Troubleshooting And Faqs; Error Messages; Faq - GenScript eZwest Lite Manual

Automated western device
Hide thumbs Also See for eZwest Lite:
Table of Contents

Advertisement

4. Troubleshooting and FAQs

Error messages

Error Message
Buffer Error
No Tank
Pump Error
System Error

FAQ

Description
Strong background
Weak signal
Fluorescent detects
only strong signal
around spacer
Possible causes
Can't read cassette chip/no
cassette
Cassette not secured in
position
Pump malfunction
System malfunction
Possible causes
Insufficient wash
1. Antibody is too
diluted
2. Instrument needs
cleaning
3. Insufficient
antibody
incubation time
Weak target signal
contrasted out by
interference
Solutions
Replace cassette/ insert
cassette
Check Pins and Reinsert
cassette securely
Contact GenScript
Contact GenScript
Solutions
1. Increase number
2. Increase pump
1. Reduce antibody
2. Clean eZwest Lite
3. Increase antibody
1. Adjust detection
2. Cover spacer
20
of wash cycles (set
to 5 cycles for 3
minutes each)
flow per wash
cycles (set to
normal)
dilution factor
instrument
incubation time or
reduce incubation
pump flow
area to omit
spacer position
signal region with
paper or tape