Overview free programs

Progr.-
Program
Indication
no.
name
U 1
Gate
Acute nociceptive
Control 1
pain, acute pain,
chronic neuropathic
pain
U 2
Gate
Alternative to
Control 2
program 1
U 3
Low
Chronic nociceptive
frequency
pain, improvement
of
blood circulation,
Kaada stimulation
U 4
Gate
Combination of
Control +
high and low fre-
Low
quency
Frequency
U 5
Adjust-
Muscular pain
ment
U 6
HAN
For nearly all pain
indications
U 7
Burst
Alternative to pro-
gram no. 3 (more
comfortable)
U 8
Modula-
Alternative to other
tion
programs if resis-
tant to therapy
U 9
Muscle
Prophylaxis treat-
Stimula-
ment of muscle
tion
atrophy
U 10
Gate
Muscle relaxation
Control
Dynamic
slow
U 11
Gate
Muscle relaxation
Control
Dynamic
fast
U 12
Deep
Muscle pain, pain
TENS
in deep areas
Description
Both channels with the same
parameters
Both channels with the same
parameters
Both channels with the same
parameters
Channel 1 and channel 2
work with different para-
meters
First 10 min.: 100 Hz
Then 20 min.: 2 Hz
alternating
3 s 100 Hz
3 s 2 Hz
0.25 s long impulse with a
frequency to be set
Automatic frequency run
from one frequency to ano-
ther within 7.5 s
Increase 2 s, working time 5
s, decrease 1 s, pause 12 s,
frequency to be set, automa-
tic intensity adaption
Intensity within 1 s to maxi-
mum, then within 1 s redu-
ced to 50%. Both channels
in alternating operation
Intensity within 0.25 s to
maximum, then within 0.25 s
reduced to 50%. Both chan-
nels in alternating operation
Impulse packages of 4 impul-
ses with a pulse width to be
set working with a frequency
carrier to be set.
English
Fre-quency/
Pulse
Hz
width/
µs
1-120
50-400
1-120
50-400
1-120
50-400
Channel 1:
50-400
4-120
Channel 2:
2 Hz fixed
First fre-
50-400
quency
1-120
First fre-
50-400
quency
1-120
50-120
50-400
First fre-
First
quency:
pulse
1-50
width:
Last fre-
100-400
quency:
80
1-120
50-400
10-120
50-400
50-120
50-400
1-120
50-400
11
Timer/
min.
1-99
1-99
1-99
1-99
1-99
1-99
1-99
1-99
1-99
1-99
1-99
1-99