MT 600
IB MT 600 EN-PL 3920, Revision Date: 2020-09-15
4
Description of this Thermometer
ON/OFF button
1
Display
2
Battery compartment cover
3
Measuring sensor / measuring tip
4
Cleaning and disinfecting area (entire thermometer surface)
5
Important Safety Instructions
 Follow instructions for use. This document provides important
product operation and safety information regarding this device.
Please read this document thoroughly before using the device and
keep for future reference.
 This device is only to be used for measuring human body tempera-
ture through oral, rectal or axillary. Do not attempt to take tempera-
tures at other sites, such as in the ear, as it may result in false read-
ings and may lead to injury.
 Do not use this device if you think it is damaged or notice anything
unusual.
 We recommend cleaning this device according to the cleaning
instructions before first use for personal hygiene.
 The minimum measurement time until the beep is heard must be
maintained without exception!
Consider that different measurement locations may require
continued measuring even after the beep, see section «Measuring
methods / Normal body temperature».
 For safety reasons (risk of rectal perforation) rectal measurement in
children younger than 3 years must be performed only by trained
health care personnel (professional user). Use another measurement
method instead. For rectal fever measurement in small children
younger than 3 years, thermometers with a flexible tip are available.
 Do not attempt rectal measurements on persons with rectal disor-
ders. Doing so may aggravate or worsen the disorder.
 Ensure that children do not use this device unsupervised; some
parts are small enough to be swallowed. Be aware of the risk of
strangulation in case this device is supplied with cables or tubes.
 Do not use this device close to strong electromagnetic fields such as
mobile telephones or radio installations. Keep a minimum distance
of 3.3 m from such devices when using this device.
 Protect the device from impact and dropping!
 Avoid ambient temperatures above 60 °C. NEVER boil this device!
 Use only the commercial disinfectants listed in the section «Cleaning
and Disinfecting» to clean the device to avoid damage to the device.
 We recommend this device is tested for accuracy every two years or
after mechanical impact (e.g. being dropped). Please contact your
local Microlife-Service to arrange the test.
WARNING: The measurement result given by this device is
not a diagnosis! Do not rely on the measurement result only.
Batteries and electronic devices must be disposed of in
accordance with the locally applicable regulations, not with
domestic waste.
Read the instructions carefully before using this device.
Type BF applied part
Turning on the Thermometer
To turn on the thermometer, press the ON/OFF button 1; a short
beep signals «thermometer ON». A display test is performed. All
segments should be displayed.
Then at an ambient temperature of less than 32 °C, an «-- -» and a
flashing «°C» appear at the display field 2. The thermometer is now
ready for use.
Function Test
Correct functioning of the thermometer is tested automatically each
time it is turned on. If a malfunction is detected (measurement inaccu-
racy), this is indicated by «ERR» on the display, and a measurement
becomes impossible. In this case, the thermometer must be replaced.
Using the Thermometer
Choose the preferred measuring method. When taking a measure-
ment, the current temperature is continuously displayed and the «°C»
symbol flashes. If the beep is heard 10 times and the «°C» is no longer
flashing, this means that the measured increase in temperature is less
than 0.1 °C in 6 seconds.
To prolong the battery life, turn off the thermometer by briefly pressing
the ON/OFF button 1. Otherwise the thermometer will automatically
turn off after about 10 minutes.
Storage of Measured Values
If the ON/OFF button 1 is pressed for more than 3 seconds when
turning on the thermometer, the automatically stored maximum
temperature during the last measurement will be displayed. At the
same time, a «M» for memory will appear on the display. About
2 seconds after the button is released, the temperature value disap-
pears and the thermometer is ready for measurement.
Measuring methods / Normal body temperature
 In the armpit (axillary) / 34.7 - 37.3 °C
Wipe the underarm with a dry towel. Place the measuring sensor 4
under the arm into the center of the armpit so the tip is touching the
skin and position the patient's arm next to the patient's body. This
ensures that the room air does not affect the reading. Because the
axillary takes more time to reach its stable temperature wait at least
5 minutes, regardless of the beep sound.
 In the mouth (oral) / 35.5 - 37.5 °C
Do not eat or drink anything hot or cold 10 minutes before the
measurement. The mouth should remain closed up to 2 minutes
before starting a reading.
Position the thermometer in one of the two pockets under the
tongue, to the left or right of the root of the tongue. The measuring
sensor 4 must be in good contact with the tissue. Close your mouth
and breathe evenly through the nose to prevent the measurement
from being influenced by inhaled/ exhaled air.
If this is not possible due to blocked airways, another method for
measuring should be used.
Approx. measuring time: 1 minute!
 In the anus (rectal) / 36.6 - 38.0 °C
Attention: For the prevention of rectal perforation in children
(younger than 3 years), we recommend using another measuring
method, or the use of a thermometer with flexible tip.
Carefully insert the measuring sensor 4 of the thermometer 2 to
3 cm into the anal aperture.
The use of a probe cover and the use of a lubricant is recommended.
If you are unsure of this measurement method, you should consult a
professional for guidance/training.
Approx. measuring time: 1 minute!
Cleaning and Disinfecting
For disinfection in home use environment, use a 70% Isopropyl alcohol
swab or a cotton tissue moistened with 70% Isopropyl alcohol to wipe
surface pollutants off the thermometer probe. Always start wiping from
the end of the thermometer probe (approx. at the middle of the ther-
mometer) towards the thermometer tip. Afterwards the entire thermom-
eter (see number 5 in the drawing) should be immersed in 70%
Isopropyl alcohol for at least 5 minutes (max. 24 hours). After immer-
sion, let the disinfectant dry off for 1 minute before next use.
For professional use: You may apply probe covers (available as
accessory). Contact Microlife for further details concerning suitable
disinfectants for your thermometer model.
Battery Replacement
When the «» symbol (upside-down triangle) appears in the display,
the battery is flat and needs replacing. To replace the battery remove
the battery compartment cover 3 from the thermometer. Insert the
5
2
EN
new battery with the + at the top. Make sure you have a battery of the
same type to hand. Batteries can be purchased at any electrical store.
Technical Specifications
Type:
Measurement range:
Measurement accuracy: ± 0.1 °C; 34 °C - 42 °C
Operating conditions:
Storage conditions:
Battery:
Battery lifetime:
IP Class:
Reference to
standards:
Expected service life:
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.
Guarantee
This device is covered by a lifetime guarantee from the date of
purchase. During this guarantee period, at our discretion, Microlife will
repair or replace the defective product free of charge.
Opening or altering the device invalidates the guarantee.
The following items are excluded from the guarantee:
 Transport costs and risks of transport.
 Damage caused by incorrect application or non-compliance with the
instructions for use.
 Damage caused by leaking batteries.
 Damage caused by accident or misuse.
 Packaging/storage material and instructions for use.
 Regular checks and maintenance (calibration).
 Accessories and wearing parts: Battery.
Should guarantee service be required, please contact the dealer from
where the product was purchased, or your local Microlife service. You
may contact your local Microlife service through our website:
www.microlife.com/support
Compensation is limited to the value of the product. The guarantee will
be granted if the complete product is returned with the original invoice.
Repair or replacement within guarantee does not prolong or renew the
guarantee period. The legal claims and rights of consumers are not
limited by this guarantee.
Microlife AG, Espenstrasse 139
9443 Widnau / Switzerland
1
3
Maximum thermometer
32.0 °C to 43.9 °C
Temp. < 32.0 °C: display «-- -» for low (too low)
Temp. > 43.9 °C: display «H» for high (too high)
± 0.2 °C; 32.0 - 33.9 °C and 42.1 - 43.9 °C
10 - 40 °C; 15-95 % relative maximum humidity
-25 - +60 °C; 15-95 % relative maximum humidity
LR41 (1.5V) / SR41 (1.55V)
approx. 4500 measurements (using a new battery)
IP67
EN 12470-3, clinical thermometers;
ASTM E1112; IEC 60601-1; IEC 60601-1-2 (EMC);
IEC 60601-1-11
5 years or 10000 measurements

Advertisement

loading
Need help?

Need help?

Do you have a question about the MT 600 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Microlife MT 600

  • Page 1 Should guarantee service be required, please contact the dealer from mobile telephones or radio installations. Keep a minimum distance where the product was purchased, or your local Microlife service. You of 3.3 m from such devices when using this device.
  • Page 2 Do użytku profesjonalnego: Możesz zastosować osłony na sondy bezpiecznego korzystania z tego urządzenia. Przeczytaj dokładnie (dostępne jako akcesoria). Skontaktuj się z Microlife fw celu uzyskania ten dokument przed pierwszym użyciem i zachowaj go na dalszych informacji dotyczących odpowiednich środki dezynfekujące przyszłość.