Page 1
Manual System Prof. Dr.-Ing. Hans Kleinwächter Electro field meter EFM 113B Small stationary static-field meter with high sensitivity to measure static DC-fields and electrostatic charge according the field mill-influenz-principle. Output: analogue current-signal ± 1mA...
Manual EFM 113B Contents General information of the principle of electrostatics ..................3 What is the result of electrostatic charge ........................ 3 Description ................................... 3 Technical Data ..................................4 Legend ....................................4 Operating instructions ..............................5 Measurement method ..............................5 Field of application ...............................
General information of the principle of electrostatics General information of the principle of electrostatics Electrostatic discharging is a today problem on many working places with the use of modern micro-electronics as e.g. Microchips and there sensibility against it. There is also a very strong danger for industrial branches as e.g. Telecommunication- , Plastics- , and manufactures of explosives.
Manual EFM 113B Technical Data Dimension (L x B x H): app. D36mm x 120mm Weight: app. 180g Calibration in a plate capacitor: 200mm x 200mm, Distance 25mm Ranges: 5kV/m, 20kV/m, 50kV/m und 200kV/m Tolerance: < 5% Power supply: 9-15V DC / ca. 60mA Operating-time : By 8 hours a day min.
Operating instructions Operating instructions Measurement method The static-field meter has a parametric amplifier. The electrical field causes an alternated current proportional to the field strength. These current can be measured with a selective amplifier, without taking away energy of the electrical field. Through using gold-plated influenz-electrodes no galvanic disturbances are induced.
Manual EFM 113B Pinning Pin 1 = Range select B Pin 2 = Power supply + 9-15V DC Pin 3 = Ground (GND) Pin 4 = Range select A Pin 5 = Current output ±1mA The maximum cable length (0,14 mm²) is 50m.
Grounding Grounding The appliance must be grounded to get real measuring results. You can connect ground by the access line (pin3) or over the optional aluminum clamp Maintenance Don´t touch the modulator system ! Time by time the modulator-system needs to be cleaned and a new offset – adjustment with shielded field-plate is necessary.