Cleaning the BIOPAC TSD180/181 probe connector Always clean the connector ferrule end face with the included connector cleaner before each connection. Failure to do so can result in permanent damage to the connector caused by hard particles trapped between the fiber optic end faces. Step 1 Step 2 Softly press connector on a pristine portion...
Page 3
Cleaning the conditioner connector Step 1 Provided in the cleaning kit, swab sticks are required to insure a proper cleaning which are available at most electronic stores. Avoid using cotton swabs. Take a swab and wet it with a few drops of isopropanol - avoid acetone.
Page 4
How to prevent damage and ensure a long life to your BIOPAC TSD180/181 probes Tips to prevent probes damages and ensure long probe life: Keep this page as a reminder at your application location. • Keep the fiber-optic connector surface clean – adopt the habit of cleaning the connector at each use.