General information on electrostatics General information on electrostatics Nowadays ESD is a problem at many workstations, because modern microelectronics is easily destroyed by the sequels of ESD. Other branches of industry like e.g. telecommunications-, plas- tics-, and explosive material industry are also heavily affected by ESD. ESD causes losses of time as well as high financial losses and can endanger the human health.
Caption Caption Cover cap 5 Spacer Modulator system 6 Zero point trimmer Button «function/on» 7 Alphanumeric LCD display Grounding jack 8 Battery case (on the back)
Manual EFM 022 V5.3 Operating instructions Measuring principle The electrostatic field meter is a parametric amplifier. The electric field influences a current proportional to the electrostatic field. The current is amplified and measured with a selective amplifier. No energy is taken from the field over time means.
Operating instructions Measurement ranges After switching the device on the distance selected is at 2cm. If you want to measure with that distance you just have to place the device 2cm in front of the object to be measured. For most cases this distance is ideal because charges up to 20kV can be measured (see specifications).
Manual EFM 022 V5.3 Changing the measurement distance Press «function/on» until the lower line shows „change distance“ (approx. 2 sec). In the upper line the current distance is shown in cm. By pressing «function/on» the distance can be changed in the following order:...
Maintenance Display The display is a double-spaced alphanumeric LCD display with 12 digits in each row (2 x 12). In the upper line the selected distance is displayed in cm. In the bottom line the measured charge is displayed in Volt. From 999V on the display changes to kV display. The displayed value is always triple-digit: Examples: 578V...
Manual EFM 022 V5.3 Zero Point Under normal circumstances a zero point adjustment is not necessary. If the device does not show U=000V while the modulator system is screened (e.g. mounted cover cap) the zero point can be adjusted using the trimmer (6) on the right side of the device. The last digit of the display may be ignored as that error is smaller than the admissible tolerance.
It has the same function as the EFM 022 with an additional analoque interface ±1V It is possible to upgrade an EFM 022 to an EFM 023 ! Together with our USB – Analoque/Digital Converter UAC 110 you can connect the instrument to a personal computer to save, change and display the results.
Manual EFM 022 V5.3 MK 023 -measuring head- VMS 022 – Volt Meter Set- with MK 023, hand electrode up to ±4kV and silicon cable Head to measure DC-Voltage with an to do a walking test Input impedance from 10 Warning Notices ...