Starting the Kaye Validator AVS Software

On the Console home page, click the Kaye Validator AVS tile. The Kaye Validator AVS user login screen
appears, followed by the Hub page. The Hub page provides access to the major functions of the Kaye
Validator software.
8.3.

Creating/Modifying a Setup

The setup defines everything required to calibrate sensors, run qualification studies, and verify sensor calibration.
The setup is created/modified using the Kaye Validator software. The Console does not have to be connected to
the Kaye Validator AVS to create/modify setups.
Procedure
1.
On the Hub page, click Assets tile, then on Asset Details page click Setups tab.
2.
Existing setups are listed by the date and time the files were saved.
To create a new setup, click New, enter up to 35 characters to describe the setup, and click OK.
To modify an existing setup, select the setup from the file list and click on edit icon.
The Sensors screen appears.
3.
Add, modify or delete sensor configurations as necessary.
Select sensors by clicking a sensor tiles in the SIM.
To add or modify a sensor definition, enter/modify the sensor label and sensor description, and select a
sensor type. Specify scale factors as required. To automatically number multiple sensor definitions,
select both Automatic Numbering check boxes to append 2 digits in sequence to the sensor label and
description label. Click OK.
4.
To remove sensor location(s), click the sensor tile again to deselect it.
5.
Identify/modify groups.
Click the Group Sensors tab.
Change group titles or add groups as necessary (up to 25 groups, 35 characters) in the group list box.
6.
Assign sensors to groups.
Select a group from the list box.
Assign sensor locations as necessary.
Repeat for all groups as necessary.
Note: Sensors of different types should not be mixed in a group.
6.
Specify calculations and events.
Click the Calculations tab.
Specify calculations as necessary.
Define the event as necessary.
Repeat for all sensors as necessary.
7.
Specify calibration parameters.
Click the Calibration tab.
Set the Temperature Setpoints as necessary.
Set the Stability Criteria as necessary.
Set the Deviation Criteria as necessary.
8.
Specify qualification study conditions.
Standard Operating Procedure for the Kaye Validator® ® ® ® AVS System
Revision: A
© 2015 Amphenol Thermometrics, Inc
Standard Operating Procedure
PAGE: 7 of 16

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Kaye Validator AVS and is the answer not in the manual?

Questions and answers

Table of Contents