D.  Power Status Indicator 
Green AC only  
Green AC and flashing Yellow DC  
Yellow DC only  
Yellow DC and flashing Red ATTN 
 
Green AC and flashing Red ATTN 
 
E.  Manual CPR Release 
In the event of a medical emergency, activation of the manual CPR release handle will mechanically flatten the head section (backrest) of 
the mattress deck allowing staff to administer CPR to the patient. The knee/foot sections will also be automatically flattened. 
Ac va
ng Manual CPR Release:  To fla
 
Rese
ng the Head Actuator: The head actuator MUST be reset every time the manual CPR handle is pulled. The bed will automatically 
reset the head actuator once the backrest has been completely flattened. The LCD display will display a message notifying staff that the 
CPR reset procedure is in progress. In the event of an incomplete manual CPR release or interrupted reset procedure, a message will be 
displayed prompting staff to PUSH and HOLD the CPR button or to PULL UP and HOLD the manual CPR handle to complete the CPR reset 
procedure. An audible "chirp" will sound to notify staff that the CPR reset procedure has been successfully completed. 
 
F.    Scale / LCD Display 
Press the ON/ENTER bu on to "wake up" the Home Screen to show Pa ent Weight . Press the UP/DOWN bu ons to scroll between 
the addi onal Home Screens which show Head & Trend angles and warnings (if any).  
 
Scenario 
Before each new pa ent is placed in the bed 
Before adding clinically necessary equipment, addi onal 
linens, pillows, etc 
 
MENU NAVIGATION: 
 Use the UP/DOWN arrow bu ons to scroll through the following menus.  
 Press the ON/ENTER bu on to make selec on 
 Press the 0.0/ZERO bu on to return back or "escape" to the home screen at any  me from any menu op on screen. 
 
 
Bed is plugged into wall and the ba ery is fully charged 
Bed is plugged into wall and the ba ery is charging  
Bed is running on ba ery power 
Bed is running on ba ery power but has used up the ba ery reserve power. Plug bed into 
wall and allow ba ery to charge for 24hr 
Bed is plugged into wall but there is a problem with ba ery or another component. 
Confirm ba ery is properly connected and/or service bed immediately 
en the backrest, PULL UP and HOLD  on the CPR handles (located on either side of the bed). 
CHG Hospital Beds • 1‐866‐516‐5446 • www.chgbeds.com 
Recommended Ac on 
Zero the scale. This will automa cally reset adjusted 
pa ent weight and bed exit alarm history. 
Use the Change Equipment feature. This will allow any 
changes to not impact displayed pa ent weight. 
 
 