Document #:
Kaye Validator
Standard Operating Procedure
Version 1.0
This is a version controlled document that is released with Amphenol internal approvals.
The Approvals section is provided for optional customer use to approve and accept this
document. It may be amended, deleted or marked as N/A at the customer's discretion.
Approvals
(Name and Title)
Technical
_____________________________________________
Reviewer:
Validation:
_____________________________________________
Technical
_____________________________________________
Services:
Quality Control:
_____________________________________________
Management:
_____________________________________________
Standard Operating Procedure for the Kaye Validator® ® ® ® AVS System
Revision: A
AVS
® ® ® ®
© 2015 Amphenol Thermometrics, Inc
Standard Operating Procedure
Date: _________________
Date: _________________
Date: _________________
Date: ________________
Date: _________________

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Kaye Validator AVS and is the answer not in the manual?

Questions and answers

Summary of Contents for Amphenol Kaye Validator AVS

  • Page 1 Standard Operating Procedure Version 1.0 This is a version controlled document that is released with Amphenol internal approvals. The Approvals section is provided for optional customer use to approve and accept this document. It may be amended, deleted or marked as N/A at the customer’s discretion.
  • Page 2: Table Of Contents

    Temperature Reference ...........................5 6.5. IRTD ................................5 SYSTEM ACCESS ..............................5 PROCEDURES ..............................6 8.1. Wiring SIMs ..............................6 8.2. Starting the Kaye Validator AVS Software ......................7 8.3. Creating/Modifying a Setup ..........................7 8.4. Connecting the System ............................9 8.5. Checking Hardware and Communications Connections..................9 8.6.
  • Page 3: Purpose

    Kaye Validator AVS audit trail. This procedure assumes that the Kaye Validator AVS software version 1.0 has been installed and the System Administrator has created user accounts. See Kaye Validator AVS User’s Guide for more information.
  • Page 4: Reference

    Generate reports from secure data files that contain original calibration and qualification data recorded during the study. Access the Kaye Validator AVS audit trail, which provides a complete listing of events that affect the integrity of the Kaye Validator AVS software and the Kaye Validator AVS instrument.
  • Page 5: Temperature Reference

    PAGE: 5 of 16 6.4. Temperature Reference Temperature references provide the stable temperature required for sensor calibration. Amphenol Advanced Sensors has five temperature reference models available. They are designed to provide different temperature ranges and are compatible with the Kaye Validator.
  • Page 6: Procedures

    AMPHENOL ADVANCED SENSORS also recommends shaving open about 1/4 inch from each of the red and blue inner wire jackets at opposite ends of the outer insulation drip cut to allow condensate drainage from within the individual wires.
  • Page 7: Starting The Kaye Validator Avs Software

    Starting the Kaye Validator AVS Software On the Console home page, click the Kaye Validator AVS tile. The Kaye Validator AVS user login screen appears, followed by the Hub page. The Hub page provides access to the major functions of the Kaye Validator software.
  • Page 8 Set the Output Relay to actuate based on “event” if necessary. Save the setup file. Click Save. The Save Setup File screen appears. Enter User Id and password. Click OK. The Setup tile screen appears. 10. Click home to return to the Home page. © 2015 Amphenol Thermometrics, Inc...
  • Page 9: Connecting The System

    PAGE: 9 of 16 8.4. Connecting the System All system connections are on the back of the Kaye Validator AVS. Each connection port is labeled with an icon representing its function. Connection ports are as follows: 2 IRTD ports (4 pin DIN connector)
  • Page 10: Operation

    (i.e., close to a sterilizer, an open door causing a draft, etc.). Power up the Kaye Validator AVS and let it run for approximately 30 minutes in the operating environment in order to acclimate to the ambient temperature.
  • Page 11: Qualification Study

    Kaye Validator AVS. Procedure If not already on, power up the Kaye Validator AVS and let it run for approximately 30 minutes in the operating environment in order to acclimate to the ambient temperature.
  • Page 12: Calibration Verification

    Kaye temperature reference and IRTD. Procedure If not already on, power up the Kaye Validator AVS, temperature reference and IRTD and let them run for approximately 30 minutes in the operating environment in order to acclimate to the ambient temperature.
  • Page 13: Generating Reports

    Setup Report. Procedure: Select the report file. On the Console, either go to Set Up tab of Asset details screen of Kaye Validator AVS application. Select Setup and click on edit icon. Go to review pane and generate Report in PDF.