Arm; Count - HP E1346A User Manual

Relay multiplexer
Table of Contents

Advertisement

ARM

Subsystem Syntax

:COUNt

Parameters
Comments
Example
:COUNt?
Parameters
Comments
Example
Chapter 5
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The ARM subsystem selects the number of scanning cycles (1 through
32767) for each INITiate command.
ARM
:COUNt <number>MIN |MAX
:COUNt? [MIN |MAX]
ARM:COUNt <number>allows scanning cycles to occur a multiple of times
(1 to 32767) with one INIT command and INIT:CONT OFF is set.
Parameter Name
Parameter Type
number
numeric
Number of Scans: Use only values between 1 to 32767 for the
number of scanning cycles.
Related Commands: ABORt, INITiate:IMMediate
*RST Condition: ARM:COUNt 1
Setting Ten Scanning Cycles
ARM:COUN 10
SCAN (@100:147)
INIT
ARM:COUNt? [MIN |MAX] returns current number of scanning cycles set
by ARM:COUNt. A value is supplied if no MIN or MAX parameter is sent.
With MIN or MAX passed as parameters, MIN returns 1 and MAX returns
32767.
Parameter Name
Parameter Type
MIN |MAX
numeric
Related Commands: INITiate:IMMediate
Query Number of Scanning Cycles
ARM:COUN 10
ARM COUN?
HP E1346A Multiplexer Command Reference 43
Range of Values
1 |32767 |MIN |MAX
Set 10 scanning cycles
Sets channel list
Starts scanning cycle
Range of Values
MIN = 1, MAX = 32767
Sets 10 scanning cycles
Query number of scanning cycles;
returned value is 10