575 | www.ugt-online.de Page 28 of 32...
Page 29
Generally it is recommended to delete all old data before each new use to guar- antee a maximum on storage capacity and avoid a mix up of the data. Umwelt-Geräte-Technik GmbH | Eberswalder Str. 58 | D-15374 Müncheberg | Tel.: +49 (0) 33 43 2 - 89 575 | www.ugt-online.de Page 29 of 32...
Page 30
The data point, from which on a constant flow as signal for steady state is meas- ured is highlighted. Umwelt-Geräte-Technik GmbH | Eberswalder Str. 58 | D-15374 Müncheberg | Tel.: +49 (0) 33 43 2 - 89 575 | www.ugt-online.de Page 30 of 32...
In case this is necessary in the course of an infiltration meas- urement, shut [V1] when filling. Micrologger IL-2700: Check the charge of the battery in the logger Remove the batterys from the logger for longer measuring breaks ...