Electronic switch for voltage transient testing (21 pages)
Summary of Contents for EMTEST CWS 500N1.3
Page 1
100 kHz to 300 MHz IEC 61000-4-6 / EN 61000-4-6 IEC 61000-4-6 The CWS 500N1.3 and CWS 500N1.4 are compact EN 61000-4-6 single box test equipment for testing conducted RF immunity per IEC 61000-4-6 Ed.3 and Ed.4 and related standards.
8.1. 8.2. Test Setup with injection clamps (EM and BCI) ..................37 Test Equipment CWS 500N1.x ....................... 38 9.1. Design of CWS 500N1.3 series ......................38 9.2. Design of CWS 500N1.4 series ......................39 9.3. Control Unit ............................. 39 9.4.
Page 4
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 Maintenance ............................44 11.1. General ..............................44 11.2. Calibration and Verification ........................44 11.2.1. Factory calibration ........................... 44 11.2.2. Guideline to determine the calibration period of EM Test instrumentation ..........44 11.2.3. Calibration of Accessories made by passive components only: ............. 44 11.2.4.
CWS500 N1.3 / N1.4 Model Overview This Manual describes the continuous frequency generator models: - CWS 500N1.3 Generator with frequency range from10 kHz to 400 MHZ and - CWS 500N1.4 Generator with frequency range from100 kHz to 300 MHZ User manual V 1.01...
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 Safety 2.1. Safety Aspects Tests performed with RF-disturbances, induced by radio-frequency fields, are immunity tests on electronic equipment. Therefore, it is the responsibility of the user to avoid critical failures and risks to the environment and the operators.
3. Before a test or a calibration may performed, the generator must be operating for warm up - CWS 500N1.3 for at least 15 minutes and - CWS 500N1.4 for at least 30 minutes. After this time the CWS 500N1.x output is stable 4.
Bridge SG OUT – AMP IN Safety Circuit Voltage selector Main Switch Interface Fig 3.2 : CWS 500N1.3 and CWS 500N1.4 rear side Connection bridge Connect the delivered N-connector bridge to the plug COUPLER IN – AMP OUT Fig 3.3 Bridge SG OUT –...
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 3.2. Part Identifications and Functions Continuous Wave Simulator CWS 500N1.3 or CWS 500N1.4 Fig 3.5 Attenuator ATT6/80 6 dB / 80W N/BNC connector Fig 3.6 Attenuator 20 dB / 1W Fig 3.7...
Page 10
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 Calibration CWS-Cal Basic Calibration Kit Transport Case 1x R100N 150 to 50 Adapter 1x BNC cable 0.5m Cal adapter according to CDN , Fig 3.12 Calibration CA EM Ed.4 Basic Calibration Kit 1 x Transport Case 2 x R100N1 150...
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 4.2. Rear View Figure 4.2: Rear side CWS 500N1.3 1 Bridge AMP OUT – Coupler IN 6 Power selector 115V – 230V 2 Air output 7 IEEE Interface 3 Bridge SG OUT – AMP IN...
Page 13
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 Figure 4.3: Rear side CWS 500N1.4 1 Bridge AMP OUT – Coupler IN 6 Power selector 115V – 230V 2 Air output 7 IEEE Interface 3 Bridge SG OUT – AMP IN 8 USB Serial Interface 4 Safety Circuit 9 Remote Control Connector...
The latest settings are stored automatically and will be recalled when the menu is selected again. EXIT: The firmware will reset to the main screen. Fig 5.2 Start-up display CWS 500N1.3 Generator Model Continuous Wave Simulator Frequency Ranges RF 9 kHz – 1 GHz V 8.50...
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 Page 2 (Test Routines) TEST ROUTINES F1 : Quickstart F2 : User test routines F3 : Standard test F1 Quickstart Menu offers easy and fast operation, with individual parameter settings. F2 User Test Routines The user saves and recalls individual specific test routines.
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 5.2. Quickstart Easy and very fast operation of all standard functions of the equipment. Page 3 (Show Parameters) Page 4 (Change) Quickstart Quickstart = 10,0V Mod = 1kHz = 5.0MHz Cal = F1 CP = F-33-2 START CHANGE 10.0...
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 5.3. User Test Routines The user can change, save and recall own specific test routines based on IEC 61000-4-6. All special functions and routines are stored in this part of the user menu. Page 3 (Selection of the function) Page 4 (Select store)
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 5.3.3. Dwell Time Sweep The dwell time changes from td1 to td2. After the dwell time td and the pause time tr, the dwell time is changed by td until the set value for td2 is reached. Page 5 (Show Parameters) DWELL TIME SWEEP Store F1...
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 5.5. Calibration and Saturation test Menu After selecting CAL in any of the internal programs of the CWS 500N1.x or selecting the “Calibration and Saturation Test” in the main menu, the display will show the calibration stores. Structure of the calibration Menu EXIT MAIN MENU...
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 Main Menu Calibration and Saturation Calibration and Saturation Test F1 : Calibration F2 : Amplifier Saturation Test 5.5.1. Calibration Menu Calibration Calibration Procedure Calibration Calibration Which calibration - store should be overwritten? Calibration - Store F2 F1 : Calibration Store F1 valid...
Page 21
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 It is very important to check the calibration setup for correctness - Cabling - Ground reference is conductive - CDN or clamp is correct connected The CAL data of cal store can only be lost due to a discharged internal battery or by overwriting existing data. In case of battery discharge, the calibration stores are cleared and marked invalid.
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 5.5.2. Saturation check For the saturation check the generator increase to the calibrated signal with +5.1dB for check Amplifier Saturation Test 10.0 V CAL = F2 START CHANGE f = 000.100 MHz CHANGE: Voltage level of the calibrated signal Calibration storage file F1 to F5...
F7 Amplifier state This function indicates the amplifier temperature status measured at the internal amplifier module. State: State of TEST ON button, safety circuit, amplifier power and temperature (CWS 500N1.3) Temperature: (CWS 500N1.3) Operating temperature: typical 45°C to 56°C under labor conditions Max temperature: 60°C (The CWS will switch off the amplifier)
CWS 500N1.3 only 60 ºC: Firmware switches off the amplifier. After approx. 5 minutes the amplifier will switch on. The CWS 500N1.3 indicates the actual temperature in the display. 63 ºC: The hardware will switch off the amplifier automatically. Over temperature should not appear normally. Please check if the cooling airflow is not free or a ventilator is not working.
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 5.7. Remote display During Remote operation the display shows different screens depend the actual software procedure Waiting TEST ROUTINES F1 : Quickstart F2 : User test routines F3 : Standard test Testing R E M O T E ProgRemote MOD = OFF...
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 Starting Operation 6.1. Generator Function Check As with all testing equipment, the continuous wave simulator type CWS 500N1.x should be checked for performance and accuracy from time to time. The check should be conducted as listed below. - Connect the RF-output of the CWS 500N1.x to the 6dB-attenuator with a coaxial cable.
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 6.2. Function Check PM 1000 This function is for checking the PM 1000 measuring input MONITOR. With a measuring procedure, the CWS 500N1.x compares the two inputs with two different measurements. After the test the CWS verify the correct function of the two measuring.
Page 28
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 Step 2 Page 3 (Service ) Function check Powermeter step 2/2 - Connect the MONITOR input to the RF OUT via a ATT 6 dB attenuator Continue break Fig 6.3 :Setup step 2 with connected 6 dB attenuator Result after the test Finally the result of the check appears in the display.
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 6.3. Calibration and Test setup Overview A calibration starts at 100 kHz the first calibration and increases the next frequency 1% of the previous value. The calibration ends at approx. 240 MHz. 6.3.1.
Page 30
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 Calibration setup acc. IEC 61000-4-6 with amplifier level control as per Ed. 4 The standard IEC 61000-4-6 Ed.4 propose as alternative to use a dual directional coupler for control the forward power. The firmware controls the forward voltage during a test. With the icd.control software the user can select the method (Ed.
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 6.3.2. EM-Clamp / BCI Clamp Calibration setup acc. IEC 61000-4-6 Fig 6.8 : Test setup acc. IEC 61000-4-6 Fig 6.9 : User manual V 1.01 31 / 47...
Page 32
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 Calibration setup for EM Clamp / BCI acc. IEC 61000-4-6 with generator level control as per Ed. 4 The standard IEC 61000-4-6 Ed.4 propose as alternative to use a dual directional coupler for control the forward power.
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 Calibration Calibration files, generated with CWS500N1.x (*.cdx) are only applicable with standard test who are based on the standard IEC 61000-4-6. Tests based on other standards uses own calibration files (*.cal icd.control software) Information 7.1.
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 7.2. Calibration Setup with CDN Maximum Voltage on the Monitor input U = 1.0 Vrms (13dBm) The CDN and the 150 to 50 adapter must well connect ground reference plane (i.e. by using copper tape).
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 7.3. Calibration Setup with EM Clamp The calibration with an EM Clamp is similar to the CDN calibration. Fig 7.4 : Calibration setup with EM Clamp 7.4. Calibration Setup for Bulk Current Injection Clamp (BCI) Fig 7.6 : Calibration setup with BCI Attention During calibration of the test setup on a ground reference plane at very low levels (very low currents), the...
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 Setup Test Setup with CDN’s 8.1. The following information is important for a correct test setup: Fig 8.1 :Test setup with CDN • Each, the EUT and the AE are 10 cm isolated above the reference ground surface. •...
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 8.2. Test Setup with injection clamps (EM and BCI) Fig 8.3 :Test Setup using Injection clamps The CDN-side has no special requirements for the connected AE unit. The AE is placed 10cm above the reference ground plane.
Test Equipment CWS 500N1.x 9.1. Design of CWS 500N1.3 series The CWS 500N1.3 continuous wave simulator is divided into three main parts. The control unit is completely separated and uncoupled from the RF power section. I Control unit 1 Power Supply Filter...
CWS500 N1.3 / N1.4 9.2. Design of CWS 500N1.4 series The CWS 500N1.3 continuous wave simulator is divided into three main parts. The control unit is completely separated and uncoupled from the RF power section. I Control unit 1 Power Supply Filter...
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 9.6. Fail 1 & 2 Input The Fail 1 and Fail 2 input are located at the rear side of the CWS 500N1.x Fail 1: A short circuit at the Fail 1 input will stop the test procedure. It is not possible to continue this test.
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 9.7. 6dB Attenuator The 6dB attenuator is connected directly to the RF-output of the CWS 500N1.x The attenuator matches the output to a 50 system. The attenuator must be connected as close as possible to the coupling/decoupling network (CDN, EM clamp or current injection clamp), using a 50cm (20 inch) coaxial cable.
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 Technical Data Test Level CWS 500N1.3 : Output Level 1 - 30Ve.m.f. after 6dB-attenuator Output power 80W (nominal) Output impedance 50 Ohm max. VSWR 1:1.2 at all phase angles and at max. power (without destruction) Harmonic distortion (at max.
Page 43
Power 150 W max. Insertion loss 0.6 dB max. Mainline VSWR 1.1:1 max. RF Amplifier ( build in ) CWS 500N1.3 CWS 500N1.4 Class of Operation Class A Class A Frequency range 10 kHz .. 400 MHz 100 kHz .. 300 MHz...
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 Maintenance 11.1. General The generator is absolutely maintenance-free by using solid state semiconductors. The generator uses forced air cooling. Take care not to cover the cooling slots. 11.2. Calibration and Verification 11.2.1. Factory calibration Every EM TEST generator is entirely checked and calibrated as per international standard regulations before delivery.
EM TEST (Switzerland) GmbH CWS500 N1.3 / N1.4 Delivery Groups 12.1. Basic equipment Continuous wave simulator type CWS 500N1.3 or CWS 500N1.4 Mains cable Calibration certificate icd.control software on delivered USB memory stick 1 x BNC cable 0.5m, 1 x N cable 0.5 m + 1 x N cable 1.5 m 20 dB / 1W attenuator N-connector bridge (AMP OUT –...
CE directives listed below using the relevant section of the following EC standards and other normative documents. Product‘s name: Continuous Wave Simulator Model Number(s) CWS 500N1.3 CWS 500N1.4 Low Voltage Directive 2014/35/EU Standard to which conformity is declared: EN 61010-1 : 2011 Safety requirements for electrical equipment for measurement, control, and laboratory use.