Page 1
Manual Elektrofeldmeter EFM 115 Small E - Field meter with high sensitivity to measure static DC-fields and electrostatic charge according the field mill-influenz-principle with USB – Interface to connect to a PC etc.
Page 2
Handbuch EFM 115 Contents General information of the principle of electrostatics ..................3 What is the result of electrostatic charge ........................ 3 Description ................................... 3 Technical Data ................................4 Legende ..................................... 4 Operating instructions ..............................4 Measurement method ..............................4 Field of application ...............................
Page 3
General information of the principle of electrostatics General information of the principle of electrostatics Electrostatic discharging is a today problem on many working places with the use of modern micro-electronics as e.g. Microchips and their sensibility against it. There is also a very strong danger for industrial branches as e.g. Telecommunication-, Plastics-, and manufactures of explosives.
Page 4
Handbuch EFM 115 Technical Data Dimension (L x B x H): app. D36mm x 120mm Weight: app. 180g Calibration in a plate capacitor: 200mm x 200mm, Distance 25mm Ranges: 5kV/m, 20kV/m, 50kV/m und 200kV/m Tolerance: < 5% Power supply: 5V max 100mA by USB Interface...
Page 5
Calibration Zero adjust First put on the red cap onto the modulator system. Then press in the EFM115_ReadOut Software the “Zero Adjust” button. If an error message appears, you should make a Zero adjust by the potentiometer (4) on back of the unit. Adjust the measure value < 50mV and press the “ok” button at the software.
Page 6
By high voltage measurements the directives of the standard EN 10100 are be followed. For using the MK1 mount it on the modulator system of the EFM 115. At the voltage head MK11 there is a PTFE isolate field plate in 1cm arranged.
Page 7
Grounding Grounding The appliance must be grounded to get real measuring results. You can connect ground by the 4mm socket on the back. Maintenance Don´t touch the modulator system! Time by time the modulator-system needs to be cleaned and a new offset – adjustment with shielded field-plate is necessary.
Page 8
Handbuch EFM 115 Parts of delivery The following parts belong to basic equipment: • Electro – Field - Meter EFM 115 • shield cap • USB – cable with Mini plug 1,5m • Spiral ground cable 2 m with alligator clip •...