Download Print this page
T&D RTR507B User Manual

T&D RTR507B User Manual

Wireless thermo recorder

Advertisement

Quick Links

Wireless Thermo Recorder
RTR501B / 502B / 503B / 507B
User's Manual
Thank you for purchasing our product.
To ensure safe and proper operation, please read this manual
thoroughly before use.
Overview
The RTR501B, RTR502B, RTR503B and RTR507B are wireless
data loggers designed to measure and record temperature
and humidity.
Recorded data is then automatically collected by the Base Unit
via wireless communication and downloaded for archive and
analysis.
As a Remote Unit, the RTR501B / 502B / 503B / 507B requires a Base
Unit to carry out wireless communication.
(Compatible Base Units: RTR500BC, RTR500BW, RTR-500MBS-A, RTR-
500DC, RTR-500, RTR-500NW/AW)
For the operation and configuration of the Base Unit, refer to
the instruction manual attached to the Base Unit or the RTR500B
Series Help available on T&D Website.
Wireless Thermo Recorder RTR501B / 502B / 503B / 507B is re-
ferred to as the "(data) logger" or "device" in this manual.
tandd.com
© Copyright T&D Corporation. All rights reserved.
2020.09 16508160003
(1st Edition) Printed on recycled paper.
Package Contents
RTR501B
RTR502B
Internal Temp
Temp Sensor
Sensor
TR-5106
(Included)
Common Accessories
• Lithium Battery LS14250 (or Large Capacity Battery Adaptor Kit RTR-500B1 for L type model)
• Strap (Not included with L type model)
• Manual Set (Warranty Included)
Part Names
Antenna
LCD Display
Optical
Communication
Port
Sensor
Jack
<Front>
Make sure that the sensor is completely
inserted until you hear a "click" sound.
RTR503B
RTR507B
Temp-Humidity
High Precision
Sensor
Temp-Humidity Sensor
TR-3310 (Included)
SHB -3101 (Included)
<Side>
<Side>
<Back>
(L type)
Installing the Battery
Recording starts automatically by inserting the battery with the
default or previous settings.
Battery
Cover
Screw
Rubber
Packing
Lithium Battery
Battery Replacement
When it is time for the battery to be
replaced, a battery warning mark will
appear. Please change the battery as
soon as possible if this mark appears.
If you continue to use the logger without changing the battery, the
current temperature and [bAtt] will be displayed alternately and
wireless communication will stop. (Recording will continue.)
• If the battery is further left unchanged, the display will automatically shut
off and all previously recorded data will be lost.
• Although the logger continues to work for a period after the battery has
been removed, leaving the device without a battery until the LCD display
goes blank will cause all recorded data to be lost.
Remote Unit Registration
Using the software of the compatible Base Unit, carry out Remote
Unit Registration via optical or Bluetooth.
Optical Communication
Connect the Base Unit to a PC with a USB cable, and place the data
logger face down to align the communication areas as shown be-
low.
®
Bluetooth
Communication
When the RTR500BW is a Base Unit, it is possible to make Base Unit
and Remote Unit settings from nearby mobile devices using Blue-
tooth.
Default Settings
Recording Mode
Endless
Recording Interval
10 minutes
Recording Start
Immediate Start
Method
• Make sure to use the proper type and
size screwdriver. (Phillips head #1
screwdriver is recommended.)
• Insert the supplied battery with the
tube attached. When using a CR2 lithi-
um battery, the tube is not necessary.
• Before closing the cover, check the
rubber packing for dust or scratches,
as they can reduce the water resis-
tance of the rubber.
• Be sure to completely close the cover.
Make sure not to over tighten the
screws.
• Appropriate Tightening Torque:
20N·cm to 30N·cm (2Kgf·cm to 3Kgf·cm)
Alternating
Display

Advertisement

loading

Summary of Contents for T&D RTR507B

  • Page 1 • Insert the supplied battery with the Overview tube attached. When using a CR2 lithi- um battery, the tube is not necessary. The RTR501B, RTR502B, RTR503B and RTR507B are wireless Rubber data loggers designed to measure and record temperature • Before closing the cover, check the Packing and humidity.
  • Page 2 Unit of Measure- Shows the unit of measurement. Mark ment Specifications Model RTR501B / 501BL RTR502B / 502BL RTR503B / 503BL RTR507B / 507BL Measurement Temperature 1ch Temperature 1ch Temperature 1ch, Humidity 1ch Temperature 1ch, Humidity 1ch Channels Sensor Thermistor (Internal)

This manual is also suitable for:

Rtr503bRtr502bRtr501b