D. Power Status Indicator
Green AC only
Green AC and flashing Yellow DC
Yellow DC only
Yellow DC and flashing Red ATTN
Green AC and flashing Red ATTN
E. Manual CPR Release
In the event of a medical emergency, activation of the manual CPR release handle will mechanically flatten the head section (backrest) of
the mattress deck allowing staff to administer CPR to the patient. The knee/foot sections will also be automatically flattened.
Ac va
ng Manual CPR Release: To fla
Rese
ng the Head Actuator: The head actuator MUST be reset every time the manual CPR handle is pulled. The bed will automatically
reset the head actuator once the backrest has been completely flattened. The LCD display will display a message notifying staff that the
CPR reset procedure is in progress. In the event of an incomplete manual CPR release or interrupted reset procedure, a message will be
displayed prompting staff to PUSH and HOLD the CPR button or to PULL UP and HOLD the manual CPR handle to complete the CPR reset
procedure. An audible "chirp" will sound to notify staff that the CPR reset procedure has been successfully completed.
F. Scale / LCD Display
Press the ON/ENTER bu on to "wake up" the Home Screen to show Pa ent Weight . Press the UP/DOWN bu ons to scroll between
the addi onal Home Screens which show Head & Trend angles and warnings (if any).
Scenario
Before each new pa ent is placed in the bed
Before adding clinically necessary equipment, addi onal
linens, pillows, etc
MENU NAVIGATION:
Use the UP/DOWN arrow bu ons to scroll through the following menus.
Press the ON/ENTER bu on to make selec on
Press the 0.0/ZERO bu on to return back or "escape" to the home screen at any me from any menu op on screen.
Bed is plugged into wall and the ba ery is fully charged
Bed is plugged into wall and the ba ery is charging
Bed is running on ba ery power
Bed is running on ba ery power but has used up the ba ery reserve power. Plug bed into
wall and allow ba ery to charge for 24hr
Bed is plugged into wall but there is a problem with ba ery or another component.
Confirm ba ery is properly connected and/or service bed immediately
en the backrest, PULL UP and HOLD on the CPR handles (located on either side of the bed).
CHG Hospital Beds • 1‐866‐516‐5446 • www.chgbeds.com
Recommended Ac on
Zero the scale. This will automa cally reset adjusted
pa ent weight and bed exit alarm history.
Use the Change Equipment feature. This will allow any
changes to not impact displayed pa ent weight.