Safety notes Please read this manual carefully and completely before you use the device for the first time. The device may only be used by qualified personnel and repaired by PCE Instruments personnel. Damage or injuries caused by non-observance of the manual are excluded from our liability and not covered by our warranty.
Page 4
• Pathogenic substances should always be processed in a closed container and using a suitable extraction hood. • Only use media that you know will not react dangerously to the additional energy generated during processing. Additional energy can also be generated by light irradiation, a certain ambient temperature or other factors.
195 x 255 x 155 mm / 7.7 x 10 x 6.1 in Weight 4.45 kg / 9.8 lbs Delivery scope 1 x magnetic stirrer PCE-MSR 340 1 x device connection cable 1 x temperature sensor 1 x holder 1 x instruction manual Preparation •...
Operation Power on Switch on the magnetic stirrer via the main switch which is located at the back of the device, on the left-hand side. Attention! The stirring platform heats up when the temperature sensor is inserted and the temperature is set.
Attention! Warning! Always reset the temperature setting to 000.0 °C after completion of the stirring process. The temperature sensor is also used as a reference sensor for temperature control of the stirring platform. If the temperature sensor is not in the stirring medium on the stirring platform when the magnetic stirrer is switched on and the temperature measured by the temperature sensor is below the specified temperature, the stirring platform is directly heated up to the maximum temperature.
Warranty You can read our warranty terms in our General Business Terms which you can find here: https://www.pce-instruments.com/english/terms. Disposal For the disposal of batteries in the EU, the 2006/66/EC directive of the European Parliament applies. Due to the contained pollutants, batteries must not be disposed of as household waste. They must be given to collection points designed for that purpose.
Page 9
PCE Instruments contact information Germany France Spain PCE Deutschland GmbH PCE Instruments France EURL PCE Ibérica S.L. Im Langel 4 23, rue de Strasbourg Calle Mayor, 53 D-59872 Meschede 67250 Soultz-Sous-Forets 02500 Tobarra (Albacete) Deutschland France España Tel.: +49 (0) 2903 976 99 0 Téléphone: +33 (0) 972 3537 17 Tel.