schwa-medico TENS eco 2 Instructions For Use Manual page 5

Transcutaneous nerve and muscle stimulator
Hide thumbs Also See for TENS eco 2:
Table of Contents

Advertisement

7.
Warning:
other devices,
use in the manner described above is still necessary, this device and the
other devices should be observed to ensure that they are working pro-
perly.
8.
Device mains plugs and/or socket outlet must be easily
accessible
the mains supply is possible quickly if necessary.
9.
Caution: The used electrodes should not be smaller than
2 cm²
as it may lead to high current densities. Any electrodes with cur-
rent densities exceeding 2 mA/cm² may require the special attention of
the operator as it may lead to high current densities.
10. Caution:
wave ME (medical electrical) device for shortwave or microwave thera-
py may cause fluctuations in the output value of the product which may
turn into painful reactions.
11.
Caution:
surgical ME (medical electrical) equipment may lead to burns beneath
the electrodes and it may damage to the product.
12. Caution:
Attach the electrodes to the skin so that the electrode sur-
face has even and complete skin contact. Additionally make sure that
the electrodes are placed at least 2 cm apart from each other. Other-
wise, high current densities may occur at certain points on the skin, re-
sulting in painful skin lesions.
13.
Before applying the electrodes, clean the skin surface on which the elec-
trodes are to be attached. Otherwise incorrect operation cannot be ex-
cluded.
14. Do not allow the product to be dropped and handled incor-
rectly.
Only use at temperatures between 10 °C and 40 °C, at a relative
humidity between 30 % and 75 % and a pressure between 70 kPA and 106
kPA.. So do not use the product in bathroom or similar humid environments.
Warning:
flammable substances or steam!
Caution:
from cold to warm, do not turn on the device until it reaches the same
temperature as the environment in which it will be used; wait at least
30 minutes. Not doing so may cause condensation inside the unit which
may result in electric shocks, fire, damage to the device and/or personal
injuries.
15.
The product is intended for use by one patient only.
Do not use this device next to or stacked with
as this may result in deficient operation. If, however,
during operation of the device, so that disconnection from
Operation in the vicinity (e.g. 1m) of a short wave or micro-
Simultaneous connection of a patient to a high frequency
Do not operate the device in the vicinity of explosive and/or
If you expose this device to sudden temperature changes
5